VectorStarâ„¢ MS4640A SeriesMicrowave Vector Network Analyzer : SCPI Commands : :CONTrol:AUXio Subsystem
 
:CONTrol:AUXio Subsystem
The :CONTrol:AUXio subsystem commands controls the rear panel limit failure configuration and display.
Rear Panel Subsystems
A related rear panel subsystem is:
• :CONTrol{1-16}:AOUT Subsystem
 
:CONTrol:AUXio:LIMit:POLarity <char1>
:CONTrol:AUXio:LIMit:POLarity?
Description
Sets the Positive/Negative Polarity of the Limit Fail indication on the Rear Panel Auxiliary IO connector. Outputs the Positive/Negative Polarity of the Limit Fail indication on the Rear Panel Auxiliary IO connector.
Cmd Parameters
<char1> POSitive | NEGative
Query Parameters
<char1> POS | NEG
Range
NA
Default Value
POS
Syntax Example
:CONT:AUX:LIM:POL POS
:CONT:AUX:LIM:POL?
 
:CONTrol:AUXio:LIMit[:STATe] <char1>
:CONTrol:AUXio:LIMit[:STATe]?
Description
Turns on/off indicating a limit failure on the Rear Panel Auxiliary I/O connector. Outputs the on/off status of indication of a limit failure on the Rear Panel Auxiliary I/O connector.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Query Parameters
<char1> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CONT:AUX:LIM ON
:CONT:AUX:LIM?