This set of commands prepares the instrument for the selected measurement. It disables any currently-enabled measurements and activates the specified measurement. It sets the instrument to single sweep mode, waiting for an :INITiate command. It will not initiate the taking of a measurement.
Current instrument settings may be changed to default values. These changes are identified with their respective measurement commands.
:CONFigure?
Title
Configure Query
Description
:CONFigure? query returns the name of the measurement previously set up using a CONFigure command or a MEASure? query. The list below shows the possible return values.
Sets the measurement to LTE Summary. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
Sets measurement to one of the Modulation screens. This can be one of:
SUMMary: Modulation Summary view
CONStln: Modulation Constellation diagram.
CCPGraph: Control Channel Powers - Bar-graph view.
CCPTable: Control Channel Powers - Table view.
It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
Parameter
SUMMary|CONStln|CCPGraph|CCPTable
Parameter Type
<char>
Front Panel Access
Measurements, Modulation, (Constellation, Control Channel Power, Modulation Summary).
If Control Channel Power is selected, press Control Channel Power again then press Display Mode to select Bar Graph or Table.
:CONFigure:OTA SCANner|TXTEst|MAPping
Title
Configure Over-the-Air Measurement
Description
Sets the measurement to one of the Over-the-Air screens.
Parameter
SCANner|TXTEst|MAPping
Parameter Type
<char>
Front Panel Access
Measurements, Over-the-Air, Scanner
:CONFigure:RF SUMMary|SPECtrum|ACLR|SEM
Title
Select RF Measurements
Description
This command configures RF measurements. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF).Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
Set measurement to one of the RF measurement screens. This can be one of: