S331E, S332E, S361E, S362E Site Master™MS2711E, MS2712E, MS2713E Spectrum Master™MT8212E, MT8213E Cell Master™ Programming Manual : LTE Commands : :SENSe Subsystem
 
:SENSe Subsystem
The commands in this subsystem relate to device‑specific parameters, not signal‑oriented parameters.
[:SENSe]:TAE:ACTive OFF|ON
[:SENSe]:TAE:ACTive?
Title
Time Alignment Error (4x4 MIMO configuration)
Description
Sets/Queries the state of the TAE measurement for OTA carrier aggregation. In 2x2 MIMO, use the command [:SENSe]:TAE:ACTive.
Parameter
OFF|ON
Parameter Type
<char>
Front Panel Access
Measurements, Over‑the‑Air, Carrier Aggregation, TAE
[:SENSe]:BANDWidth[:RESolution] 1.4|3|5|10|15|20
[:SENSe]:BANDWidth[:RESolution]?
Title
Bandwidth Resolution
Description
Sets/Queries the channel bandwidth. Note that available bandwidths may be instrument and option dependent.
Parameter
1.4|3|5|10|15|20
Parameter Type
<char>
Default Value
The default bandwidth is 10 MHz.
Front Panel Access
Setup, BW
[:SENSe]:CC{1|2|3|4|5}:ACTive OFF|ON
[:SENSe]:CC{1|2|3|4|5}:ACTive?
Title
Active Component Carrier
Description
Sets/Queries the state of the specified Component Carrier (CC) to be on or off.
Parameter
OFF|ON
Parameter Type
<boolean>
Front Panel Access
Measurements, Over‑the‑Air, Carrier Aggregation, CC#, CC#
[:SENSe]:CC{1|2|3|4|5}:BANDWidth 1.4|3|5|10|15|20
[:SENSe]:CC{1|2|3|4|5}:BANDWidth?
Title
Component Carrier Bandwidth
Description
Sets the channel bandwidth of the specified Component Carrier (CC).
Parameter
1.4|3|5|10|15|20
Front Panel Access
Measurements, Over‑the‑Air, Carrier Aggregation, CC#, CC#, BW
[:SENSe]:CC{1|2|3|4|5}:CHANnel <number>
[:SENSe]:CC{1|2|3|4|5}:CHANnel?
Title
Component Carrier Channel
Description
Sets/Queries the channel number for the selected signal standard for the specified Component Carrier (CC).
Parameter
<number>
Front Panel Access
Measurements, Over‑the‑Air, Carrier Aggregation, CC#, CC#, Channel
[:SENSe]:CC{1|2|3|4|5}:FREQuency <freq>
[:SENSe]:CC{1|2|3|4|5}:FREQuency?
Title
Component Carrier Frequency
Description
Sets/Queries the center frequency of the specified Component Carrier (CC).
Parameter
<freq>
Parameter Type
<NRf>
Front Panel Access
Measurements, Over‑the‑Air, Carrier Aggregation, CC#, CC#, Center Freq
[:SENSe]:CC{1|2|3|4|5}:SIGStandard <string>
[:SENSe]:CC{1|2|3|4|5}:SIGStandard?
Title
Component Carrier Signal Standard
Description
Sets/Queries the desired signal standard from the list. The <string> argument is the name of the desired signal standard as displayed in the instrument’s current signal standard list. The list can be displayed on the instrument by choosing the Signal Standard submenu button in the Freq menu. The list can also be downloaded remotely and viewed using Anritsu Master Software Tools. For example, if the desired Signal Standard is P-GSM 900 - Uplink, then the value of the <string> argument would be “P-GSM 900 - Uplink”.
The query form of this command returns the name of the currently selected Signal Standard from the list.
Parameter
<string>
Parameter Type
<char>
Front Panel Access
Measurements, Over‑the‑Air, Carrier Aggregation, CC#, CC#, Signal Standard
[:SENSe]:CELLID <char>
[:SENSe]:CELLID?
Title
Cell ID
Description
Sets/Queries the Cell ID value used when Sync Type is set to ‘RS’.
Parameter
(the known Cell ID)
Parameter Type
<char>
Front Panel Access
Setup, Sync (RS), Cell ID
[:SENSe]:EMF:STATe OFF|ON|0|1
[:SENSe]:EMF:STATe?
Title
EMF State
Description
Switches EMF ON or OFF.
The query form of this command returns a 0 or 1 when EMF state is OFF or ON, respectively.
Note: EMF turnsturn ON only if the start and stop frequencies are within the frequency range of the signal analyzer and isotropic antenna used. The antenna must be connected.
Parameter
OFF|ON|0|1
Parameter Type
<boolean>
Default Value
OFF
Front Panel Access
Measurements, Over‑the‑Air, EMF
[:SENSe]:EVM:MODe AUTO|PBCHonly
[:SENSe]:EVM:MODe?
Title
Set EVM Mode
Description
Sets/Returns the EVM Mode to either Auto or PBCH Only.
Parameter
AUTO|PBCHonly
Parameter Type
<char>
Default Value
AUTO
Front Panel Access
Setup, EVM Mode
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer?
Title
Center Frequency
Description
Sets the center frequency.
Parameter
<freq>
Default Value
1 GHz
Default Unit
Hz
Range
10 MHz to 4 GHz
Front Panel Access
Freq, Center Freq
[:SENSe]:FREQuency:SIGStandard:CHANnel <number>
[:SENSe]:FREQuency:SIGStandard:CHANnel?
Title
Channel Selection
Description
Sets the channel number for the selected signal standard.
Parameter
<number>
Front Panel Access
Frequency, Channel
[:SENSe]:FREQuency:SIGStandard:NAMe [String]
[:SENSe]:FREQuency:SIGStandard:NAMe?
Title
Signal Standard
Description
Selects the desired signal standard from the list. The <string> argument is the name of the desired signal standard as displayed in the instrument’s current signal standard list. The list can be displayed on the instrument by choosing the Signal Standard submenu button in the Freq menu. The list can also be downloaded remotely and viewed using Anritsu Master Software Tools. For example, if the desired Signal Standard is P‑GSM 900 - Uplink, then the value of the <string> argument would be “P‑GSM 900 - Uplink”.
The query form of this command turnsreturn the name of the currently selected Signal Standard from the list.
Parameter
<string>
Front Panel Access
Freq, Signal Standard
[:SENSe]:LTE:STATus?
Title
Sweep Status
Description
Returns 1 when the sweep is complete. Returns 0 when the sweep is in progress.
Front Panel Access
None
[:SENSe]:POWer[:RF]:RANGe[:IMMediate]
Title
Amplitude Range
Description
Re‑calculates amplitude range. Note that issuing this command sets the automatic dynamic range to OFF.
Related Command
[:SENSe]:POWer[:RF]:RANGe:AUTO
Front Panel Access
Amplitude, Adjust Range
[:SENSe]:POWer[:RF]:RANGe:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:RANGe:AUTO?
Title
Automatic Amplitude Range
Description
Sets the automatic amplitude range. Setting the value to ON or 1 results in the amplitude range being coupled to the detected input signal level. Setting the value to OFF or 0 results in the input attenuation being uncoupled from the input signal level. That is, changing the input signal level turnsnot change the amplitude range. When this command is issued, the amplitude range itself turnsnot change. The default value is ON. That is, sending :SENS:POW:RANG:AUTO is equivalent to sending :SENS:POW:RANG:AUTO ON.
Parameter
OFF|ON|0|1
Parameter Type
<boolean>
Default Value
ON
Related Command
[:SENSe]:POWer[:RF]:RANGe
Front Panel Access
Amplitude, Auto Range
[:SENSe]:RF:SPECtrum:SPAN Auto|1.4|3|5|10|15|20|30
[:SENSe]:RF:SPECtrum:SPAN?
Description
Sets the span for the Spectrum view. For example, setting the value to 5 sets the span for the Spectrum view to 5 MHz and setting the value to 20 sets the span for the Spectrum view to 20 MHz.
Parameter
Auto|1.4|3|5|10|15|20|30
Default Value
Auto
Default Unit
MHz
Front Panel Access
Measurements, RF, Channel Spectrum, Span
[:SENSe]:SYNC:TYPe SS|RS
[:SENSe]:SYNC:TYPe?
Title
Set Sync Type
Description
Sets the sync type setting.
Parameter
SS|RS
Parameter Type
<char>
Default Value
SS
Front Panel Access
Setup, Sync, Sync Type
[:SENSe]:TAE:ACTive OFF|ON
[:SENSe]:TAE:ACTive?
Title
Time Alignment Error
Description
Sets the state of the TAE measurement for OTA carrier aggregation.
Parameter
OFF|ON
Parameter Type
<boolean>
Front Panel Access
Measurements, Over‑the‑Air, Carrier Aggregation, TAE