The instrument can display 1-port phase measurements. The following example compares the phase of two cables using a 1-port phase measurement.
1. Press the Measurements main menu key
2. Press the More submenu key.
3. Press the 1-Port Phase key.
4. Press the Freq/Dist main menu key and set the start frequency and stop frequency.
5. Press Shift and Calibrate (2) to calibrate the instrument. See Calibration for details.
6. Connect device under test (Cable A) and press Copy Trace To Display Memory.
7. Remove the first device under test and connect the second device under test (Cable B).
8. Press the Trace – Memory key to view the difference between Cable A and Cable B.
Smith Chart
The instrument can display 1-port measurements in a standard Normalized 50 ohm Smith Chart. When markers are used, the real and imaginary components of the Smith Chart value are displayed.
Anritsu Line Sweep Tools or Master Software Tools include additional options and a calculator that can easily show what the return loss, VSWR, or reflection coefficient values of a specific Smith Chart value are.
It is possible to change the zoom size in the Amplitude menu. Expand 10 dB zooms in the Smith Chart so that the reflection coefficient is between 0 and 0.3162. Expand 20 dB expands the Smith Chart to show rho between 0 and 0.1 and Expand 30 dB expands to show rho between 0 and 0.0316.
Smith Chart Measurement
The following example shows how a Smith Chart can be used to measure the match of an antenna.
1. Press the Measurements main menu key.
2. Press the More submenu key and select Smith Chart.
3. Press the Freq/Dist main menu key and set the start frequency and stop frequency.
4. Press Shift and Calibrate (2) to calibrate the instrument. See Calibration for details.
5. Connect the antenna to the RF Out connector on the instrument.