Spectrum Analyzer Measurement Guide : Interference Analyzer (Option 25) : Zero Span IF BW Menu (Option 89)
 
Zero Span IF BW Menu (Option 89)
Note 
Option 89 is not available on all instrument models. Refer to your instrument Technical Data Sheet.
Zero Span IF bandwidth provides a nominal 140 MHz IF signal out of a BNC female connector. The IF output signal is present only when the instrument span is set to zero by pressing the Zero Span key in the Span menu. Press the Zero Span submenu key again to access the Zero Span IF BW menu.
When Normal is selected, the IF bandwidth is influenced by the RBW setting. The selectable bandwidth values may differ among instrument types.
Key Sequence: Freq > Span > Zero Span 
Zero Span IF BW Menu
 
Normal
Press this key to select a bandwidth based on the RBW setting.
The selectable fixed bandwidth values listed below may differ depending on your instrument.
7 MHz BW
Press this button to select a fixed IF bandwidth of 7 MHz.
 
10 MHz BW
Press this button to select a fixed IF bandwidth of 10 MHz.
 
 
16 MHz BW
Press this button to select a fixed IF bandwidth of 16 MHz.
 
32 MHz BW
Press this button to select a fixed IF bandwidth of 32 MHz.
 
 
 
 
Back
Returns to the Span menu.