This subsystem contains commands related to the generator.
:GENerator:ADDitional:CALibration
Title
Perform Additional Calibration
Description
Performs an additional calibration sweep to improve the output of the source. Use [:SENSe:]:ENhanced:CALibration:COMPlete[:STATe]? to determine when calibration is complete.
Note: This command works only after the Generator has been turned ON and Enhanced Accuracy Calibration has been turned ON.
Related Command
[:SENSe]:GENerator:ENhanced:CALibration[:STATe] ON [:SENSE]:MATChing:PADS <num> [:SENSe]:ENhanced:CALibration:COMPlete[:STATe]?
Generator Output Power Offset or External Gain/Loss Setting
Description
Sets the output power level offset or the External Gain/Loss Setting. Please note that changing this value causes the displayed output power to adjust with the new offset. For example, if the output power is set to -25 dBm and the External Gain/Loss is then set to 10 dB external gain, the power limits are adjusted by 10 dBm and value of the output power displayed on the instrument are adjusted to -15 dBm. The query returns a value from -100 to 100. A negative sign indicates external loss and no sign indicates external gain.
Parameter
<value>
Default Value
0 dB
Range
100 dB Ext Loss to 100 dB Ext Gain
Default Unit
dB
Example
To set the signal generator offset to 10 dB external gain: