1. Press the Zero/Cal main menu key and press the Cal Factor submenu key. Enter the Center Frequency, or press the Signal Standard key and the Up/Down arrow keys to select a particular standard. The calibration factors are derived for the corresponding center frequency. The channel number is not required because the calibration factor frequencies are rounded to the nearest 500 MHz. The Cal Factor message in the display window shows Cal Factor ON if the Cal Factor command has been properly sent to the sensor.
2. With no power applied to the sensor, press the Zero submenu key to zero the sensor. This step is recommended when making power measurements below – 20 dBm.