S331E, S332E, S361E, S362E Site Master™MS2711E, MS2712E, MS2713E Spectrum Master™MT8212E, MT8213E Cell Master™ Programming Manual : Tracking Generator Commands : :FETCh Subsystem
 
:FETCh Subsystem
This set of commands returns the most recent measurement data of the active measurement. They will not switch to another measurement.
To make a new measurement, use the INITiate command. To get new measurement data, use the READ or MEASure query commands.
:FETCh:TGENerator:POWStats?
Title
Fetch Tracking Generator Absolute Power Statistics
Description
(Only valid for Tracking Mode) Returns the most recent tracking generator absolute power statistic measurement results displayed in the Tracking Generator Summary table.
Data is returned as 3 comma-separated values: Absolute Max Power, Absolute Min Power, and Absolute Average Power. "---" is returned for each data that is not valid at that instance.
Default Unit
Current amplitude units
Related Command
:CONFigure:GENerator:TRACking POWStats
Front Panel Access
Shift-4 (Measure), Generator, Settings, Power Statistics
:FETCh:TGENerator:TXStats?
Title
Fetch Tracking Generator Transmission Statistics
Description
(Only valid for Tracking Mode and Normalize On) Returns the most recent tracking generator transmission statistic measurement results.
Data is returned as 3 comma-separated values: Max Power, Min Power, Average Power "---" is returned for each data that is not valid at that instance.
Default Unit
Current amplitude units
Related Command
:CONFigure:GENerator:TRACking TXStats
Front Panel Access