S331E, S332E, S361E, S362E Site Master™MS2711E, MS2712E, MS2713E Spectrum Master™MT8212E, MT8213E Cell Master™ Programming Manual : Cable & Antenna Commands : :CALibration Subsystem
 
:CALibration Subsystem
This subsystem controls the system calibration.
:CALibration:STATe?
Title
Calibration State
Description
Reports the calibrated state. This command returns a 1 if the instrument has been calibrated with discrete Open, Short, and Load components and returns 0 if the instrument has not been calibrated.
Related Command
[:SENSe]CORRection:COLL:LOAD
[:SENSe]CORRection:COLL:OPEN
[:SENSe]CORRection:COLL:SHORt