BW: Opens the Select Bandwidth list. Choose from the following bandwidths:
1.4 MHz3 MHz
5 MHz
10 MHz
15 MHz (Option 543 required)
EVM Mode: Press to select either Auto or PBCH Only. Selecting Auto measures PDSCH if data is available, otherwise, it measures PBCH. Selecting PBCH Only forces a PBCH measurement and the constellation measurement will only show QPSK. The instrument will automatically choose PBCH Only mode when the OTA Scanner is used.
Sync Type: Press to choose between the Normal (SS) or RS Sync Type. Only choose RS when there is no sync signal present in the transmitter under test. If RS is chosen and a sync signal is present, no modulation measurements will be available.
Cell ID: Identifying information sent by the transmitter in the sync signal. If Normal (SS) is selected as the Sync Type, the Cell ID is extracted automatically from the Sync Signal. The Cell ID must be entered manually when RS Sync Type is selected. For user convenience, the Cell ID is automatically populated from the last measured signal with an SS.
Back: Returns to the Setup menu.
Ext. Ref. Freq (Only Some Models): Press to set the external reference frequency. The External Reference Frequency list opens. Highlight the desired frequency and press
Enter.