General Information : Cable and Antenna Analyzer : Limit Menu

Limit Menu
A limit line can be used for visual reference only, or for pass/fail criteria using the limit alarm. Limit alarm failures are reported whenever a signal is above the upper limit line or below the lower limit line.
A limit line can consist of a single segment, or as many as 40 segments across the entire frequency span of the instrument. These limit segments are retained regardless of the current frequency span of the instrument, allowing the configuring of specific limit envelope at various frequencies of interest without having to re-configure them each time the frequency is changed. To clear the current limit setup configuration and return to a single limit segment starting at the current start frequency and ending at the current stop frequency, press the Clear Limit submenu key.
Key Sequence: Shift > Limit
This key create a single segment limit line. The amplitude of the limit line is adjusted with the arrow keys, rotary knob, or the numeric keypad.
The “Limit Edit Menu” is displayed to allow the creation or editing of single or multi-segment limit lines. The currently active limit point is marked by a red circle on the display.
This submenu key selects, for the currently active limit line, if an alarm beep will occur when a data point exceeds the limit.

Copyright 2010 Anritsu Company
Anritsu Company
1-800-ANRITSU
10450-00024, Rev. B