General Information : Programming Commands : :SENSe Subsystem

:SENSe Subsystem
The commands in this subsystem relate to device-specific parameters, not signal-oriented parameters.
:[SENSe:]PIManalyzer:OUTput:POWer
:[SENSe:]PIManalyzer:OUTput:POWer?
Title
PIM Analyzer Output Power
Description
Sets the deliverable output from the PIM Analyzer to the system under test.
Parameter
0 = 40 Watts
1 = 30 Watts
2 = 20 Watts
Default Value
2 (20 Watts)
Range
0, 1 or 2
Example
To set the PIM Output Power:
:SENSe:PIManalyzer:OUTput:POWer 0
Sets the PIM output to 40 Watts.
Front Panel Access
Setup, Output Power
:[SENSe:]PIManalyzer:TESt:DURation
:[SENSe:]PIManalyzer:TESt:DURation?
Title
PIM Test Measurement Test Duration
Description
Sets the amount of time in seconds the PIM Master will be on for intermodulation distortion measurements.
Parameter
<time>
Default Value
20
Default Unit
Seconds
Range
1.0 to 60.0 seconds
Example
To set the test duration time to 5 seconds:
:SENSe:PIManalyzer:TESt:DURation 5.0
Front Panel Access
Setup, Test Duration
:[SENSe:]PIManalyzer:FREQuency:F1|F2
:[SENSe:]PIManalyzer:FREQuency:F1|F2?
Title
PIM Master Frequency Setup
Description
Sets the PIM Master LO frequencies, calculates the IMx Order frequency and sets the Anritsu handheld instrument for the corresponding IMx frequency.
Parameter
1930 to 1990 MHz and 2110 to 2155 MHz frequency ranges are valid ranges. Frequency input in Hz.
Default Unit:
Hz
Range:
PIM Master Model MW8219A
Valid ranges are 1930 to 1990 MHz and 2110 to 2155 MHz.
Example
To set the PIM Test frequency F2 for 1990 MHz:
:SENSe:PIManalyzer:FREQuency:F2 1990000000
Front Panel Access
Freq, Carrier F1|Carrier F2
:[SENSe:]PIManalyzer:AUTorange OFF|ON|0|1
:[SENSe:]PIManalyzer:AUTorange?
Title
Auto Range Setup
Description
Sets the Anritsu handheld instrument to auto range the amplitude of the PIM signal received from the PIM Master. There is an initial 1 second measurement of the PIM power so the Anritsu handheld instrument can adjust reference level and span appropriately. This is followed by the actual measurement at the pre described test duration. The query part of the command returns the value of the current Auto Range setup.
Default Value
0
Range
0|Off, Auto Ranging is OFF
1|ON, Auto Ranging is ON
Example
To set the Anritsu handheld instrument for Auto range:
:SENSe:PIManalyzer:AUTorange 1
Front Panel Access
Amplitude, Auto Range
:[SENSe:]PIManalyzer:MEASure:VALue?
Title
Measurement Receiver Measured Value From PIM Master Measurement Setup
Description
The query returns an inter modulation measured value in both dBc and dBm.
Parameter:
<amplitude>
Default Unit
dBc/dBm
Front Panel Access
N/A. Inter modulation distortion value is displayed in the lower measurement box as “PIM =”.
:[SENSe:]PIManalyzer:MEASure:STATus?
Title
Measurement Receiver/ PIM Master Current Measurement Status
Description
The query returns the current measurement status of the measurement receiver/PIM Analyzer measurement system. Since the measurement system can run for 60 seconds, the query serves to indicate if the measurement is still in process.
Range
0, Measurement is OFF
1, Measurement is ON
:[SENSe:]PIManalyzer:IMD:ORDer
:[SENSe:]PIManalyzer:IMD:ORDer?
Title
Measurement Receiver / Pim Master Inter-Modulation Distortion (IMD) Order Setup
Description
Sets the measurement receiver center frequency to receive one of the following IMDs from the PIM Master measurement system: 3/5/7. The query command returns the possible strings “3rd”, “5th” and “7th” depending on the current selection of IMD Order.
Default Value
3
Range: 3, 5 and 7 are the only acceptable values.
Example
To set the measurement receiver for the 5th order IMD:
:SENSe:PIManalyzer:IMD:ORDer 5
Sets the measurement receiver center frequency to the 5th order IMD.
Front Panel Access
Freq, IMD Order

Copyright 2010 Anritsu Company