The Anritsu handheld instrument can not lock onto the reference signal from the PIM Master and switches to its own internal frequency reference.PIM front panel EMERGENCY STOP button has been pressed. This will turn off the PIM RF and cancel the measurement on the HHSPA side. Measurement results may be invalid.The entire PIM Master unit is beyond the normal temperature range and degradation in measurement or signal dropout can occur.The amplifier power supply is out of range and degradation in measurement or signal dropout can occur.If the PIM Master has exceeded an On Time greater than allowed (~68 seconds), then the unit will automatically turn RF Off. Possible communication failure between Anritsu handheld instrument and PIM unit. Disregard current measurement results and power cycle both units.