Spectrum Analyzer : Measure Menu : Channel Power Menu

Channel Power Menu
Key Sequence: Shift > Measure (4) key > Channel Power
Begins or ends the channel power measurement. When the measurement is on, Ch Pwr will appear below the display. The detection method will automatically be changed to RMS Average when the measurement is started. The detection method can be modified by pressing the Shift and the Sweep keys and pressing the Detection submenu key.
Activates the center frequency function, and sets the center frequency of the instrument for the channel power measurement. Use the keypad, the directional arrow keys, or the rotary knob to enter the center frequency. The Up/Down arrows change the frequency by the frequency step size entered in the “Freq (Frequency) Menu”. The Left/Right arrows change the frequency by 10% of the span.
Sets the width for the channel power. Use the keypad, the directional arrow keys, or the rotary knob to enter the channel power width. The Up/Down arrow keys change the Channel Power Width by the frequency step value. The Left/Right arrow keys change the value by 10% of the span.
Sets the span for channel power measurement. Use the keypad, the directional arrow keys, or the rotary knob to enter the span.

Copyright 2010 Anritsu Company
Anritsu Company
1-800-ANRITSU
10450-00025, Rev. A