LMR Masterâ„¢ S412E Programming Manual : P25 Phase 1 Commands : :CONFigure Subsystem
 
:CONFigure Subsystem
This set of commands prepares the instrument for the selected measurement. It disables any currently-enabled measurements and activates the specified measurement.
Current instrument settings may be changed to default values. These changes are identified with their respective measurement commands.
Note 
Sending a non-query :CONFigure command changes the Sweep setting from Run to Hold.
:CONFigure?
Description
:CONFigure? query returns the name of the measurement previously set up using a CONFigure command or a MEASure? query. The list below shows the possible return values and the actual names of each configuration.
Returns Value
Actual Name
SIGA
P25 Analyzer
COV
P25 Coverage
CONT
P25 Control
BITC
P25 Bit Capture
Cmd Parameters:
NA
Query Parameters:
NA
Range
NA
Default Value
NA
Default Unit
NA
Example
To query the current measurement type:
:CONFigure?
Front Panel Access
Measurement
:CONFigure:BITCap
Description
This command configures the P25 Bit Capture measurement. Certain settings from the previous measurement (Ref Lvl, Scale, Graph Types, Maximize) are backed up when exiting this measurement and restored when entering this measurement. All other settings carry over from one measurement to the other. Please note that you must have the Rx pattern set to VOICe to set the P25 Bit Capture measurement.
Cmd Parameters:
NA
Query Parameters:
NA
Range
NA
Default Value
NA
Default Unit
NA
Example
To switch the measurement to P25 Bit Capture:
:CONFigure:BITCap
Related Command
:DM:PATTern VOICe
Front Panel Access
Measurement, P25 Bit Capture
:CONFigure:CONTrol
Description
This command configures the P25 Control measurement. Certain settings from the previous measurement (Ref Lvl, Scale, Graph Types, Maximize) are backed up when exiting this measurement and restored when entering this measurement. All other settings carry over from one measurement to the other. Please note that you must have the Rx pattern set to VOICe or CTRLchan to set the P25 Control measurement.
Cmd Parameters:
NA
Query Parameters:
NA
Range
NA
Default Value
NA
Default Unit
NA
Example
To switch the measurement to P25 Control:
:CONFigure:CONTrol
Related Command
:DM:PATTern VOICe
:DM:PATTern CTRLchan
Front Panel Access
Measurement, P25 Control
:CONFigure:COVerage
Description
This command configures the P25 Coverage measurement. Certain settings from the previous measurement (Mapping Type) are backed up when exiting this measurement and restored when entering this measurement. All other settings carry over from one measurement to the other.
Cmd Parameters:
NA
Query Parameters:
NA
Range
NA
Default Value
NA
Default Unit
NA
Example
To switch the measurement to P25 Coverage:
:CONFigure:COVerage
Front Panel Access
Measurement, P25 Coverage
:CONFigure:SIGAnalyzer
Description
This command configures the P25 Analyzer measurement. Certain settings from the previous measurement (Ref Lvl, Scale, Graph Types, Maximize) are backed up when exiting this measurement and restored when entering this measurement. All other settings carry over from one measurement to the other.
Cmd Parameters:
NA
Query Parameters:
NA
Range
NA
Default Value
NA
Default Unit
NA
Example
To switch the measurement to P25 Analyzer:
:CONFigure:SIGAnalyzer
Front Panel Access
Measurement, P25 Analyzer