LMR Masterâ„¢ S412E Programming Manual : NBFM Commands : :UNIT Subsystem
 
:UNIT Subsystem
The unit subsystem is used to modify the default units used for related parameters. These changes affect parameters in both commands and responses.
:UNIT:POWer:RX DBM|WATTs|VOLT
:UNIT:POWer:RX?
Description
Sets the receiver unit to dBm or Watts or Volts. If the unit is set to dBm, the NBFM Analyzer received power (from FETCh:SIGAnalyzer? or READ:SIGAnalyzer? or MEASURE:SIGAnalyzer?) and the squelch setting will be set and queried in dBm.If the unit is set to Watt, the NBFM Signal Analyzer received power and squelch setting will be set and queried in fW. If the unit is set to Volts, the NBFM Signal Analyzer received power and squelch setting will be set and queried in fV.
Cmd Parameters
DBM|WATTs|VOLT
Query Parameters
NA
Range
DBM|WATT|VOLT
Default Value
DBM
Default Unit
NA
Example
To set the receiver units to watts:
:UNIT:POWer:RX WATT
Front Panel Access
Amplitude, Units, Rx Units
:UNIT:POWer:TX DBM|WATTs|VOLT|EDBUv
:UNIT:POWer:TX?
Description
Sets the generator unit to dBm or Volts. If the unit is set to dBm, the Tx Output Lvl setting will be set and queried in dBm. If the unit is set to Watts, the Tx Output Lvl setting will be set and queried in fW. If the unit is set to Volts, the Tx Output Lvl will be set and queried in fV. If the unit is set to dBuV(EMF), the Tx Output Lvl will be set and queried in EDBU.
Cmd Parameters
DBM|WATTs|VOLT|EDBUv
Query Parameters
NA
Range
DBM|WATT|VOLT|EDBU
Default Value
DBM
Default Unit
NA
Example
To set the generator units to Volts:
:UNIT:POWer:TX VOLT
Front Panel Access
Amplitude, Units, Tx Units