LMR Master™ S412E Programming Manual : All Mode Commands : :SYSTem Subsystem
 
:SYSTem Subsystem
This subsystem contains commands that affect instrument functionality. This functionality does not directly relate to data collection, display, or transfer.
:SYSTem:MSIS[:STATe]? INTernal|USB
Title
Query Memory State
Description
Queries the ready state of the instrument’s internal memory or of the USB Flash drive. Use this command to check the ready state of the memory device before sending a command, such as :MMEMory:STORe or :MMEMory:MSIS:COPY, that requires the memory location to be available.
The USB query returns a 1 when a USB device is plugged into the USB port. It returns 0 if no USB drive is present.
Internal memory in the LMR Master S412E should always be available and the INT query should always return a 1.
Parameter
INTernal|USB
Parameter Type
<char>
Related Command
:MMEMory:MSIS
:MMEMory:MSIS:DESTination
:MMEMory:MSIS:COPY
:SYSTem:OPTions?
Title
Query Installed Options
Description
Returns a string of the installed options. Options are separated by a “/”. The string returns NONE if no options are installed.
Cmd Parameter
NA (query only)
Query Response
Options are separated by “/”.
Related Command
*IDN?
:SYSTem:PRESet
Title
Preset
Description
This command restores all application parameters to their factory preset values. This command does not modify system parameters such as Ethernet configuration, language, volume, or brightness.
Note: After issuing this command, wait 30 seconds for the instrument to complete the preset condition before sending another command.
Syntax
:SYSTem:PRESet
Cmd Parameter
NA
Query Response
NA (no query)
Related Command
*RST
Front Panel Access
Shift‑1 (Preset), Preset