LMR Master™ S412E Programming Manual : Vector Voltmeter Commands : :FETCh Subsystem
 
:FETCh Subsystem
This subsystem contains commands to fetch the VVM data and relative data.
:FETCh:VVM:DATA?
Description
Query only. Returns the most recent VVM measurement results. Data are returned as two or four comma‑separated values depending upon the measurement type, measurement format, measurement mode, port, and the reference setting. A “” is returned for any data that are not valid at that instance.
VVM Measurement Results  
If
Then
Data Values
If the measurement type is Insertion, and if the measurement mode is CW,
then data are returned as four comma‑separated values
Amplitude
Phase
Reference Amplitude
Reference Phase.
If the measurement type is Insertion, and if the measurement mode is CW with save new reference set,
then data are returned as four comma‑separated values
Relative Amplitude
Relative Phase
Reference Amplitude
Reference Phase.
If the measurement type is Return, and if the measurement mode is CW, and if format is set to dB,
then data are returned as four comma‑delimited values
Amplitude
Phase
Reference Amplitude
Reference Phase.
If the measurement type is Return, and if the measurement mode is CW with save new reference set, and if format is set to dB,
then data are returned as four comma‑delimited values
Relative Amplitude
Relative Phase
Reference Amplitude
Reference Phase.
If the measurement type is Return, and if format is set to VSWR,
then data are returned as two comma‑delimited values
VSWR
Reference VSWR.
If the measurement type is Return, and if the measurement mode is CW with save new reference set, and if format is set to VSWR,
then data are returned as two comma‑delimited values
Relative VSWR
Reference VSWR.
If the measurement type is Return, and if the measurement mode is CW, and if format is set to Impedance,
then data are returned as four comma‑delimited values
Real
Imaginary
Reference Real
Reference Imaginary.
If the measurement type is Return, and if the measurement mode is CW with save new reference set, and if format is set to Impedance,
then data are returned as four comma‑delimited values
Relative Real
Relative Imaginary
Reference Real
Reference Imaginary.
If the measurement mode is Table with save new reference set,
then data are returned as four comma‑separated values
Amplitude
Phase
Relative Amplitude
Relative Phase.
Cmd Parameter
NA (query only)
Query Response
NA (comma separated values)
Example
To fetch the VVM data:
:FETCh:VVM:DATA?
Front Panel Access
NA
:FETCh:VVM:REFerence:DATA?
Description
Returns the reference data depending upon the measurement type,
the measurement format, and the current port.
Syntax
:FETCh:VVM:REFerence:DATA?
Cmd Parameter
NA (query only)
Query Response
NA (comma separated values)
Example
To fetch the VVM reference data:
:FETCh:VVM:REFerence:DATA?
Front Panel Access
NA