This subsystem contains commands that affect instrument functionality that does not directly relate to data collection, display or transfer.
:SYSTem:MSIS[:STATe]? INTernal|USB
Title
Query Memory State
Description
Queries the ready state of the instrument’s internal memory or of the USB Flash drive. Use this command to check the ready state of the memory device before sending a command, such as :MMEMory:STORe or :MMEMory:MSIS:COPY, that requires the memory location to be available.
The USB query returns a 1 when a USB device is plugged into the USB port. It returns 0 if no USB drive is present.
Internal memory should always be available and the INT query should always return a 1.
Returns a string of the installed options. Options are separated by a “/”. The string will return “NONE” if no options are installed.
Related Command
*IDN?
:SYSTem:PRESet
Title
Preset
Description
This command restores all application parameters to their factory preset values. This does not modify system parameters such as language, volume or brightness.