S331E, S332E, S361E, S362E Site Master™MS2711E, MS2712E, MS2713E Spectrum Master™MT8212E, MT8213E Cell Master™ Programming Manual : GSM Commands : :CONFigure Subsystem
 
:CONFigure Subsystem
This set of commands prepares the instrument for the selected measurement. It disables any currently-enabled measurements and activates the specified measurement. It sets the instrument to single sweep mode, waiting for an :INITiate command. It will not initiate the taking of a measurement.
Current instrument settings may be changed to default values. These changes are identified with their respective measurement commands.
:CONFigure:SUMMary
Title
Configure Summary Measurement
Description
This command configures the summary of all the related numerical measurement results. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
Front Panel Access
Shift‑4 (Measure), GSM/EDGE Summary
:CONFigure:DEMod
Title
Configure Demodulation Measurement
Required Option
41
Description
This command configures the Demodulator measurement. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
Front Panel Access
Shift‑4 (Measure), Demodulator
:CONFigure:PFail
Title
Configure PASS/FAIL Measurement
Description
This command configures the PASS/FAIL measurement. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement. The measurement results are equivalent to the results that are displayed by choosing Measurements, then Pass/Fail mode from the front panel.
Related Command
:FETCh:PFail?
MEASure:PFail?
READ:PFail?
Front Panel Access
Shift‑4 (Measure), Pass/Fail Mode
:CONFigure:PVTFrame
Title
Configure Power vs. Time (Frame) Measurement
Required Option
40
Description
This command configures the Power vs. Time (Frame) measurement. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
Related Command
MEASure:PVTFrame?
READ:PVTFrame?
FETCh:PVTFrame?
Front Panel Access
Shift‑4 (Measure), Power vs. Time (Frame)
:CONFigure:PVTSlot
Title
Configure Power vs. Time (Slot) Measurement
Required Option
40
Description
This command configures the Power vs. Time (Slot) measurement. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
Related Command
MEASure:PVTSlot?
READ:PVTSlot?
FETCh:PVTSlot?
Front Panel Access
Shift‑4 (Measure), Power vs. Time (Slot)
:CONFigure:SPECtrum:MULTi
Title
Configure Multi-Channel Spectrum Measurement
Required Option
40
Description
This command configures the Multi-Channel Spectrum measurement. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe]:RF commands before initiating a measurement.
Front Panel Access
Shift‑4 (Measure), Spectrum, Multi-Channel Spectrum
:CONFigure:SPECtrum:SINGle
Title
Configure Single Channel Spectrum Measurement
Required Option
40
Description
This command configures the Single Channel Spectrum measurement. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe]:RF commands before initiating a measurement.
Related Command
MEASure:SPECtrum:SINGle?
READ:SPECtrum:SINGle?
FETCh:SPECtrum:SINGle?
Front Panel Access
Shift‑4 (Measure), Spectrum, Channel Spectrum
:CONFigure?
Title
Configure Query
Description
:CONFigure? query returns the name of the measurement previously set up using a CONFigure command or a MEASure? query. The list below shows the possible return values and the actual names of each configuration.
Returns Value Actual Name
SPEC, SING Single Channel Spectrum
SPEC, MULT Multi-Channel Spectrum
PVTF Power vs. Time (Frame)
PVTS Power vs. Time (Slot)
DEM Demodulator
PF Pass/Fail Mode
SUMM Summary