S331E, S332E, S361E, S362E Site Master™MS2711E, MS2712E, MS2713E Spectrum Master™MT8212E, MT8213E Cell Master™ Programming Manual : WCDMA Commands : :MEASure Subsystem
 
:MEASure Subsystem
These commands take the instrument from its current state, enable the specified measurement and put the instrument into single sweep mode. They correct any parameters that are invalid given the new measurement state such that a valid measurement can take place. Other settings may be changed; see the documentation of CONFigure for each measurement. They then initiate the measurement. When the measurement is complete, they return the result.
To make a measurement with settings other than the “default” measurement settings applied by CONFigure, do the following:
Send the appropriate CONFigure command to set the desired measurement.
Modify the settings as required.
Send the appropriate READ command to measure and return the result.
To get the current measurement data, use the appropriate FETCh command.
:MEASure:ACLR?
Title
Measure Adjacent Channel Leakage Ratio
Description
Sets the active measurement to adjacent channel power ratio, sets the default measurement parameters, triggers a new measurement and returns the main channel(s) power, and adjacent channel(s) power. It is a combination of the commands :CONFigure:RF ACLR and :READ:ACLR? For a description of the default adjacent channel power ratio measurement parameters, see :CONFigure:RF ACLR. To make an adjacent channel power ratio measurement with settings other than the default values, send:
:CONFigure:RF ACLR
Commands to set desired settings
:READ:ACLR?
Data is returned as 12 comma-separated values: -20 MHz channel power, -15 MHz channel power, -10 MHz channel power, -5 MHz channel power, Main channel power 1, Main channel power 2, Main channel power 3, Main channel power 4, +5 MHz channel power, +10 MHz channel power, +15 MHz channel power, +20 MHz channel power.
Default Unit
dBm
:MEASure:CHPower?
Title
Measure Channel Power
Description
Sets the active measurement to Code Domain Power (CDP), sets the default measurement parameters, triggers a new measurement and returns the channel power results. It is a combination of the commands :CONFigure:DEMod CDPower and :READ:CHPower? For a description of the default channel power measurement parameters, see :CONFigure:DEMod CDPower. To make a channel power measurement with settings other than the default values, send:
:CONFigure:DEMOd CDPower
Commands to set desired settings
:READ:CHPower?
Default Unit
dBm
Related Command
:CONFigure:DEMod CDPower
:MEASure:DEMod:ACTChannels?
Title
Measure Active Code Domain Channels
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the number of active traffic channels. It is a combination of the commands :CONFigure:DEMod CDP and :READ:DEMod:ACTChannels? For a description of the default code domain power measurement parameters, see :CONFigure:DEMod CDP.
Related Command
:CONFigure:RF ACLR
:CONFigure:DEMod:CDP
:MEASure:DEMod:CDPower?
Title
Measure Code Domain Power
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the channel power results. It is a combination of the commands :CONFigure:DEMod CDPower and :READ:DEMod:CDPower? For a description of the default channel power measurement parameters, see :CONFigure:DEMod CDPower. To make a code domain power measurement with settings other than the default values, send:
:CONFigure:DEMod CDP
Commands to set desired settings
:READ:DEMod:CDPower?
Results are returned as 4 comma-delimited values: CPICH power, PCCPCH power, PICH power, SCCPCH power.
Default Unit
dBm
Related Command
:CONFigure:DEMod CDPower
:READ:DEMod:CDPower?
FETCh:DEMod:CDPower?
:MEASure:DEMod:CFTHrough?
Title
Measure Carrier Feed Through
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the carrier feed through results. It is a combination of the commands :CONFigure:DEMod CDP and :READ:DEMod:CFTHrough? For a description of the default code domain power measurement parameters, see :CONFigure:DEMod CDP. To make a carrier feed through measurement with settings other than the default values, send:
:CONFigure:DEMod CDP
Commands to set desired settings
:READ:DEMod:CFTHrough?
Default Unit
dB
Related Command
:CONFigure:DEMod CDP
:MEASure:DEMod:EVM?
Title
Measure Error Vector Magnitude
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the EVM results. It is a combination of the commands :CONFigure:DEMod CDP and :READ:DEMod:EVM? For a description of the default code domain power measurement parameters, see :CONFigure:DEMod CDP. To make a carrier feed through measurement with settings other than the default values, send:
:CONFigure:DEMod CDP
Commands to set desired settings
:READ:DEMod:EVM?
Default Unit
%
Related Command
:CONFigure:DEMod CDP
:MEASure:DEMod:FERRor:PPM?
Title
Measure Frequency Error PPM
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the frequency error PPM results. It is a combination of the commands :CONFigure:DEMod CDP and :READ:DEMod:FERRor:PPM? For a description of the default code domain power measurement parameters, see :CONFigure:DEMod CDP. To make a frequency error measurement with settings other than the default values, send:
:CONFigure:DEMod CDP
Commands to set desired settings
:READ:DEMod:FERRor:PPM?
Related Command
:READ:DEMod:FERRor:PPM?
:FETCh:DEMod:FERRor:PPM?
:CONFigure:DEMod:CDP
:MEASure:DEMod:FERRor:PRECise?
Title
Measure Frequency Error
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the frequency error results. It is a combination of the commands :CONFigure:DEMod CDP and :READ:DEMod:FERRor:PRECise? For a description of the default code domain power measurement parameters, see :CONFigure:DEMod CDP. To make a frequency error measurement with settings other than the default values, send:
:CONFigure:DEMod CDP
Commands to set desired settings
:READ:DEMod:FERRor:PRECise?
Note that this command is the same as :MEASure:DEMod:FERRor?, but returns the result with a precision of one-tenth Hertz.
Default Unit
Hz
Related Command
:CONFigure:DEMod CDP
:MEASure:DEMod:FERRor?
:MEASure:DEMod:FERRor?
Title
Measure Frequency Error
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the frequency error results. It is a combination of the commands :CONFigure:DEMod CDP and :READ:DEMod:FERRor? For a description of the default code domain power measurement parameters, see :CONFigure:DEMod CDP. To make a frequency error measurement with settings other than the default values, send:
:CONFigure:DEMod CDP
Commands to set desired settings
:READ:DEMod:FERRor?
Default Unit
Hz
Related Command
:CONFigure:DEMod CDP
:MEASure:DEMod:HSDPa:FERRor:PPM?
Title
Measure HSDPA Frequency Error PPM
Description
Sets the active measurement to HSDPA sets the default measurement parameters, triggers a new measurement and returns the HSDPA frequency error PPM results. It is a combination of the commands :CONFigure:DEMod HSDPa and :READ:DEMod:HSDPa:FERRor:PPM? For a description of the default HSDPA measurement parameters, see :CONFigure:DEMod HSDPa. To make a frequency error measurement with settings other than the default values, send:
:CONFigure:DEMod HSDPa
Commands to set desired settings
:READ:DEMod:HSDPa:FERRor:PPM?
Related Command
:READ:DEMod:HSDPa:FERRor:PPM?
:FETCh:DEMod:HSDPa:FERRor:PPM?
:CONFigure:DEMod HSDPa
:MEASure:DEMod:HSDPa:FERRor?
Title
Measure HSDPA Frequency Error
Description
Sets the active measurement to HSDPA, sets the default measurement parameters, triggers a new measurement and returns the HSDPA frequency error results. It is a combination of the commands :CONFigure:DEMod HSDPa and :READ:DEMod:HSDPA:FERRor? For a description of the default HSDPA measurement parameters, see :CONFigure:DEMod HSDPa. To make a frequency error measurement with settings other than the default values, send:
:CONFigure:DEMod HSDPa
Commands to set desired settings
:READ:DEMod:HSDPa:FERRor?
Default Unit
Hz
Related Command
:READ:DEMod:HSDPa:FERRor?
:FETCh:DEMod:HSDPa:FERRor?
:CONFigure:DEMod HSDPa
:MEASure:DEMod:HSDPa:TXFRequency?
Title
Measure HSDPA Transmitter Frequency
Description
Sets the active measurement to HSDPA, sets the default measurement parameters, triggers a new measurement and returns the HSDPA transmitter frequency results. It is a combination of the commands :CONFigure:DEMod HSDPa and :READ:DEMod:HSDPa:TXFRequency? For a description of the default HSDPA measurement parameters, see :CONFigure:DEMod HSDPa. To make a transmitter frequency measurement with settings other than the default values, send:
:CONFigure:DEMod HSDPa
Commands to set desired settings
:READ:DEMod:HSDPa:TXFRequency?
Default Unit
Hz
Related Command
:READ:DEMod:HSDPa:TXFRequency?
:FETCh:DEMod:HSDPa:TXFRequency?
:CONFigure:DEMod:HSDPa
:MEASure:DEMod:NFLoor?
Title
Measure Code Domain Noise Floor
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the noise floor results. It is a combination of the commands :CONFigure:DEMod CDP and :READ:DEMod:NFLoor? For a description of the default code domain power measurement parameters, see :CONFigure:DEMod CDP. To make a noise floor measurement with settings other than the default values, send:
:CONFigure:DEMod HSDPa
Commands to set desired settings
:READ:DEMod:NFLoor?
Default Unit
dB
Related Command
:CONFigure:DEMod CDP
:MEASure:DEMod:NUMChannels?
Title
Measure Code Domain Channel Count
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the number of channels. It is a combination of the commands :CONFigure:DEMod CDP and :READ:DEMod:NUMChannels? For a description of the default code domain power measurement parameters, see :CONFigure:DEMod CDP. To make a channel count measurement with settings other than the default values, send:
:CONFigure:DEMod HSDPa
Commands to set desired settings
:READ:DEMod:NUMChannels?
Default Unit
dB
Related Command
:CONFigure:DEMod CDP
:MEASure:DEMod:SPOWer?
Title
Measure Synch Channel Power
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the synch channel power results. It is a combination of the commands :CONFigure:DEMod CDP and :READ:DEMod:SPOWer? For a description of the default code domain power measurement parameters, see :CONFigure:DEMod CDP. To make a synch channel power measurement with settings other than the default values, send:
:CONFigure:DEMod CDP
Commands to set desired settings
:READ:DEMod:SPOWer?
Results are returned as 3 comma-delimited values: SCH power, PSCH power, SSCH power.
Default Unit
dBm
Related Command
:CONFigure:DEMod CDP
:MEASure:DEMod:TXFRequency:PRECise?
Title
Measure Transmitter Frequency
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the transmitter frequency results. It is a combination of the commands :CONFigure:DEMod CDP and :READ:DEMod:TXFRequency:PRECise? For a description of the default code domain power measurement parameters, see :CONFigure:DEMod CDP. To make a transmitter frequency measurement with settings other than the default values, send:
:CONFigure:DEMod CDP
Commands to set desired settings
:READ:DEMod:TXFRequency:PRECise?
Note that this command is the same as :MEASure:DEMod:TXFRequency?, but returns the result with a precision of one-tenth Hertz.
Default Unit
Hertz
Related Command
:CONFigure:DEMod CDP
:MEASure:DEMod:TXFRequency?
:MEASure:DEMod:TXFRequency?
Title
Measure Transmitter Frequency
Description
Sets the active measurement to code domain power, sets the default measurement parameters, triggers a new measurement and returns the transmitter frequency results. It is a combination of the commands :CONFigure:DEMod CDP and :READ:DEMod:TXFRequency? For a description of the default code domain power measurement parameters, see :CONFigure:DEMod CDP. To make a transmitter frequency measurement with settings other than the default values, send:
:CONFigure:DEMod CDP
Commands to set desired settings
:READ:DEMod:TXFRequency?
Default Unit
Hertz
Related Command
:CONFigure:DEMod CDP
:MEASure:EMISsion:FAIL?
Title
Measure Spectral Emission Mask Pass/Fail status
Description
Sets the active measurement to spectral emission mask, sets the default measurement parameters, triggers a new measurement and returns the spectral emission mask pass/fail results. It is a combination of the commands :CONFigure:RF EMISsion and :READ:EMISsion:FAIL? For a description of the default spectral emission mask measurement parameters, see :CONFigure:RF EMISsion. To make a spectral emission mask pass/fail status measurement with settings other than the default values, send:
:CONFigure:RF EMISsion
Commands to set desired settings
:READ:EMISsion:FAIL?
Related Command
:MEASure:EMISsion
:MEASure:EMISsion:TEMPlate?
:CONFigure:RF EMISsion
:MEASure:EMISsion:TEMPlate:DATA?
Title
Measure Spectral Emission Mask Template Data
Description
Sets the active measurement to spectral emission mask, sets the default measurement parameters, triggers a new measurement and returns the spectral emission mask template power level, frequency, and PASS/FAIL data for all ten frequency ranges shown in the Spectral Emission Summary. It is a combination of the commands :CONFigure:RF EMISsion and :READ:EMISsion:TEMPlate:DATA? For a description of the default spectral emission mask measurement parameters, see :CONFigure:RF EMISsion. To make a spectral emission mask template measurement with settings other than the default values, send:
:CONFigure:RF EMISsion
Commands to set desired settings
:READ:EMISsion:TEMPlate:DATA?
For each frequency range, the data is returned as 3 comma delimited values: Power Level in dBm, Frequency in Hz, and PASS/FAIL.
Related Command
:READ:EMISsion:TEMPlate:DATA?
:FETCh:EMISsion:TEMPlate:DATA?
:CONFigure:RF EMISsion
:MEASure:EMISsion:TEMPlate?
Title
Measure Spectral Emission Mask Template
Description
Sets the active measurement to spectral emission mask, sets the default measurement parameters, triggers a new measurement and returns the spectral emission mask template results. It is a combination of the commands :CONFigure:RF EMISsion and :READ:EMISsion:TEMPlate? For a description of the default spectral emission mask measurement parameters, see :CONFigure:RF EMISsion. To make a spectral emission mask template measurement with settings other than the default values, send:
:CONFigure:RF EMISsion
Commands to set desired settings
:READ:EMISsion:TEMPlate?
Related Command
:MEASure:EMISsion?
:MEASure:EMISsion:FAIL?
:CONFigure:RF EMISsion
:MEASure:EMISsion?
Title
Measure Spectral Emission Mask
Description
Sets the active measurement to spectral emission mask, sets the default measurement parameters, triggers a new measurement and returns the spectral emission mask measurement results. It is a combination of the commands :CONFigure:RF EMISsion and :READ:EMISsion? For a description of the default spectral emission mask measurement parameters, see :CONFigure:RF EMISsion. To make a spectral emission mask measurement with settings other than the default values, send:
:CONFigure:RF EMISsion
Commands to set desired settings
:READ:EMISsion?
Results are returned as 2 comma-delimited values: pass/fail status, and template number.
Related Command
:MEASure:EMISsion:FAIL?
:MEASure:EMISsion:TEMPlate?
:MEASure:OBWidth?
Title
Measure Occupied Bandwidth
Description
Sets the active measurement to RF spectrum, sets the default measurement parameters, triggers a new measurement and returns the occupied bandwidth results. It is a combination of the commands :CONFigure:RF SPECtrum and :READ:OBWidth? For a description of the default occupied bandwidth measurement parameters, see :CONFigure:RF SPECtrum. To make an occupied bandwidth measurement with settings other than the default values, send:
:CONFigure:RF SPECtrum
Commands to set desired settings
:READ:OBWidth?
Default Unit
Hz
Related Command
:CONFigure:OBWidth
:CONFigure:RF SPECtrum
:MEASure:OTA{1|2|3|4|5|6}?
Title
Measure Over The Air
Description
Sets the active measurement to over-the-air, sets the default measurement parameters, triggers a new measurement and returns the OTA measurement results. It is a combination of the commands :CONFigure:OTA and :READ:OBWidth? For a description of the default occupied bandwidth measurement parameters, see :CONFigure:RF SPECTRUM. To make an OTA measurement with settings other than the default values, send:
:CONFigure:OTA
Commands to set desired settings
:READ:OTA?
The measurement results are returned as 5 comma delimited values: scrambling code, CPICH power, chip energy, Ec/I0 ratio, pilot dominance.
Default Unit
Scrambling Code: No units
CPICH power: dBm
Chip Energy: dBm
EC/I0: dB
Pilot dominance: No units
Related Command
:CONFigure:OTA
:MEASure:PFail?
Title
Measure PASS/FAIL
Description
Sets the active measurement to PASS/FAIL, sets the default measurement parameters, triggers a new measurement and returns the PASS/FAIL results. It is a combination of the commands :CONFigure PFail and :READ:PFail? To make a PASS/FAIL measurement with settings other than the default values, send:
:CONFigure PFail
Commands to set desired settings
:READ:PFail?
Related Command
:CONFigure PFail
READ:PFail?
FETCh:PFail?
Front Panel Access
Measurements, Pass/Fail Mode
:MEASure:PTAPower?
Title
Measure Peak to Average Power
Description
Sets the active measurement to RF spectrum, sets the default measurement parameters, triggers a new measurement and returns the PTAP measurement results. It is a combination of the commands :CONFigure:RF SPECtrum and :READ:PTAPower? For a description of the default occupied bandwidth measurement parameters, see :CONFigure:RF SPECtrum. To make a peak to average power measurement with settings other than the default values, send:
:CONFigure:RF SPECtrum
Commands to set desired settings
:READ:PTAPower?
Default Unit
dB
Related Command
:CONFigure:RF SPECtrum