VNA Master™Programming Manual : VNA Commands : :CALCulate<Tr>:FILTer[:GATE] Subsystem
 
:CALCulate<Tr>:FILTer[:GATE] Subsystem
This subsystem includes commands that allow you to set up the gate configuration.
Gate Coupling State
:CALCulate:FILTer[:GATE]:COUPled[:STATe]
Description
Sets the gate coupling state. Setting the value to ON or to 1 turns on the gate coupling, which implies that the gate settings for all traces are identical. Setting the value to OFF or to 0 turns off the gate coupling, which implies that each trace can have different gate settings. The query version of this command returns 1 if gate coupling is currently on, otherwise returns 0 for off.
Syntax
:CALCulate:FILTer[:GATE]:COUPled[:STATe] ON|OFF|1|0
:CALCulate:FILTer[:GATE]:COUPled[:STATe]?
Cmd Parameter
<boolean> ON|OFF|1|0
Query Response
<bNR1> 1|0
Default Value
1
Example
To set gate coupling to off:
:CALC:FILT:COUP OFF
:CALCulate:FILTer:GATE:COUPled:STATe 0
Front Panel Access
Shift‑8 (System), Application Options, Time Domain, Gate Coupled