VNA Master™Programming Manual : VNA Commands : [:SENSe]:CORRection:CKIT:USER Subsystem
 
[:SENSe]:CORRection:CKIT:USER Subsystem
This subsystem contains commands to configure the user device under test (DUT).
DUT User Inductance Coefficient value
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:L0|1|2|3
Description
Sets the DUT inductance value for the specified Short component of the user-defined SOLT Cal Kit.
Syntax
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:
L0|1|2|3 <inductance>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:
L0|1|2|3?
Cmd Parameter
<NRf> <inductance> (e‑12, e‑24, e‑34, e‑42)
Query Response
<NR3> <inductance> (e‑12, e‑24, e‑34, e‑42)
Example
To set the DUT inductance L1 for User 1 with line type coax to 5:
:SENS:CORR:CKIT:USER:COAX:SOLT:L1 5
To query the DUT inductance L1 for User 1:
:SENS:CORR:CKIT:USER:COAX:SOLT:L1?
The query response would be: 5.000
Front Panel Access
Shift-2 (Calibrate), DUT Port Setup, Setup User‑Defined, Short
 
DUT User Capacitance Coefficient value
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:C0|1|2|3
Description
Sets the DUT capacitance value for the specified user.
Syntax
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:C0|1|2|3 <capacitance>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:C0|1|2|3?
Cmd Parameter
<NRf> <capacitance> (e‑15, e‑27, e‑36, e‑45)
Query Response
<NR3> <capacitance> (e‑15, e‑27, e‑36, e‑45)
Example
To set the DUT capacitance #1 for User 1 with line type coax to 5:
:SENS:CORR:CKIT:USER:COAX:SOLT:C1 5
Front Panel Access
Shift-2 (Calibrate), DUT Port Setup, Setup User‑Defined, Open
 
DUT User Cutoff Frequency
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:WGUide:SSLT|SSST
:FCUToff
Description
Sets the DUT cutoff frequency for the specified user.
Syntax
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:WGUide:SSLT|SSST
:FCUToff <freq>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:WGUide:SSLT|SSST
:FCUToff?
Cmd Parameter
<NRf> <freq> (hertz)
Query Response
<NR3> <freq> (hertz)
Default Unit
Hz
Range
5 kHz to 20 GHz for MS2028C, MS2038C
5 kHz to 15 GHz for MS2027C, MS2037C
5 kHz to 6 GHz for MS2026C, MS2036C
Example
To set the DUT cutoff frequency for User 1 with calibration method SSLT to 1 GHz:
:SENS:CORR:CKIT:USER:WGU:SSLT:FCUT 1GHz
Front Panel Access
NA
DUT User Name
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT|SSLT|SSST:NAME
Description
Sets the DUT name for the specified user.
Syntax
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT|SSLT|SSST:NAME <string>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT|SSLT|SSST:NAME?
Cmd Parameter
<string> (no parameter data or units)
Query Response
<string> (no parameter data or units)
Example
To set the DUT name for User 1 with line type coax and calibration method SOLT:
:SENS:CORR:CKIT:USER:COAX:SOLT:NAME “SOLT1”
Front Panel Access
NA
 
DUT User Open Offset
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide:SOLT
:OPEN
Description
Sets the DUT capacitance value for the specified user.
Syntax
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:OPEN <length>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:OPEN?
Cmd Parameter
<NRf> <length> (millimeters)
Query Response
<NR3> <length> (millimeters)
Example
To set the DUT open offset for User 1 with line type coax to 3 mm:
:SENS:CORR:CKIT:USER:COAX:SOLT:OPEN 3
Front Panel Access
NA
 
DUT User Short Offset (SSLT)
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSLT:SHORt[1]|2
Description
Sets the DUT Short offset for the specified user.
Syntax
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSLT:SHORt[1]|2 <length>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSLT:SHORt[1]|2?
Cmd Parameter
<NRf> <length> (millimeters)
Query Response
<NR3> <length> (millimeters)
Example
To set the DUT short offset 2 for User 1 with line type coax and calibration method SSLT to 3 mm:
:SENSe:CORRection:CKIT:USER:COAX:SSLT:SHORt2 3
Front Panel Access
NA
 
DUT User Short Offset (SSST)
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSST:SHORt[1]|2|3
Description
Syntax
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSST:SHORt[1]|2|3 <length>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SSST:SHORt[1]|2|3?
Cmd Parameter
<NRf> <length> (millimeters)
Query Response
<NR3> <length> (millimeters)
Example
To set the DUT short offset 2 for User 1 with line type coax and calibration method SSST to 3 mm:
:SENSe:CORRection:CKIT:USER:COAX:SSST:SHORt2 3
Front Panel Access
NA
 
DUT User Short Offset for calibration method SOLT
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide:SOLT
:SHORt
Description
Sets the DUT capacitance value for the specified user.
Syntax
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:SHORt <length>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX|WGUide
:SOLT:SHORt?
Cmd Parameter
<NRf> <length> (millimeters)
Query Response
<NR3> <length> (millimeters)
Example
To set the DUT short offset for User 1 with line type coax to 3 mm:
:SENSe:CORRection:CKIT:USER:COAX:SOLT:SHORt 3
Front Panel Access
NA