VNA Master™Programming Manual : Vector Voltmeter Commands : :MMEMory:STORe Subsystem
 
:MMEMory:STORe Subsystem
The Mass Memory Store subsystem contains commands to transfer from the internal memory to the mass memory device.
Save Setup
:MMEMory:STORe:STATe
Description
Stores the current setup into the file that is specified by14
<filename>. <filename> must be enclosed in either single quotes (‘ ’) or double quotes (“ ”) and must not contain a file extension. The <integer> is used to distinguish whether the calibration should be saving with the setup. Send a 1 to save setup without a calibration. Send a 2 to save setup with calibration.
Syntax
:MMEMory:STORe:STATe <integer>,<filename>
Cmd Parameter
<integer>,<string> (1|2, filename)
Query Response
NA (no query)
Front Panel Access
NA
 
Save Measurement
:MMEMory:STORe:TRACe
Note 
This command is specific for vector network analyzer modes, not for spectrum analyzer mode.
Description
Stores the trace into the file that is specified by <filename>. <filename> must be enclosed in either single quotes (‘ ’) or double quotes (“ ”) and must not contain a file extension. Note that existing files of the same name will not be overwritten. The <integer> parameter is used to distinguish which type of files to save. The following types are available:
<Integer>
:
File type
1
:
Measurement file (default, if number is not 2 to 6)
2
:
S2P Real/Imag
3
:
S2P Lin Mag/Phase
4
:
S2P Log Mag/Phase
5
:
Text
6
:
CSV
Syntax
:MMEMory:STORe:TRACe <integer>,<filename>
Cmd Parameter
<integer>,<string> (1|2, filename)
Query Response
NA (no query)
Example
To save the trace into the file named “trace”.
:MMEMory:STORe:TRACe 1,”trace”
Related Command
:MMEMory:LOAD:TRACe
Front Panel Access
Shift‑7 (File), Save, Change Type (select file type from list)