Display Subsystem This subsystem provides commands that modify the display of data for the user. They do not modify the way in which data is returned to the controller. Scale :DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision <rel ampl> :DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision? Sets the scale (dB/division) for the y-axis. Factory Preset / *RST: 10 dB/div Range: 1 dB to 15 dB Default Unit: dB Front Panel Access: Amplitude, Scale Reference Level :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel <amplitude> :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel? Sets the reference level amplitude value for the y-axis. Note that this may cause a change in attenuation if the automatic input attenuation coupling is enabled. Factory Preset / *RST: 10 dBm Range: With reference level offset = 0 dB: 30 dBm to -130 dBm Default Unit: Current active amplitude unit Front Panel Access: Amplitude, Reference Level See Also: :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFset <rel ampl> :UNIT:POWer Reference Level Offset :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet <rel ampl> :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFSet? Sets the reference level offset value for the y-axis. Factory Preset / *RST: 0 dB Range: -100 dB to 100 dB Default Unit: dB Front Panel Access: Amplitude, RL Offset