Deletes the measurement or setup file specified by <file name>.<file name> should be enclosed in either single quotes (‘’) or double quotes (“”) and should contain a file extension “.stp” for setup and “.spa” for measurement.
Loads the current setup from the file specified by <file name>.The instrument must be in the mode of the saved setup in order to recall that setup. Use :INSTrument:SELect or :INSTrument:NSELect to set the mode.<file name> should be enclosed in either single quotes (‘’) or double quotes (“”) and should include the extension “.stp”.The <integer> parameter is not currently used, but it must be sent. Send a 0.
The instrument must be in the mode of the saved trace in order to recall that trace. Use :INSTrument:SELect or :INSTrument:NSELect to set the mode.Valid <integer> values are 1, 2 and 3. They correspond to traces A, B and C, respectively.<file name> should be enclosed in either single quotes (‘’) or double quotes (“”) and should include a file extension.
Stores the current setup into the file specified by <file name>.<file name> should be enclosed in either single quotes (‘’) or double quotes (“”) and should not contain a file extension.The <integer> parameter is not currently used, but it must be sent. Send a 0.
<file name> should be enclosed in either single quotes (‘’) or double quotes (“”) and should not contain a file extension.The <integer> parameter is not currently used, but it must be sent. Send a 0.