The commands in this subsystem relate to device-specific parameters, not signal-oriented parameters.
Factory Preset / *RST: Setting the TYPE to NONE is the equivalent of setting the Trace Mode to “Norm” on the front panel. The displayed value for a point will be the current measured value for that point.Setting the TYPE to SCALar is the equivalent of setting the Trace Mode to “Avg” on the front panel. The displayed value for a point will be the average of the last <integer> measured values where <integer> is set by [:SENSe]:AVERage:COUNt.Setting the TYPE to MAXimum is the equivalent of setting the Trace Mode to “Max” on the front panel (Max Hold). The displayed value for a point will be the maximum measured value for that point over sweeps.Setting the TYPE to MINimum is the equivalent of setting the Trace Mode to “Min” on the front panel (Min Hold). The displayed value for a point will be the minimum measured value for that point over sweeps.
Factory Preset / *RST: