Spectrum Master™ Programming Manual : 3 Remote Commands : CORRection Subsystem

CORRection Subsystem
This subsystem provides commands for losses or gains external to the instrument.
Input Impedance
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] <integer>
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]?
Sets the input impedance that will be used for amplitude correction and conversion between units (dBm vs. dBV vs. Volts, etc.).
If the value of <integer> is 50 no correction will be performed.
If the value of <integer> is 75 correction will be done based on Anritsu adapter 12N50-75B.
If the value of <integer> is anything else it will look to the value set by [:SENSe]:CORRection:IMPedance[:INPut]:OFFSet for the amplitude correction value. Unit conversion will be done as if the impedance is 75Ω.
If a value other than 50 or 75 is set, it will be returned as 2 by the query form of this command.
 
“Other” Input Impedance Loss
[:SENSe]:CORRection:IMPedance[:INPut]:OFFSet <rel ampl>
[:SENSe]:CORRection:IMPedance[:INPut]:OFFSet?
Sets the value that will be used for amplitude correction when the value set by [:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] is something other than 50 or 75. This value is not applied if the impedance is set to either 50 or 75.
 