Spectrum Masterâ„¢ Programming Manual : 3 Remote Commands : 3.4 WCDMA

3.4 WCDMA
Abort Subsystem
The abort subsystem includes commands that allow the user to stop current measurement activities on the instrument.
Abort
:ABORt
Restarts the current sweep and/or measurement. Resets the trigger system.
If :INITiate:CONTinuous is OFF (i.e. the instrument is in single sweep mode), send the command :INITiate[:IMMediate] to trigger the next sweep.
If :INITiate:CONTinuous is ON (i.e. the instrument is in continuous sweep mode) a new sweep will start immediately.
 
CALCulate Subsystem
The commands in this subsystem process data that has been collected via the SENSe subsystem.
Turn All Markers Off
:CALCulate:MARKer:AOFF
Turns off all markers.
 
Marker State
:CALCulate:MARKer[1]|2|3|4|5|6[:STATe] OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4|5|6[:STATe]?
Sets the specified marker on/off.
 
Marker X Value
:CALCulate:MARKer[1]|2|3|4|5|6:X <x-parameter>
:CALCulate:MARKer[1]|2|3|4|5|6:X?
Sets the location of the marker on the x-axis at the specified location. <x-parameter> is defined in the current x-axis units.
The query version of the command returns the location of the marker on the x-axis.
 
Marker Read Y Value
:CALCulate:MARKer[1]|2|3|4|5|6:Y?
Reads the current Y value for the specified marker. The units are the units of the y-axis.
 
Display Subsystem
This subsystem provides commands that modify the display of data for the user. They do not modify the way in which data is returned to the controller.
Power Offset
:DISPlay:WINDow:TRACe:Y[:SCALe]:OFFSet <rel ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:OFFSet?
Sets the power offset value for the y-axis.
 
Scale
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision <rel ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision?
Sets the scale (dB/division) for the y-axis.
 