Spectrum Masterâ„¢ Programming Manual : 4 Measurement Commands : 4.3 WCDMA

4.3 WCDMA
CONFigure Commands
This set of commands prepares the instrument for the selected measurement. It disables any currently-enabled measurements and activates the specified measurement. All measurement activity is stopped pending an INITiate command.
Current instrument settings may be changed to default values. These changes will be identified with their respective measurement commands.
The :CONFigure? query returns the name of the current measurement.
Configure Demodulation Measurement
:CONFigure:DEMod SUMMary|CDPower|CODogram|HSDPa
The HSDPa measurement is only available if option 65 is installed.
This command configures the selected demodulation measurement. It disables any other active measurements.
None of the instrument parameters are changed by the execution of this command.
Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF).
Measurement settings can be modified by using the [:SENSe]:DEMod commands before initiating a measurement.
When the SUMMARY option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then Demod, then Summary from the front panel.
When the CDP option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then Demod, then CDP from the front panel.
When the CODOGRAM option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then Demod, then Codogram from the front panel.
When the HSDPa option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then Demod, then Hsdpa from the front panel.
 
 
Configure Over-The-Air Measurement
:CONFigure:OTA
This command configures the over-the-air measurement. It disables any other active measurements.
None of the instrument parameters are changed by the execution of this command.
Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF).
Measurement settings can be modified by using the [:SENSe]:OTA commands before initiating a measurement.
The measurement results are equivalent to the results that are displayed by choosing Measurements, then OTA from the front panel.
 
Configure PASS/FAIL Measurement
:CONFigure PFail
This command configures the PASS/FAIL measurement. It disables any other active measurements.
None of the instrument parameters are changed by the execution of this command.
Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF).
Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
The measurement results are equivalent to the results that are displayed by choosing Measurements, then PASS/FAIL from the front panel.
 
Configure RF Measurement
:CONFigure:RF SUMMary|SPECtrum|ACLR|EMISsion|BANDspectrum
This command configures the selected RF measurement. It disables any other active measurements.
None of the instrument parameters are changed by the execution of this command.
Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF).
Measurement settings can be modified by using the [:SENSe]:RF commands before initiating a measurement.
When the SUMMary option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then RF, then Summary from the front panel.
When the SPECtrum option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then RF, then Channel Spectrum from the front panel.
When the ACLR option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then RF, then ACLR from the front panel.
When the EMISsion option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then RF, then Spectral Emission Mask from the front panel.
When the BANDspectrum option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then RF, then Band Spectrum from the front panel.
 
 
Configure Summary Measurement
:CONFigure SUMMARY|PFail
This command configures the UMTS Summary or PASS/FAIL measurement. It disables any other active measurements.
None of the instrument parameters are changed by the execution of this command.
Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF).
Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
In SUMMARY mode, the measurement results are equivalent to the results that are displayed by choosing Measurements, then WCDMA Summary from the front panel. In PFail mode, the results are equivalent to those that are displayed by choosing Measurements, then Pass Fail Mode from the front panel.
 