Spectrum Masterâ„¢ Programming Manual : 4 Measurement Commands : READ Commands

READ Commands
This set of commands combines the ABORt, INITiate and FETCh commands. It aborts any current triggering sequence and sets the trigger state to idle. It then initiates a new active measurement (i.e. begins the collection of new data). When the measurement is complete, it returns the result. These commands will not switch to another measurement.
To get the current measurement data, use the FETCh command.
Read Adjacent Channel Leakage Ratio (ACLR)
:READ:ACLR?
Triggers a new adjacent channel power ratio measurement and returns the results: main channel power, lower adjacent and upper adjacent channel power, and lower alternate and upper alternate channel power. It is a combination of the commands :ABORT;:INITiate;:FETCh:ACLR?
The ACLR measurement must be the active measurement (specified by :CONFigure:RF ACLR). The current measurement can be queried using :CONFigure?
Data is returned as 12 comma-separated values: -20 MHz channel power, -15 MHz channel power, -10 MHz channel power, -5 MHz channel power, Main channel power 1, Main channel power 2, Main channel power 3, Main channel power 4, +5 MHz channel power, +10 MHz channel power, +15 MHz channel power, +20 MHz channel power.
 
Read Channel Power
:READ:CHPower?
Triggers a new channel power measurement and returns the results. It is a combination of the commands :ABORT;:INITiate;:FETCh:CHPower?
The channel power measurement must be active. The current measurement can be queried using :CONFigure?
 
Read Active Code Domain Channel Count
:READ:DEMod:ACTChannels?
Triggers a new measurement and returns the active channel count. It is a combination of the commands :ABORT;:INITiate;:FETCh:DEMod:ACTChannels?
A demodulation measurement must be active. The current measurement can be queried using :CONFigure?
 
Read Code Domain Power
:READ:DEMod:CDPower?
Triggers a new code domain power measurement and returns the results. It is a combination of the commands :ABORT;:INITiate;:FETCh:DEMod:CDPower?
The code domain power measurement must be active. The current measurement can be queried using :CONFigure?
Results are returned as 4 comma-delimited values: CPICH power, PCCPCH power, PICH power, SCCPCH power.
 
Read Carrier Feed Through
:READ:DEMod:CFTHrough?
Triggers a new measurement and returns the carrier feed through results. It is a combination of the commands :ABORT;:INITiate;:FETCh:DEMod:CFTHrough?
The carrier feed through measurement must be active. The current measurement can be queried using :CONFigure?
 
Read Error Vector Magnitude (EVM)
:READ:DEMod:EVM?
Triggers a new measurement and returns the EVM results. It is a combination of the commands :ABORT;:INITiate;:FETCh:DEMod:EVM?
The EVM measurement must be active. The current measurement can be queried using :CONFigure?
 
Read Frequency Error
:READ:DEMod:FERRor?
Triggers a new measurement and returns the frequency error results. It is a combination of the commands :ABORT;:INITiate;:FETCh:DEMod:FERRor?
The frequency error measurement must be active. The current measurement can be queried using :CONFigure?
 
Read Noise Floor
:READ:DEMod:NFLoor?
Triggers a new measurement and returns the noise floor results. It is a combination of the commands :ABORT;:INITiate;:FETCh:DEMod:NFLoor?
The noise floor measurement must be active. The current measurement can be queried using :CONFigure?
 
Read Code Domain Channel Count
:READ:DEMod:NUMChannels?
Triggers a new measurement and returns the channel count. It is a combination of the commands :ABORT;:INITiate;:FETCh:DEMod:NUMChannels?
A demodulation measurement must be active. The current measurement can be queried using :CONFigure?
 
Read Synch Channel Power
:READ:DEMod:SPOWer?
Triggers a new measurement and returns the synch channel power results. It is a combination of the commands :ABORT;:INITiate;:FETCh:DEMod:SPOWer?
The synch channel power measurement must be active. The current measurement can be queried using :CONFigure?
Results are returned as 3 comma-delimited values: SCH power, PSCH power, SSCH power.
 
Read Transmitter Frequency
:READ:DEMod:TXFRequency?
Triggers a new measurement and returns transmitter frequency results. It is a combination of the commands :ABORT;:INITiate;:FETCh:DEMod:TXFRequency?
The carrier frequency measurement must be active. The current measurement can be queried using :CONFigure?
 
Read Spectral Emission Mask
:READ:EMISsion?
Triggers a new spectral emission mask measurement and returns the results: pass/fail status and template number. It is a combination of the commands :ABORT;:INITiate;:FETCh:EMISsion?
The spectral emission mask measurement must be active. The current measurement can be queried using :CONFigure?
Results are returned as 2 comma-delimited values: pass/fail status, and template number.
 
Read Spectral Emission Mask Pass/Fail Status
:READ:EMISsion:FAIL?
Triggers a new spectral emission mask measurement and returns the pass/fail results. It is a combination of the commands :ABORT;:INITiate; :FETCh:EMISsion:FAIL?
The spectral emission mask measurement must be active. The current measurement can be queried using :CONFigure?
 
Read Spectral Emission Mask Template
:READ:EMISsion:TEMPlate?
Triggers a new spectral emission mask measurement and returns the template number result. It is a combination of the commands :ABORT;:INITiate; :FETCh:EMISsion:TEMPlate?
The spectral emission mask measurement must be active. The current measurement can be queried using :CONFigure?
 
Read Occupied Bandwidth
:READ:OBWidth?
Triggers a new occupied bandwidth measurement and returns the result. It is a combination of the commands :ABORT;:INITiate;:FETCh:OBWidth?
The occupied bandwidth measurement must be active. The current measurement can be queried using :CONFigure?
 
Read Over The Air
:READ:OTA[1]|2|3|5|6?
Triggers a new over-the-air measurement and returns the result. It is a combination of the commands :ABORT;:INITiate;:FETCh:OTA?
The over-the-air measurement must be active. The current measurement can be queried using :CONFigure?
The measurement results are returned as 5 comma delimited values: scrambling code, CPICH power, chip energy, Ec/I0 ratio, pilot dominance.
 
Read PASS/FAIL
:READ:PFail?
Triggers a new PASS/FAIL measurement and returns the results. It is a combination of the commands :ABORT;:INITiate;:FETCh:PFail?
The PASS/FAIL measurement must be active. The current measurement can be queried using :CONFigure?
See :FETCh:PFail?
 
Read Peak to Average Power Ratio
:READ:PTAPower?
Triggers a new measurement and returns the peak to average power ratio result. It is a combination of the commands :ABORT;:INITiate;:FETCh:PTAPower?
The peak to average power measurement must be active. The current measurement can be queried using :CONFigure?
 

Copyright 2011 Anritsu Company
Anritsu Company
1800-ANRITSU
10450-00034, Rev. A
mmdtechpubs@anritsu.com