Spectrum Master Programming Manual : Programming with SCPI : Parameter Names
 
Parameter Names
The following tables list the parameter options for the :TRACe:PREamble? command in each supported measurement mode.
Spectrum Analyzer Parameter Names
Available Parameters in Spectrum Analyzer Mode (1 of 6)
Parameter Name
Description
SN
Instrument serial #
UNIT_NAME
Instrument name
DESCR
Trace name
DATE
Trace date/time
BASE_VER
Base FW version
APP_NAME
Application name
APP_VER
Application FW version
UNITS
Amplitude units
CENTER_FREQ
Center frequency
SPAN
Frequency span
FREQ_STEP
Frequency step size
RBW
Resolution bandwidth
RBW_TYPE
RBW coupling auto/manual
VBW
Video bandwidth
VBW_TYPE
VBW coupling auto/manual
RBW_VBW_RATIO
RBW/VBW ratio
SPAN_RBW_RATIO
Span/RBW ratio
INPUT_ATTEN
Input attenuation
ATTEN_TYPE
Attenuation coupling auto/manual
REFERENCE_LEVEL
Reference level
SCALE
y‑axis scale
PREAMP_SET
Preamp state
REF_LEVEL_OFFSET
Reference level offset
DETECTION
Detection type
TRACE_AVERAGE
Number of traces to average
SWEEP_TYPE
Single/continuous
CURRENT_SIGNAL
Current signal index
CURRENT_CHANNEL
Current signal channel
TRACE_MODE
Normal/Avg/Max
TRACE_STATUS
TRACE_A_VIEW_NOT_BLANK: 0x0000000000000001
TRACE_A_WRITE_NOT_HOLD: 0x0000000000000002
TRACE_A_DATA_VALID: 0x0000000000000004
TRACE_B_VIEW_NOT_BLANK: 0x0000000000010000
TRACE_B_WRITE_NOT_HOLD: 0x0000000000020000
TRACE_B_DATA_VALID: 0x0000000000040000
TRACE_C_VIEW_NOT_BLANK: 0x0000000100000000
TRACE_C_WRITE_NOT_HOLD: 0x0000000200000000
TRACE_C_DATA_VALID: 0x0000000400000000
TRACE_C_IS_B_MINUS_A_ON: 0x0000001000000000
TRACE_C_IS_A_MINUS_B_ON: 0x0000002000000000
TRACE_COUNT
Number of traces averaged
UI_DATA_POINTS
Number of display points
IMPEDANCE
Input impedance
REFERENCE_FREQUENCY
Reference freq
SET_SWEEP_TIME
Minimum sweep time setting
TRIGGER_TYPE
Trigger type
VIDEO_TRIGGER_LEVEL
Video trigger level
TRIGGER_POSITION
Trigger position as a percent of the display
PEAK_THRESHOLD
Marker peak search threshold
MARKER_TABLE
Marker table status
ACTIVE_ MEASUREMENT
Current measurement
ANTENNA
Antenna index
OCC_BW_METHOD
Occupied bandwidth method
OCC_BW_PERCENT
Occupied bandwidth % of power setting
OCC_BW_DBC
Occupied bandwidth dBc setting
OCC_BW_MEASURED_ dB
Occupied bandwidth measured dBc value
OCC_BW_MEASURED_ PERCENT
Occupied bandwidth measured % value
OCC_BW_VALUE
Measured occupied bandwidth
OCC_BW_LINE_ MARKER_INFO
Mask off 16 bits at a time to get the display point location of the 3 OBW display indicators
CH_PWR_WIDTH
Channel power integration bandwidth
CH_PWR_VALUE
Measured channel power
CH_PWR_DENSITY
Measured channel power density
CH_PWR_LINE_ MARKER_INFO
Mask off 16 bits at a time to get the display point location of the 2 channel power display indicators
ACPR_MAIN_CH_BW
ACPR main channel bandwidth
ACPR_ADJC_CH_BW
ACPR adjacent channel bandwidth
ACPR_CHANNEL_ SPACING
ACPR channel spacing
ACPR_MAIN_CH_PWR
ACPR measured main channel power
ACPR_UPPER_CH_PWR
ACPR measured upper channel power
ACPR_LOWER_CH_ PWR
ACPR measured lower channel power
ACPR_LOWER_CH_ LINE_MARKER_INFO
Mask off 16 bits at a time to get the display point location of the 2 ACPR lower channel display indicators
ACPR_MAIN_CH_LINE_ MARKER_INFO
Mask off 16 bits at a time to get the display point location of the 2 ACPR main channel display indicators
ACPR_UPPER_CH_ LINE_MARKER_INFO
Mask off 16 bits at a time to get the display point location of the 2 ACPR upper channel display indicators
AM_FM_DEMOD_VOL
AM/FM demod volume
AM_FM_DEMOD_ FREQUENCY
AM/FM demod freq
AM_FM_DEMOD_TYPE
AM/FM demod type
AM_FM_DEMOD_TIME
AM/FM demod time
AM_FM_LINE_ MARKER
Display point location of the demodulation frequency
BEAT_FREQUENCY_ OSC_FREQUENCY
BFO oscillator freq
CI_C_TYPE
C/I measurement carrier type
CI_C_VALUE
C/I measurement measured carrier power
CI_I_BB_VALUE
C/I measurement measured broadband interference power
CI_I_NB_VALUE
C/I measurement measured narrowband interference power
CI_I_WB_VALUE
C/I measurement measured wideband interference power
CI_BB_VALUE
C/I measurement with broadband interference
CI_NB_VALUE
C/I measurement with narrowband interference
CI_WB_VALUE
C/I measurement with wideband interference
MKR_SPA_FREQNx
Marker x frequency (where x is the marker number 0 to 11, 0 represent the reference marker #1 and 1 represent delta marker #1, 2 represent reference marker #2, and 3 represent delta marker #2, and so on)
MKR_SPA_POINTx
Reference marker x display point
MKR_SPA_MAGNTx
Reference marker x magnitude
MKR_SPA_PRCNTx
Reference marker x display percentage
MKR_SPA_FLAGSx
Reference marker x flags:
SPA_MKR_FLAG_ON_OFF: 0x00000001
SPA_MKR_FLAG_DELTA_MKR: 0x00000002
SPA_MKR_FLAG_SELECTED: 0x00000004
SPA_MKR_FLAG_DATA_INVALID: 0x00000008
SPA_MKR_FLAG_DATA_STALE: 0x00000010
SPA_MKR_FLAG_FIXED: 0x00000020
SPA_MKR_FLAG_MASK: 0x000000FF
SPA_MKR_FLAG_DISPL_AMPL_HZ: 0x00000100
SPA_MKR_FLAG_DISPL_AMPL_PER_HZ: 0x00000200
SPA_MKR_FLAG_DISP_FLAG: 0x00000F00
SPA_MKR_FLAG_RELATIVE: 0x00001000
SPA_MKR_STANDARD: 0x10000000
SPA_MKR_FIELD_STRENGHT: 0x20000000
SPA_MKR_NOISE: 0x30000000
SPA_MKR_COUNTER: 0x40000000
SPA_MKR_TIME: 0x50000000
MKR_SPA_REF_TOx
Specifies which marker is the marker x reference to
MKR_SPA_TRACex
Specifies which trace the marker x is for.
LIM_LFLAGS_UP
Upper limit flags:
LIMIT_FLAG_ON: 0x00000004
LIMIT_FLAG_ALARM_ON: 0x00000002
LIM_FREQNC_UPx
Upper limit point x freq (where x is the limit point number starting with 0)
LIM_MAGNTD_UPx
Upper limit point x amplitude
LIM_LFLAGS_LO
Lower limit flags:
LIMIT_FLAG_ON: 0x00000004
LIMIT_FLAG_ALARM_ON: 0x00000002
LIM_FREQNC_LOx
Lower limit point x freq (where x is the limit point number starting with 0)
LIM_MAGNTD_LOx
Lower limit point x amplitude
Parameters Related to Tracking Generator
EXTERNAL_GAIN_LOSS
Generator External Gain/Loss Setting
SG_FREQ
Generator CW Frequency
TG_GEN_SELECT Generator
CW/Tracking Mode Selection
TG_IL_ABS_AVG_PWR_VALUE
Generator Power Statistics (Average Value)
TG_IL_ABS_MAX_PWR_VALUE
Generator Power Statistics (Max Value)
TG_IL_ABS_MIN_PWR_VALUE
Generator Power Statistics (Min Value)
TG_LEVEL_1 Generator
Output Power Level
TG_S21_NORMALIZE_ON_OFF
Generator Normalize Measurement
GENERATOR_OUTPUT
Generator State
TG_REF_POSITION
Normalized Trace Reference Position
TG_REF_AMPLITUDE
Reference Amplitude at the Reference Position
TG_SCALE
Normalized Trace Scale
ACPR_UPPER_ALT_CH_PWR
ACPR measured upper alternate channel power
ACPR_LOWER_ALT_CH_PWR
ACPR measured lower alternate channel power
TG_S21_NORMALIZE_ON_OFF
TG Normalization State
TG_IL_REL_STATS
TG Transmission Measurement Statistics Table State
TG_IL_ABS_STATS
Generator Power Statistics Table State
TG_REF_LEVEL_OFFSET
TG Transmission Offset
TG_IL_REL_MAX_PWR_VALUE
TG Transmission Measurement Statistics (Max Value)
TG_IL_REL_AVG_PWR_VALUE
TG Transmission Measurement Statistics (Average Value)
TG_IL_REL_MIN_PWR_VALUE
TG Transmission Measurement Statistics (Min Value)
 
 
GSM Parameter Names
Available Parameters in GSM Mode
Parameter Name
Description
SN
Instrument serial #
UNIT_NAME
Instrument name
TYPE
The data type (Setup or Data)
DESCR
Trace name
DATE
Trace date/time
BASE_VER
Base FW version
APP_NAME
Application name
APP_VER
Application FW version
APP_MODE
Application Mode
REFERENCE_LEVEL
Reference Level
CENTER_FREQ
Center freq
SIGNAL_STANDARD
Current signal standard (the value is the index of the signal standard list, where a value of 1 is the first index in the list)
CHANNEL
Current channel
POWER_OFFSET
Applied power offset
REFERENCE_FREQUENCY
Selected external reference frequency
UNITS
Amplitude units
CURRENT_VIEW
Current view
CURRENT_MEASUREMENTS
Current measurements
DYNAMIC_ATTENUATION
Dynamic range on/off
SPAN
Frequency span
MANUAL_SYSTEM_SELECT
GSM/EDGE select
CURRENT_SPECTRUM_VIEW
Current spectrum view (single or multiple channel spectrum)
I_Q_VIEW
IQ view
RUN_HOLD
Run/Hold on/off
MC_SPECTRUM_START_FREQ
Multi‑channel spectrum start frequency
MC_SPECTRUM_STOP_FREQ
Multi‑channel spectrum stop frequency
MC_CHANNEL_CURSOR
Multi‑channel spectrum channel cursor (used for markers)
MC_FREQ_CURSOR
Multi‑channel spectrum frequency cursor (used for markers)
TEST_MODEL
Current pass fail model being tested
WiMAX, Mobile WiMAX Parameter Names
Available Parameters in WiMAX and Mobile WiMAX Mode (1 of 2)
Parameter Name
Description
SN
Instrument serial #
UNIT_NAME
Instrument name
TYPE
The data type (Setup or Data)
DESCR
Trace name
DATE
Trace date/time
BASE_VER
Base FW version
APP_NAME
Application name
APP_VER
Application FW version
APP_MODE
Application Mode
REFERENCE_LEVEL
Reference Level
UNITS
Amplitude units
SCALE
Y axis scale
CENTER_FREQ
Center freq
SIGNAL_STANDARD
Current signal standard (the value is the index of the signal standard list, where a value of 0 is the first index in the list)
CHANNEL
Current channel
POWER_OFFSET
Applied power offset
REFERENCE_FREQUENCY
Selected external reference frequency
UNITS
Amplitude units
CURRENT_VIEW
Current view
CURRENT_MEASUREMENTS
Current measurements
DYNAMIC_ATTENUATION
Dynamic range on/off
SPAN
Frequency span
PVT_FRAME_START_TIME
Power vs. Time start time
PVT_FRAME_STOP_TIME
Power vs. Time stop time
BW_SELECT
Current Bandwidth (MHz)
CURRENT_SPECTRUM_VIEW
Current spectrum view (single or multiple channel spectrum)
I_Q_VIEW
IQ view
RUN_HOLD
Run/Hold on/off
TEST_MODEL
Current pass fail model being tested
CP_RATIO
Cyclic Prefix Ratio (the value is the index of the CP Ratio list, where a value of 0 is the first index in the list)
SPECTRUM_SPAN
Spectrum frequency span (the value is the index of the span list, where a value of 0 is the first index in the list
AUTO_SPAN
Auto span on/off
MAX_HOLD
Max hold on/off
EVM_SUB_CARRIER_TOP
Y Axis Max for EVM vs. Symbol/Sub‑Carrier
EVM_SUB_CARRIER_SCALE
EVM vs. Symbol/Sub‑Carrier scale
SPECTRAL_FLATNESS_SCALE
Spectral Flatness scale
ACPR_DISPLAY_GRAPH
Display ACPR trace on/off
NUM_OF_ACPR_MAIN_CHANNELS
Number of main channels
NUM_OF_ACPR_ADJC_CHANNELS
Number of adjacent channels
CONSTELLATION_REFERENCE_POINTS
Constellation reference points on/off
SPECTRAL_FLATNESS_TOP
Y Axis Max for Spectral Flatness
SPECTRAL_FLATNESS_EVM_SUB_START
Spectral Flatness/EVM vs. sub‑carrier start
SPECTRAL_FLATNESS_EVM_SUB_STOP
Spectral Flatness/EVM vs. sub‑carrier stop
EVM_SYM_START
EVM vs. Symbol start
EVM_SYM_STOP
EVM vs. Symbol stop
WCDMA Parameter Names
Available Parameters in WCDMA Mode (1 of 2)
Parameter Name
Description
SN
Instrument serial #
UNIT_NAME
Instrument name
TYPE
The data type (Setup or Data)
DESCR
Trace name
DATE
Trace date/time
BASE_VER
Base FW version
APP_NAME
Application name
APP_VER
Application FW version
APP_MODE
Application Mode
CENTER_FREQ
Center frequency
AUTO_SCRAMBLING_CODE
Auto scrambling code detection on/off
SCRAMBLING_CODE
Selected manual scrambling code
MAX_SPREADING_FACTOR
Maximum spreading factor
REFERENCE_LEVEL
Reference Level
THRESHOLD
Active channel threshold
CURRENT_MEASUREMENTS
Currently active measurements
SCALE
Display scale (dB/division)
POWER_OFFSET
Applied power offset
SCCPCH_CODE
Selected S‑CCPCH channelization code
PICH_CODE
Selected PICH channelization code
CURRENT_VIEW
Current view
CDP_ZOOM
Zoom range for code domain power
CDP_ZOOM_START
Start code for zoom window
REFERENCE_FREQUENCY
Reference Frequency
SCCPCH_SPREAD_FCT
Selected S‑CCPCH spreading factor
RUN_HOLD
Run/hold state
EXT_REF_FREQ
Selected external reference frequency
DYNAMIC_ATTENUATION
Dynamic range on/off
ACLR_DISPLAY_GRAPH
Display ACLR trace on/off
MARKER_TABLE
Marker table on/off
OTA_AUTO_SC
OTA auto scrambling code on/off
OTA_CODE_LOCK
OTA code lock on/off
OTA_DISPLAY_VALUE
Selected OTA display value
OTA_SORT_PARAM
Selected OTA sort parameter
OTA_MANUAL_SC_1
Selected manual scrambling code 1
OTA_MANUAL_SC_2
Selected manual scrambling code 2
OTA_MANUAL_SC_3
Selected manual scrambling code 3
OTA_MANUAL_SC_4
Selected manual scrambling code 4
OTA_MANUAL_SC_5
Selected manual scrambling code 5
OTA_MANUAL_SC_6
Selected manual scrambling code 6
CODOGRAM_SINGLE_SWEEP_TIME
Single sweep time for codogram
TEST_MODEL
Current pass fail model being tested
CONSTELLATION_CODE
Code number for which IQ data is being displayed
CONSTELLATION_SPREAD
Spreading factor of code for which IQ data is being displayed
NUM_OF_ACLR_MAIN_CHANNELS
Number of main channels
NUM_OF_ACLR_MAIN_CHANNELS
Number of adjacent channels
BAND_SPECTRUM_START_FREQ
Band spectrum start frequency
BAND_SPECTRUM_STOP_FREQ
Band spectrum stop frequency
DL_CHANNEL_NUMBER
Currently selected band spectrum channel
AUTO_THRESHOLD
Auto threshold value
HSDPA_DEMOD_SINGLE_SWEEP_TIME
Single sweep time for hsdpa
BAND_SPECTRUM_REFERENCE_LEVEL
Reference level for band spectrum
CONSTELLATION_PERSISTENCE
Persistence level for IQ data
SIGNAL_STANDARD
Current Signal standard
CHANNEL
Current channel
CONSTELLATION_MARKER
Display location of the IQ marker
RF_MKR_POINT_[1]|2|3|4|5|6
RF marker point number
RF_MKR_STATE_[1]|2|3|4|5|6
RF marker on/off
DEMOD_MKR_POINT_[1]|2|3|4|5|6
Code domain marker point number
DEMOD_MKR_STATE_[1]|2|3|4|5|6
Code domain marker on/off
CODOGRAM_MKR_POINT_1
Codogram marker point number
CODOGRAM_MKR_STATE_1
Codogram marker on/off
CODOGRAM_MKR_TIME_1
Codogram marker time offset
CDMA Parameter Names
Available Parameters in CDMA Mode (1 of 2)
SN
Instrument serial #
UNIT_NAME
Instrument name
TYPE
The data type (Setup or Data)
DESCR
Trace name
DATE
Trace date/time
BASE_VER
Base FW version
APP_NAME
Application name
APP_VER
Application FW version
APP_MODE
Application Mode
CENTER_FREQ
Center frequency
REFERENCE_LEVEL
Reference Level
THRESHOLD
Active channel threshold
CURRENT_MEASUREMENTS
Currently active measurements
SCALE
Display scale (dB/division)
SPAN
Span
POWER_OFFSET
Applied power offset
CURRENT_VIEW
Current view
CDP_ZOOM
Zoom range for code domain power (16/32/64)
CDP_ZOOM_START
Start code for zoom window
REFERENCE_FREQUENCY
Reference Frequency
RUN_HOLD
Run/hold state
DYNAMIC_ATTENUATION
Dynamic range on/off
MARKER_TABLE
Marker table on/off
TEST_MODEL
Current pass fail model being tested
NUM_OF_ACLR_MAIN_CHANNELS
Number of main channels
NUM_OF_ACLR_ADJC_CHANNELS
Number of adjacent channels
SIGNAL_STANDARD
Current Signal standard
CHANNEL
Current channel
CHANNEL_WIDTH
Channel width
DATA_MODULATION
Data channel modulation type
PN_OFFSET
PN offset
WALSH_CODE
Walsh code (64/128)
CDP_SCALE
Code Domain Power scale
MEAS_SPEED
Measurement speed (Fast/Normal/Slow)
PN_TYPE
PN trigger type (No Trig/GPS/Ext)
PN_SEARCH_TYPE
PN search type (Auto/Manual)
TRIGGER_POLARITY
Trigger polarity (Rising/Falling)
PN_INCREMENT
PN increment
SPUR_EMM_DISPLAY_POINTS
Number of Spurious Emission display points
CDP_UNITS
Code Domain Power unit type (Relative/Absolute)
USER_PWR_UNIT
Default unit (dBm/W/mW)
BAND_CLASS
Band class
USER_CARRIER_BW
Carrier bandwidth (1.23/1.24/1.25)
RF_MKR_POINT_[1]|2|3|4|5|6
RF marker point number
RF_MKR_STATE_[1]|2|3|4|5|6
RF marker on/off
DEMOD_MKR_POINT_[1]|2|3|4|5|6
Code domain marker point number
DEMOD_MKR_STATE_[1]|2|3|4|5|6
Code domain marker on/off
EVDO Parameter Names
Available Parameters in EVDO Mode (1 of 2)
Parameter Name
Description
SN
Instrument serial #
UNIT_NAME
Instrument name
TYPE
The data type (Setup or Data)
DESCR
Trace name
DATE
Trace date/time
BASE_VER
Base FW version
APP_NAME
Application name
APP_VER
Application FW version
APP_MODE
Application Mode
CENTER_FREQ
Center frequency
REFERENCE_LEVEL
Reference Level
THRESHOLD
Active channel threshold
CURRENT_MEASUREMENTS
Currently active measurements
SCALE
Display scale (dB/division)
SPAN
Span
POWER_OFFSET
Applied power offset
CURRENT_VIEW
Current view
CDP_ZOOM
Zoom range for code domain power (16/32/64)
CDP_ZOOM_START
Start code for zoom window
REFERENCE_FREQUENCY
Reference Frequency
RUN_HOLD
Run/hold state
DYNAMIC_ATTENUATION
Dynamic range on/off
MARKER_TABLE
Marker table on/off
TEST_MODEL
Current pass fail model being tested
NUM_OF_ACLR_MAIN_CHANNELS
Number of main channels
NUM_OF_ACLR_ADJC_CHANNELS
Number of adjacent channels
SIGNAL_STANDARD
Current Signal standard
CHANNEL
Current channel
CHANNEL_WIDTH
Channel width
DATA_MODULATION
Data channel modulation type
PN_OFFSET
PN offset
WALSH_CODE
Walsh code (64/128)
POWER_TIME
Power vs. Time slot type
CDP_SCALE
Code Domain Power scale
SPUR_EMM_DISPLAY_POINTS
Number of Spurious Emission display points
POWER_VS_TIME_THRESHOLD
Power vs. Time threshold
MEAS_SPEED
Measurement speed (Fast/Normal/Slow)
PN_TYPE
PN trigger type (No Trig/GPS/Ext)
PN_SEARCH_TYPE
PN search type (Auto/Manual)
TRIGGER_POLARITY
Trigger polarity (Rising/Falling)
PN_INCREMENT
PN increment
CDP_UNITS
Code Domain Power unit type (Relative/Absolute)
POWER_VS_TIME_START
Power vs. Time start chip
POWER_VS_TIME_STOP
Power vs. Time stop chip
USER_PWR_UNIT
Default unit (dBm/W/mW)
BAND_CLASS
Band class
USER_CARRIER_BW
Carrier bandwidth (1.23/1.24/1.25)
RF_MKR_POINT_[1]|2|3|4|5|6
RF marker point number
RF_MKR_STATE_[1]|2|3|4|5|6
RF marker on/off
DEMOD_MKR_POINT_[1]|2|3|4|5|6
Code domain marker point number
DEMOD_MKR_STATE_[1]|2|3|4|5|6
Code domain marker on/off
TDSCDMA Parameter Names
Available Parameters in TDSCDMA Mode (1 of 2)
SN
Instrument serial #
UNIT_NAME
Instrument name
TYPE
The data type (Setup or Data)
DESCR
Trace name
DATE
Trace date/time
BASE_VER
Base FW version
APP_NAME
Application name
APP_VER
Application FW version
APP_MODE
Application Mode
CENTER_FREQ
Center frequency
SIGNAL_STANDARD
Current Signal Standard
CHANNEL
Current Signal Standard Channel
SPAN
Span
REFERENCE_LEVEL
Reference Level
SCALE
Display Scale
POWER_OFFSET
Applied power offset
UNITS
0 = dBm, 1 = Watts
DYNAMIC_ATTENUATION
Dynamic range on/off
REFERENCE_FREQUENCY
Reference Frequency
OTA_DISPLAY_MODE
0 = Code Scan, 1 = Tau Scan
CURRENT_VIEW
Current view
CURRENT_MEASUREMENTS
Currently active measurements
POWER_TIME_VIEW
0 = Sub‑Frame, 1 = Slot View
POWER_VS_TIME_START
Start time
POWER_VS_TIME_STOP
Stop time
SLOT_NUMBER_SET
Slot number (7 = auto)
SLOT_NUMBER_RETURN
Slot number
Dynamic Range
0 = Normal, 1 = Wide
MARKER_TABLE
Marker table on/off
NUM_OF_ACLR_MAIN_CHANNELS
Number of main channels
NUM_OF_ACLR_ADJC_CHANNELS
Number of adjacent channels
CHANNEL_WIDTH
Channel width
SCRAMBLING_CODE_SET
Scrambling code (128 = auto)
SCRAMBLING_CODE_SET_RETURN
Actual detected scrambling code
SCRAMBLING_CODE_1
Scrambling code 1
SCRAMBLING_CODE_2
Scrambling code 2
SCRAMBLING_CODE_3
Scrambling code 3
SCRAMBLING_CODE_4
Scrambling code 4
SPREADING_FACTOR
Spread factor
SYNC_DL_CODE_SET
Sync DL code (32 = auto)
SYNC_DL_CODE_RETURN
Actual detected sync DL code set
DWPTS
DWPTS Power
NUM_OF_CARRIERS
0 = 1 carrier, 1 = 3 carriers
MAX_USERS_SET
Max users set (0 = auto)
MAX_USERS_RETURN
Max users detected
UPLINK_SLOT_SWITCH_POINT
Uplink slot switch point set
THRESHOLD
Threshold
MEAS_SPEED
Measurement speed (Fast/Normal/Slow)
TAU_OFFSET
Tau time offset
TRIGGER_TYPE
Trigger type (No Trig/GPS/External)
TRIGGER_POLARITY
Trigger polarity (Rising/Falling)
TRIGGER_ACCURACY
Trigger accuracy
CDP_ZOOM
CDP Zoom
CDP_ZOOM_START
CDP Zoom Start
WALSH_CODE
Walsh Code
CDP_SCALE
CDP scale
CDP_UNITS
CDP units (Absolute/Relative)
DATA_MODULATION
Data Modulation (QPSK/8PSK/16QAM)
DEMOD_MODE
Demod mode
RUN_HOLD
Run/hold
CODE_AUTOSAVE
Code autosave on/off
TEST_MODEL
Current pass fail model being tested
RF_MKR_POINT_[1]|2|3|4|5|6
RF marker point number
RF_MKR_STATE_[1]|2|3|4|5|6
RF marker on/off
DEMOD_MKR_POINT_[1]|2|3|4|5|6
Code domain marker point number
DEMOD_MKR_STATE_[1]|2|3|4|5|6