Spectrum Master Programming Manual : Tracking Generator Commands : :GENerator Subsystem
 
:GENerator Subsystem
This subsystem contains commands related to the generator.
:GENerator:OUTput:POWer <amplitude>
:GENerator:OUTput:POWer?
Title
Generator Output Power
Description
Sets the output power for the Generator.
Query returns actual output power setting without regards to any external gain or loss setting.
Note: Returned value will not match displayed output power on the instrument if external gain/loss setting is not 0 dB.
Parameter
<amplitude>
Default Value
–40 dBm
Range
–50 dBm to +15 dBm
Default Unit
Current active amplitude unit
Example
To set the output to –30 dBm when the active amplitude units are dBm:
:GENerator:OUTput:POWer 30
Related Command
:INITiate:GENerator
:GENerator:OUTput:POWer:OFFSet[:MAGNitude]
Front Panel Access
Shift‑4 (Measure), Generator, Output Power
:GENerator:OUTput:POWer:OFFSet[:MAGNitude] <value>
:GENerator:OUTput:POWer:OFFSet[:MAGNitude]?
Title
Generator Output Power Offset or External Gain/Loss Setting
Description
Sets the output power level offset or the External Gain/Loss Setting. Note that changing this value also causes the displayed output power to adjust with the new offset. For example, if the output power is set to –25 dBm, and if the External Gain/Loss is then set to 10 dB external gain, then the power limits will be adjusted by 10 dB, and the value of the output power that is displayed on the instrument will be adjusted to –15 dBm. The query returns a value from –100 to +100. A negative sign indicates external loss, and no sign indicates external gain.
Parameter
<value>
Default Value
0 dB
Range
100 dB Ext Loss to 100 dB Ext Gain
Default Unit
dB
Example
To set the signal generator offset to 10 dB external gain:
:GENerator:OUTput:POWer:OFFSet 10
Related Command
:GENerator:OUTput:POWer
Front Panel Access
Shift‑4 (Measure), Generator, Settings, Ext Gain/Loss