Spectrum Master Programming Manual : TDSCDMA Commands : :CONFigure Subsystem
 
:CONFigure Subsystem
This set of commands prepares the instrument for the selected measurement. It disables any currently‑enabled measurements and activates the specified measurement. It sets the instrument to single sweep mode, waiting for an :INITiate command. It will not initiate the taking of a measurement.
Current instrument settings may be changed to default values. These changes are identified with their respective measurement commands.
:CONFigure SUMMary|PFail
Title
Configure Summary and Pass/Fail Measurements
Description
This command configures the summary and PASS/FAIL measurements. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
Parameter
SUMMary|PFail
Parameter Type
<char>
Front Panel Access
Shift‑4 (Measure), TD‑SCDMA Summary or Pass/Fail Mode
:CONFigure:DEMod <char>
Title
Configure Demodulation Measurement
Description
Valid <char> parameter options: SUMMary|CDPData
This command configures the selected demodulation measurement. It disables any other active measurements. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe]:DEMod commands before initiating a measurement. When the SUMMARY option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then Demodulator, then Modulation Summary from the front panel. When the CDPData option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then Demodulator, then CDP from the front panel.
Parameter
<char>
Front Panel Access
Measurements, Demodulator
 
:CONFigure:OTA
Title
Configure Over‑The‑Air Measurement
Description
Valid parameter options: CSCan|TSCan
This command configures the selected Over The Air measurement. It disables any other active measurements. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified before initiating a measurement. When the CSCan option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then OTA, then Code Scan from the front panel. When the TSCan option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then OTA, then Tau Scan from the front panel.
:CONFigure:RF <char>
Title
Configure RF Measurement
Description
Valid parameter options: SUMMary|SPECtrum|PVTSlot|EMISsion
This command configures the selected RF measurement. It disables any other active measurements. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe]:RF commands before initiating a measurement.
When the SUMMary option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then RF Measurements, then RF Summary from the front panel. When the SPECtrum option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then RF Measurements, then Channel Spectrum from the front panel. When the PVTSlot option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then RF, then Power vs. Time from the front panel. When the EMISsion option is selected, the measurement results are equivalent to the results that are displayed by choosing Measurements, then RF Measurements, then Spectrum Emission from the front panel.
Parameter
<char>
Related Command
[:SENSe]:FREQuency