Spectrum Master Programming Manual : TD‑LTE Commands : :CONFigure Subsystem
 
:CONFigure Subsystem
This set of commands prepares the instrument for the selected measurement. It disables any currently‑enabled measurements and activates the specified measurement. It sets the instrument to single sweep mode, waiting for an :INITiate command. It will not initiate the taking of a measurement.
Current instrument settings may be changed to default values. These changes are identified with their respective measurement commands.
:CONFigure SUMMary
Title
Set Measurement to TD‑LTE Summary
Description
Sets the measurement to TD‑LTE Summary. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
Parameter
SUMMary
Parameter Type
<char>
Front Panel Access
Measurements, TD‑LTE Summary
:CONFigure?
Title
Configure Query
Description
:CONFigure? query returns the name of the measurement previously set up using a CONFigure command or a MEASure? query. The list below shows the possible return values.
SUMM
RF,SPEC
RF,ACLR
RF,SEM
RF, PVT, SUBF
RF, PVT, FRAM
RF,SUMM
DEM,CONS
DEM,CCPG
DEM,CCPT
DEM,PVRB
DEM,SUMM
PF
OTA,OTAS
OTA,OTATX
OTA,MAP
:CONFigure:DEMod SUMMary|CONStln|CCPGraph|CCPTable
Title
Select the Modulation Measurement
Description
Sets measurement to one of the Modulation screens. This can be one of:
SUMMary: Modulation Summary view
CONStln: Modulation Constellation diagram.
CCPGraph: Control Channel Powers - Bar‑graph view.
CCPTable: Control Channel Powers - Table view.
It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF). Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
Parameter
SUMMary|CONStln|CCPGraph|CCPTable
Parameter Type
<char>
Front Panel Access
Measurements, Modulation, (Constellation, Control Channel Power, Modulation Summary). If Control Channel Power is selected, press Control Channel Power again then press Display Mode to select Bar Graph or Table.
:CONFigure:OTA SCANner|TXTEst|MAPping
Title
Select the Over‑the‑Air Scanner Measurement
Description
Sets the measurement to one of the Over‑the‑Air screens.
Parameter
SCANner|TXTEst|MAPping
Parameter Type
<char>
Front Panel Access
Measurements, Over‑the‑Air, Scanner
 
:CONFigure:RF SUMMary|SPECtrum|ACLR|SEM|PVTime
Title
Select RF Measurements
Description
This command configures RF measurements. It disables any other active measurements. None of the instrument parameters are changed by the execution of this command. Sets the instrument to single sweep mode (:INITiate:CONTinuous OFF).Measurement settings can be modified by using the [:SENSe] commands before initiating a measurement.
Set measurement to one of the RF measurement screens. This can be one of:
SUMMary: RF measurement summary screen.
SPECtrum: RF Channel Spectrum measurement screen.
ACLR: ACLR measurement view.
SEM: Spectral Emission Mask measurement.
PVTime: Power vs. Time measurement
 
Parameter
SUMMary|SPECtrum|ACLR|SEM|PVTime
Front Panel Access
Measurements, RF, (Channel Spectrum, Power vs. Time, ACLR, Spectral Emission Mask, RF Summary)
:CONFigure:RF:PVTime FRAMe|SUBFrame
Title
Configure Power vs Time Measurements
Description
Sets the current display mode for Power vs. Time measurements, either ‘Frame’ or ‘Sub‑frame’.
Parameter
FRAMe|SUBFrame
Parameter Type
<char>
Front Panel Access
Measurements, RF, Power vs. Time, View