PIM Master™ Passive Intermodulation Analyzer : PIM versus Time : Frequency (Freq) Menu
  
Frequency (Freq) Menu
Key Sequence: Freq
Frequency Menu
l
 
Carrier F1
Press this submenu key and enter the desired frequency using the number keypad or the Arrow keys. Enter a frequency within the acceptable range by referring to Table: PIM Master Carrier Bands and Frequencies.
Carrier F2
Press this submenu key and enter the desired frequency using the keypad or the Arrow keys. Enter a frequency within the acceptable range by referring to Table: PIM Master Carrier Bands and Frequencies.
Intermod Order
3rd 5th 7th
Select the 3rd, 5th, or 7th order of intermodulation. If the selected F1 and F2 do not produce the desired IM product, the PIM Master automatically changes F1 and F2 to factory default frequencies that will generate an in‑band IM product. F1and F2 can be further adjusted if desired.
 
Band Select
High Low
Press this submenu key to toggle the band selection to High or Low. This submenu key appears only on instruments with multiple receive ranges. For Option 193, the toggle choices are PCS and PCS/AWS.
PIM Aid
Press this submenu key to display the PIM Parameters dialog box (see Figure: PIM Parameters Dialog Box. This dialog box allows editing of the current carrier frequencies and carrier power. It also displays the intermodulation products generated by the carrier frequencies.
Note 
The left‑most main menu key is labeled Freq in the PIM vs. Time and the Swept PIM measurement settings. It is labeled Distance in the Distance‑to‑PIM measurement setting.