PIM Masterâ„¢ Passive Intermodulation Analyzer : Quick Start Guide : Calibrate Menu
  
Calibrate Menu
Key Sequence: Shift > Cal (2)
Calibration Menu
 
 
Start Calibration
Press this submenu key to begin calibration of the PIM Master. You need a PIM standard and a Low PIM Load. Follow screen prompts until calibration is complete.
 
 
Custom Calibrations
Press this submenu key to display the Custom Calibrations Menu. Note that custom calibrations relate to PIM versus Time measurements only.
Calibration Due Date
Press this submenu key to enter the date (Day, Month, and Year) when your PIM Master requires calibration.
Calibration Reminder
Press this submenu key to display a Calibration Reminder notice in Days. You can use the number keypad to key in a new reminder duration. Press Esc or any other main menu key or submenu key to dismiss the reminder notice. Set to zero to disable.
If calibration is due in less time than this setting, then the reminder notice is displayed when instrument power is turned On.
Reset Calibration
Press this submenu key to reset the calibration of the PIM Master. This deletes any saved calibration files that match the current instrument settings (power level, PIM Order, and so forth). These can be calibrations for any PIM analyzer function at the set power level.
Back
Press the Back submenu key to return to the previous menu.
Note 
During calibration, RF is On, but the RF ON message is not displayed. The CALIBRATION IN PROCESS... message is displayed in red letters on the measurement screen.