PIM Master™ Passive Intermodulation Analyzer : PIM Analyzer Programming Commands : :CALibration Subsystem
  
:CALibration Subsystem
The commands in this subsystem control the system calibration.
:CALibration:PIManalyzer:FULL
:CALibration:PIManalyzer:FULL?
Title
Full Calibration
Description
Sets or resets the calibration for the instrument. This calibration is done for all 3 measurement types at their current configuration (in other words, output power, frequency (PVT only), and IMD order for PVT and Swept PIM). The query returns 0 for “CAL OFF” or 1 for “CAL ON”.
Parameter
OFF|ON|0|1
Parameter Type
<boolean>
Default Value
None. Returns error
Example
To start full calibration:
:CALibration:PIManalyzer:FULL ON
To query the calibration state:
:CAL:PIM:FULL?
Front Panel Access
Shift‑Cal (2), Start Calibration
Shift‑Cal (2), Reset Calibration