PIM Master™ Passive Intermodulation Analyzer : PIM Analyzer Programming Commands : :SENSe Subsystem
  
:SENSe Subsystem
The commands in this subsystem relate to device‑specific parameters, not signal‑oriented parameters.
[:SENSe]:DTPMeas:AVERage:TYPE
[:SENSe]:DTPMeas:AVERage:TYPE?
Title
DTP Trace Mode
Description
Sets and queries DTP trace mode
Parameter
NORMal = normal
MAXHold = max trace hold
Default Value:
None – Returns error
Example
To set DTP trace mode to normal:
:SENSe:DTPMeas:AVERage:TYPE NORMal
Front Panel Access
Setup, Normal -> A / Max Hold -> A
 
[:SENSe]:DTPMeas:DISPlay:TOP|BOTTom
[:SENSe]:DTPMeas:DISPlay:TOP?|BOTTom?
Title
DTP Display Top and Bottom
Description
Sets and queries DTP display magnitude
Parameter
top/bottom magnitude
Range
top = 0, bottom = 260
Default Value:
None – Returns error
Example
To set the DTP display top to 10 dB:
:SENSe:DTPMeas:DISPlay:TOP 10.0
Front Panel Access
Amplitude, Top/Bottom
 
[:SENSe]:DTPMeas:ENREsolution ON|OFF
[:SENSe]:DTPMeas:ENREsolution?
Title
PIM Master Enhanced Resolution
Description
This feature estimates the location of PIM sources on the measurement trace and displays up to 4 vertical impulse bars from bottom of measurement grid to point on measurement trace that indicates a predicted PIM source location. This is useful for resolving PIM sources that are physically close together.
Parameter
ON|OFF
Example
To set Enhanced Resolution to On:
:SENSe:DTPMeas:ENREsolution ON
Front Panel Access
Setup, Enhanced Resolution
 
[:SENSe]:PIManalyzer:AVERaging FAST|LOWNoise
[:SENSe]:PIManalyzer:AVERaging?
Title
PIM vs. Time Trace Mode.
Description
Low Noise provides improvement in measurement range. Fast provides faster measurement updates (measurements per second). The query returns FAST or LOWN (LOWNoise).
Parameter
FAST|LOWNoise
Range
FAST, Averaging is minimum
LOWNoise, Averaging is maximum
Example
To set the PIM Master measurement for maximum averaging:
:SENSe:PIManalyzer:AVERaging LOWNoise
Front Panel Access
Setup, Trace Mode
 
[:SENSe]:PIManalyzer:AUTorange OFF|ON|0|1
[:SENSe]:PIManalyzer:AUTorange?
Title
PIM vs. Time Auto Range
Description
Sets and queries the Auto Range condition for PIM vs. Time measurement.
Parameter
None
Range
0|OFF, Auto Ranging is OFF
1|ON, Auto Ranging is ON
Example
To set the PIM Master for Auto Range:
:SENSe:PIManalyzer:AUTorange 1
Front Panel Access
Amplitude, Auto Range
 
[:SENSe]:PIManalyzer:FREQuency:F1|F2
[:SENSe]:PIManalyzer:FREQuency:F1|F2?
Title
PIM Master Frequency Setup
Description
Sets the PIM vs. Time frequencies, calculates the IMx Order frequency, and sets the instrument to display the corresponding IMx frequency.
Parameter
Frequency in MHz
Default Unit
Hz
Ranges
PIM Master option dependent (for a list of ranges, refer to Table: PIM Master Carrier Bands and Frequencies)
Example
To set the PIM vs. Time frequency F2 for 1990 MHz:
:SENSe:PIManalyzer:FREQuency:F2 1990000000
Front Panel Access
Freq, Carrier F1|Carrier F2
 
[:SENSe]:PIManalyzer:IMD:ORDer
[:SENSe]:PIManalyzer:IMD:ORDer?
Title
Measurement Receiver / PIM Master Intermodulation Distortion (IMD) Order Setup
Description
Sets the measurement receiver center frequency to receive one of the following IMDs from the PIM Master measurement system: 3/5/7. The query command returns the possible strings “3rd”, “5th”, and “7th” depending on the current selection of IMD Order.
Default Value
3
Range
3, 5, and 7 are the only acceptable values.
Example
To set the measurement receiver for the 5th order IMD:
:SENSe:PIManalyzer:IMD:ORDer 5
Sets the measurement receiver center frequency to the 5th order IMD.
Front Panel Access
Freq, Intermod Order
 
[:SENSe]:PIManalyzer:IMFReq:BAND LOW|HIGH
[:SENSe]:PIManalyzer:IMFReq:BAND?
Title
Option MW82119A‑0700 High/Low IMD frequency band selection
Description
Sets the receive IMD measurement frequency band only for the LTE‑based 700 MHz options. The command and the query are ignored by MW82119A PIM Masters with other instrument options.
Parameter
LOW|HIGH
Default Value
None
Example
To set the measurement receiver to High Band:
:SENSe:PIManalyzer:IMFReq:BAND HIGH
Front Panel Access
PIM vs. Time and Swept PIM:
 Freq, High/Low Band Select
Distance‑to‑PIM:
 Distance, High/Low Band Select
 
[:SENSe]:PIManalyzer:MEASure:STATus?
Title
PIM Analyzer Current Measurement Status
Description
The query returns the current measurement status of the measurement receiver/PIM Analyzer measurement system. Because the measurement system can run for the maximum time duration, the query serves to indicate if the measurement is still in process. This command works with any of the available measurement modes: PIM vs. Time, DTP, and Swept PIM.
Range
0, Measurement is OFF
1, Measurement is ON
[:SENSe]:PIManalyzer:MEASure:VALue?
Title
Measurement Receiver Measured Value From PIM Master Measurement Setup
Description
The query returns an intermodulation measured value in both dBc and dBm.
Parameter:
<amplitude>
Default Unit
dBc/dBm
Front Panel Access
N/A. Intermodulation distortion value is displayed in the lower measurement box as “PIM”.
 
[:SENSe]:PIManalyzer:MODe PIM|PIMSwp|DTP
[:SENSe]:PIManalyzer:MODe?
Title
PIM Analyzer Mode, Set or Request
Description
Puts the system into either PIM vs. Time (PIM), Swept PIM (PIMSwp), or Distance‑to‑PIM (DTP) measurement mode. The query reports the current system mode.
Parameter
PIM|PIMSwp|DTP
Query Response
PIM (PIM vs. Time)
PIMSwp (Swept PIM)
DTP (Distance‑to‑PIM)
Default Value
None
Example
To set the PIM Analyzer mode to measure Distance‑to‑PIM:
:SENSe:PIManalyzer:MODe DTP
To query the state of the PIM Analyzer system:
:SENS:PIM:MOD?
Front Panel Access
Measurements, PIM|Distance-to-PIM|Swept PIM
 
[:SENSe]:PIManalyzer:OUTPut:POWer
[:SENSe]:PIManalyzer:OUTPut:POWer?
Title
PIM Analyzer Output Power
Description
Sets the deliverable output (in dBm) from the PIM Analyzer to the system under test.
Parameter
dBm input with 0.1 dBm resolution
Default Value
None, returns error
Range
25 dBm to 46 dBm
Example
To set the PIM Analyzer output power to 43.5 dBm:
:SENSe:PIManalyzer:OUTPut:POWer 43.5
Front Panel Access
PIM vs. Time and Swept PIM: Setup, PIM Aid, Output Power
PIM vs. Time and Swept PIM: Freq, PIM Aid, Output Power
DTP: Distance, DTP Aid, Output Power
DTP: Setup, DTP Aid, Output Power
 
[:SENSe]:PIManalyzer:RF:OUTput ON|OFF
[:SENSe]:PIManalyzer:RF:OUTput?
Title
PIM vs. Time – Set RF Out On/Off
Description
This controls whether the RF is ON during a measurement, which corresponds to Normal and Noise Floor PVT measurements.
Parameter
ON|OFF
Default Value
ON|OFF
Range
25 dBm to 46 dBm
Example
To turn on RF:
:SENSe:PIManalyzer:OUTPut:POWer 43.5
Front Panel Access
Measurements, PIM vs. Time, Test
 
[:SENSe]:PIManalyzer:SWEEp:SPAN:TYPE IM|RX
[:SENSe]:PIManalyzer:SWEEp:SPAN:TYPE?
Title
Swept PIM Analyzer Sweep Span View
Description
In Swept PIM measurement mode, sets the Frequency Span to either the licensed range associated with the option (IM) or to the widest span capability between the Fixed F1 and Fixed F2 displayed sweeps (RX). The query command returns either IM or RX.
Parameter
IM|RX
Example
To set the span view to IM:
:SENSe:PIManalyzer:SWEEp:SPAN:TYPE IM
Front Panel Access
Freq, Span Type
[:SENSe]:PIManalyzer:SWEEp:IMD:ORDer
[:SENSe]:PIManalyzer:SWEEp:IMD:ORDer?
Title
PIM Analyzer Swept PIM Intermodulation Distortion (IMD) Order Setup
Description
Sets the measurement receiver center frequency to receive one of the following IMDs from the PIM Analyzer measurement system: 3/5/7. The query command returns the possible strings “3rd”, “5th”, and “7th”, depending on the current selection of IMD Order.
Default Value
3
Range
3, 5, and 7 are the only acceptable values.
Example
To set the measurement receiver center frequency to the 5th order IMD:
:SENSe:PIManalyzer:SWEEp:IMD:ORDer 5
Front Panel Access
Freq, Intermod Order
 
[:SENSe]:PIManalyzer:TESt:DURation
[:SENSe]:PIManalyzer:TESt:DURation?
Title
PIM Test Measurement Test Duration
Description
Sets the amount of time in seconds the PIM Master will be on for intermodulation distortion measurements.
Parameter
<time>
Default Value
20
Default Unit
Seconds
Range
1.0 to 1200.0 seconds
Example
To set the test duration time to 5 seconds:
:SENSe:PIManalyzer:TESt:DURation 5.0
Front Panel Access
Setup, Test Duration