VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :SENSe{1-16}:CORRection:DISPlay Subsystem – 2-Port VNAs
 
:SENSe{1-16}:CORRection:DISPlay Subsystem – 2-Port VNAs
:SENSe{1-16}:CORRection:DISPlay:RSWeep[:STATe] <char>
:SENSe{1-16}:CORRection:DISPlay:RSWeep[:STATe]?
Description
Only applies when full 2-port cal is present.
Sets the on/off state of displaying the data when sweeping reverse sweep only on the indicated channel.
Returns the on/off status of displaying the data when sweeping reverse sweep only on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:SENSe1:CORR:DISP:RSW 1
:SENSe1:CORR:DISP:RSW?