VectorStar™ MS464xB Series Microwave Vector Network Analyzer User Interface Reference : Response Menus – 4-Port VNAs : Noise Figure Response Menu (Option 48)
 
Noise Figure Response Menu (Option 48)
NF RESPONSE Menu – 4-Port VNAs (Option 48)
Full Name
Differential Noise Figure Response
Previous
Primary Response Menus – 4-Port VNAs
Prerequisites
The VNA must be equipped with Option 48 – Differential Noise Figure.
The application mode must be set to Noise Figure on the APPLICATION menu.
MAIN | Application | APPLICATION
Differential Noise Figure (Option 48)
Navigation
MAIN | Response | RESPONSE | Noise Figure | NF RESPONSE
NF RESPONSE (Noise Figure Response) Menu (4-port DUT)
Standard Mode
3739 Mode
Note 
All of the noise figure response types are real only. Keep this in mind when selecting graph types. Noise figure and noise power quantities are computed using a 10 × log10 convention when a log mag graph type is selected.
Button Selection Groups
The top five buttons on the NF RESPONSE menu form a button selection group where selection of any one button de-selects all other buttons.
The Configuration buttons on the NF RESPONSE menu form a button selection group where selection of any one button de-selects all other buttons. These buttons are only present if the DUT type is 4-port.
Noise Figure
Select sets the noise figure measurement to Noise Figure. Noise figure is measured using the “cold source” technique. The units are set by the graph type.
Available Gain
Displays the available gain (in a static display, as it is based on loaded DUT S-parameters) of the DUT. This does include match effects.
Insertion Gain
Displays the insertion gain (in a static display, as it is based on loaded DUT S-parameters) of the DUT. This does not include match effects.
Noise Temperature
Temperature Kelvin. Read-only display of the IEEE default temperature of 290 K. The noise temperature is calculated from the noise figure. For reference:
290 K = 17 ºC (62.6 ºF)
0 K = −273.15 °C (−459.67 °F)
Noise Power
Select sets the noise figure measurement to Noise Power trace response. The measurement units are set by the display graph type. The units will be in absolute power when a receiver calibration is applied.
Configuration (Only available for 4-port DUT)
Each of the above parameters requires a port/receiver definition (with a 4-port DUT) and these selections are either single-ended (into VNA b1 or b2 receivers), differential, or common-mode. If the DUT type is 2-port, the response variable will be assigned to the current receiver path.