BTS Master Programming Manual : All Mode Commands : :STATus Subsystem
 
:STATus Subsystem
The commands in this subsystem relate to the current operating state of the instrument.
:STATus:OPERation?
Title
Query Operation Status
Description
This command requests information about the current status of the instrument. Each bit of the return value represents some operation. Only a subset of the bits are implemented for each application. The number returned is the decimal representation of the bit‑wise OR of the enabled bits:
Bit
Decimal Value
Description
0
1
Not implemented
1
2
Not implemented
2
4
Not implemented
3
8
Not implemented
4
16
Not implemented
5
32
Not implemented
6
64
Not implemented
7
128
Not implemented
8
256
Sweep Complete
This bit is set to 0 when the command :INITiate[:IMMediate] is sent to trigger a sweep. It will have a value of 1 when the sweep has completed.
9
512
Not implemented
10
1024
Not implemented
11
2048
Not implemented
12
4096
Not implemented
13
8192
Not implemented
14
16384
Not implemented
15
0
Will always be 0