Line Sweep Tools™ : Measurement Tools : Marker Toolbar & Functions (S331L, S331P)
 
Marker Toolbar & Functions (S331L, S331P)
The Marker toolbar allows you to activate markers within the measurement plot, open the Plot Properties dialog to edit marker properties manually, and toggle the marker table for viewing in the measurement plot. The S331L and S331P has two Marker Style windows to setup markers – Default and Advanced. The select button is located at the bottom of the window titled Marker Style. If Advanced is selected the setup features replicate those of the Advance Marker setups in a VNA Master instrument.
Note 
Any markers added to the measurement using the Advance Marker feature in LST cannot be loaded back up to the instrument.
Advanced Marker settings can be changed to Default. Reference markers will retain their settings while Delta makers are converted into Reference markers. Any peak or valley information will be removed. And the current location of the former delta marker is placed in the Distance window.
Click Plot Properties in the Edit menu or the Marker Edit button on the Marker Toolbar to open the Plot Properties dialog with the Markers tab selected.
Activate a Marker
Press the desired marker button, Marker 1-8, to display the markers in the measurement display. The marker is a vertical line with a diamond that resides on the trace.
Move a Marker
Place the cursor on the desired marker, press the left mouse button, and drag it to the desired frequency.
Display the Marker Table
Press the Marker Table Toggle button to toggle on the db/freq values for the active markers.
Edit Markers using the Plot Properties Dialog
Use the Plot Properties dialog to move a marker to the highest peak, to the lowest valley, edit its frequency manually, or activate the associated delta marker. Delta Makers are useful in calculating the bandwidth of a signal or finding the dB/VSWR difference between two data points. The difference between the reference marker and delta marker is displayed when the marker table is enabled.
Plot Properties VNA Marker Dialog
Instructions
1. Click on a window to enlarge.
2. Press the Marker Edit button in the Marker toolbar to open the Plot Properties dialog with the Markers tab active. Or, place the cursor over the graph and click the right mouse button.
3. Click on Markers… in the menu.
Activate a Marker
1. Press the desired marker buttons, M1-M8, to activate a marker.
• The pressed button’s color changes and the TYPE, PEAK, VALLEY, BETWEEN, and FREQ window become active.
Activate and Set Reference Markers
1. Press the desired marker button.
• The button changes color.
2. Enter a frequency in the associated entry window of the FREQ column.
3. Press Apply and the Reference marker will move to that frequency in the display.
Activate and Set Delta Markers
Markers 2-8 can be set as delta markers to reference Marker 1.
1. Press the associated delta marker button.
• The button changes color.
2. Enter a frequency in the delta frequency entry window.
3. Press Apply
• The delta marker will move to that delta frequency position relative to Marker 1.
Marker Type
• Reference markers display absolute frequency.
• Delta markers display frequency relative to Marker 1.
• Marker 2- Marker 8 can be set as a reference marker or a delta marker. If set as a delta marker, Marker 1 will be the reference marker.
Move Marker to Peak
Using the default BETWEEN setting, Marker 5 and 7 can be used to find either the peak or valley between Markers 1 and 2. Marker 6 and 8 can be used to find either the peak or valley between Markers 3 and 4.
1. Press the PEAK button associated with the desired active marker and the frequency of the highest peak value of the measurement plot will be entered into the FREQ window.
2. Press Apply and the marker will move to that peak position on the active measurement plot.
Move Marker to Valley
Using the default BETWEEN setting, Marker 5 and 7 can be used to find either the peak or valley between Markers 1 and 2. Marker 6 and 8 can be used to find either the peak or valley between Markers 3 and 4.
1. Press the VALLEY button associated with the desired active marker and the lowest valley value of the measurement plot will be entered into the FREQ window.
2. Press Apply and the marker will move to that valley position.