PIM Master™ Passive Intermodulation Analyzer with Site Master™ Cable & Antenna Analyzer Option Programming Manual : All Modes Programming Commands : :INSTrument Subsystem
 
:INSTrument Subsystem
One instrument may contain many logical instrument “modes”. This subsystem controls the selection of the current instrument mode.
:INSTrument:CATalog:FULL?
Title
Query Available Modes
Description
Returns a comma-separated list of available modes. Mode names are enclosed in double quotes (“ “). Immediately following the string name is the application number. For example, an instrument with the High Accuracy Power Meter (Option 19) would return the string: “HI_PM”10,”MINIPIM”46. And an instrument with the Site Master Option 331 and the High Accuracy Power Meter (Option 19) would return the string: “VNA”2,“HI_PM”10,“MINIPIM”46.
Front Panel Access
Shift‑Mode (9), or Menu
 
:INSTrument:NSELect <integer>
:INSTrument:NSELect?
Title
Select Mode by Number
Description
Sets the instrument mode based on the value of <integer>. The query version returns the number associated with the current mode. Use :INSTrument:CATalog:FULL? to get a list of available mode names and their integer representations.
Parameter
<integer>
Parameter Type
<integer>
Related Command
:INSTrument:CATalog:FULL?
:INITiate:CONTinuous
:INSTrument[:SELect]
:STATus:OPERation?
Front Panel Access
Shift‑Mode (9), or Menu
Note 
Switching modes can take longer than 80 seconds, depending on the application. Add a delay of at least 90 seconds between mode switch commands. Anritsu Company advises you to set the remote PC time‑out to 120 seconds in order to avoid unexpected time‑out errors.
:INSTrument[:SELect] <string>
:INSTrument[:SELect]?
Title
Select Mode by Name
Description
Sets the instrument mode based on the mode name specified by <string>. Enclose the <string> argument in single or double quotes. The query version returns the name of the current mode. Use :INSTrument:CATalog:FULL? to get a list of available modes.
Parameter
<string>
Related Command
:INSTrument:CATalog:FULL?
:INSTrument:NSELect
Front Panel Access
Shift‑Mode (9), or Menu
Note 
Switching modes can take longer than 80 seconds, depending on the application. Add a delay of at least 90 seconds between mode switch commands. Anritsu Company advises you to set the remote PC time‑out to 120 seconds in order to avoid unexpected time‑out errors.