PIM Masterâ„¢ Passive Intermodulation Analyzer with Site Masterâ„¢ Cable & Antenna Analyzer Option Programming Manual : PIM Analyzer Programming Commands : :INITiate Subsystem
 
:INITiate Subsystem
The commands in this subsystem control the triggering of measurements.
:INITiate:PIManalyzer:MEASure OFF|ON|0|1
Title
Trigger PIM Analyzer Measurement
Description
This command triggers the start of the PIM Master measurement to measure intermodulation distortion generated between the PIM Analyzer and the system under test. It works with any of the available measurement types: PIM vs. Time, DTP, Swept PIM. The measurement will continue until the defined test duration time has elapsed.
Sending this command with the OFF parameter or the 0 parameter cancels the measurement.
Parameter
OFF|ON|0|1
Parameter Type
<boolean>
Front Panel Access
Measurements, Test (Measure underlined)
 
:INITiate:PIManalyzer:PVT:ALLPower:CAL
Title
PvT All Power Levels Calibration
Description
This command triggers the start of the PIM vs. Time calibration for all power levels.
Front Panel Access
Shift-Cal (2), Custom Calibrations, Start Cal
(full name: Start Cal PIM vs. Time Only All Power Levels)
 
:INITiate:PIManalyzer:RESidual:CAL
Title
PIM Calibration: Current Mode Only
Description
This command triggers the start of the residual PIM calibration for the current measurement mode only. The command only performs the residual PIM calibration and not a full DTP calibration. A PIM load must be attached before sending the command.
Front Panel Access
None