PIM Master™ Passive Intermodulation Analyzer with Site Master™ Cable & Antenna Analyzer (331) and 2-Port Options (703) User Guide : Quick Start Guide : Cable & Antenna Analyzer Measurements (Option 331) : Calibrate with OSL Calibration
 
Calibrate with OSL Calibration
Note 
Refer to the Cable & Antenna Measurement Guide listed in Other Documents for calibration details.
1. Press the Freq/Dist main menu key and enter the appropriate frequency range
2. Press Shift then Calibrate (2) key.
3. Select Standard or FlexCal.
4. Press Start Cal and follow instructions on screen.
5. Connect Open to RF Out and press the Enter key.
6. Connect Short to RF Out and press the Enter key.
7. Connect Load to RF Out and press the Enter key.
8. Verify that the calibration has been properly performed by checking that the Cal Status message is now displaying “ON, Standard” or “ON, FlexCal”.
Calibration Setup with OSL Cal Standards