Field Master Pro™ MS2090A User Guide : Instrument Notifications : Self-Test Messages
 
Self-Test Messages
Field Master Pro has built-in diagnostics that allow you to run a built-in self test. The test results can be viewed on the screen and saved to a log file. To run self test, use the 3-line menu to access DIAGNOSTICS, then select SELF TEST and RUN SELF TEST. Below is a sample list of self tests with passing results:
TEST DATE: 2019-03-21 09:51:47 MDT
 
OVERALL STATUS : PASS
 
SYSTEM : OVERALL STATUS : PASS
System voltage : 13.00 V, 4.00 V ,15.01 V (PASS)
System power : 32.2083 W
External current : 2.50 A, 2.50 A ,2.32 A (PASS)
Battery current : 4.00 A, 4.00 A ,0.18 A (PASS)
+0.85V LP core supply : 0.85 V, 0.09 V ,0.85 V (PASS)
+3.3V LP supply : 3.30 V, 0.33 V ,3.25 V (PASS)
+1.8V LP Vaux supply : 1.80 V, 0.18 V ,1.82 V (PASS)
+1.8V LP IO supply : 1.80 V, 0.18 V ,1.81 V (PASS)
+0.85V FP core supply : 0.85 V, 0.09 V ,0.85 V (PASS)
+1.8V DDR PLL supply : 1.80 V, 0.18 V ,1.81 V (PASS)
+0.85V PS transceiver AVCC : 0.85 V, 0.09 V ,0.85 V (PASS)
+1.8V PS transceiver AVTT : 1.80 V, 0.18 V ,1.83 V (PASS)
+2.5V DDR SODIMM supply : 2.50 V, 0.25 V ,2.49 V (PASS)
+1.2V DDR SODIMM supply : 1.20 V, 0.12 V ,1.20 V (PASS)
+0.85V FPGA core supply : 0.85 V, 0.09 V ,0.86 V (PASS)
+0.9V transceiver AVCC supply : 0.90 V, 0.09 V ,0.90 V (PASS)
+1.2V transceiver AVTT supply : 1.20 V, 0.12 V ,1.21 V (PASS)
+1.0V Ethernet supply : 1.00 V, 0.10 V ,0.99 V (PASS)
+1.8V FPGA supply : 1.80 V, 0.18 V ,1.79 V (PASS)
+2.5V FPGA supply : 2.50 V, 0.25 V ,2.48 V (PASS)
+3.3V FPGA supply : 3.30 V, 0.33 V ,3.29 V (PASS)
+5.0V USB/misc supply : 5.00 V, 0.50 V ,5.00 V (PASS)
+3.9V analog supply : 3.90 V, 0.39 V ,3.96 V (PASS)
+5.7V analog supply : 5.70 V, 0.57 V ,5.67 V (PASS)
-5.7V analog supply : -5.70 V, 0.57 V ,-5.26 V (PASS)
+12.8V analog supply : 12.80 V, 1.28 V ,12.82 V (PASS)
+24.0V analog supply : 24.00 V, 2.40 V ,23.92 V (PASS)
Backlight voltage : 22.00 V, 4.40 V ,19.56 V (PASS)
Backlight current : 0.20 A, 0.20 A ,0.18 A (PASS)
12.0V for fan : 12.00 V, 1.20 V ,12.00 V (PASS)
CPU Temperature : 61.0 C
Motherboard Temperature : 46.5 C
Fan speed, percent of full speed : 7 %
Fan speed : 2638 RPM
Temperature at Reference : 45.3 C
 
SPA : OVERALL STATUS : PASS
+12VG : 12.00 V, 1.20 V ,11.90 V (PASS)
+5V_PREAMP : 5.00 V, 0.50 V ,4.99 V (PASS)
+5VA : 5.00 V, 0.50 V ,5.02 V (PASS)
+5VG : 5.00 V, 0.50 V ,4.99 V (PASS)
+5VH : 5.00 V, 0.50 V ,5.02 V (PASS)
+5VI : 5.00 V, 0.50 V ,5.05 V (PASS)
+5V_VCO : 5.00 V, 0.50 V ,5.09 V (PASS)
+5VM : 5.00 V, 0.50 V ,5.01 V (PASS)
+5VP : 5.00 V, 0.50 V ,5.05 V (PASS)
+5V_SAW : 5.00 V, 0.50 V ,4.99 V (PASS)
+3V3_HBPA : 3.30 V, 0.33 V ,3.31 V (PASS)
+3V3_IF : 3.30 V, 0.33 V ,3.29 V (PASS)
+3V6 : 3.60 V, 0.36 V ,3.55 V (PASS)
+3V3_CAL : 3.30 V, 0.33 V ,3.36 V (PASS)
+3V3G : 3.30 V, 0.33 V ,3.33 V (PASS)
+3V3H : 3.30 V, 0.33 V ,3.29 V (PASS)
+3V3M : 3.30 V, 0.33 V ,3.32 V (PASS)
+3V3P : 3.30 V, 0.33 V ,3.33 V (PASS)
ADC_AVDD3_3V3 : 3.30 V, 0.33 V ,3.33 V (PASS)
+3V3IFA : 3.30 V, 0.33 V ,3.33 V (PASS)
+3V3DET : 3.30 V, 0.33 V ,3.35 V (PASS)
+3V3IFC : 3.30 V, 0.33 V ,3.37 V (PASS)
+3V3IFD : 3.30 V, 0.33 V ,3.35 V (PASS)
+3V3IF_OUT : 3.30 V, 0.33 V ,3.33 V (PASS)
SPARE : 0.00 V, 1.00 V ,0.00 V (PASS)
-5V : -5.00 V, 0.50 V ,-5.13 V (PASS)
+24VH : 24.00 V, 2.40 V ,24.18 V (PASS)
+3V3D : 3.30 V, 0.33 V ,3.27 V (PASS)
+2V5D : 2.50 V, 0.25 V ,2.50 V (PASS)
ADC_AVDD2_2V5 : 2.50 V, 0.25 V ,2.50 V (PASS)
ADC_AVDD1_1V25 : 1.25 V, 0.13 V ,1.25 V (PASS)
ADC_DRVDD_1V25 : 1.25 V, 0.13 V ,1.25 V (PASS)
DCM Not Available For This Option : 0.00 V, 0.00 V ,0.00 V (PASS)
Thermal Sensor 1 (Mid-band 1st IF) : 35.00 C, 90.00 C ,44.75 C (PASS)
Thermal Sensor 2 (Mid-band 2nd IF) : 35.00 C, 90.00 C ,46.00 C (PASS)
Thermal Sensor 3 (Low-band RF) : 35.00 C, 90.00 C ,42.00 C (PASS)
DCM Temperature Not Available For This Option : 0.00 C, 0.00 C ,0.00 C (PASS)
ADC Thermal Sensor : 35.00 C, 90.00 C ,39.93 C (PASS)
If any self test fails, try resetting the instrument with a RESET ALL and reboot, and run the test again. If the condition persists, contact your local Anritsu Service Center (http://www.anritsu.com/contact-us).