Rubidium™ MG362x1A SeriesLow-Noise RF/Microwave SignalGenerators : Error Messages : Parser Errors
 
Parser Errors
An <error code> in the range [201, 212] is generated by the instrument’s parser in response to the error condition described.
 
Error Code
Error Description [description/explanation/examples]
201
“Query only”
Indicates the command is a query command only.
202
“No query allowed”
Indicates that a query form of the command is not available.
203
“Parameter(s) not expected”
Indicates that the command does not take any parameter(s).
204
“Constant not allowed in STATUS subsystem”
207
“Enumeric value not in union”
Indicates an illegal parameter was sent with the command.
208
“Illegal number of parameters”
Indicates an illegal number of parameters was sent with the command.
210
“Run out of memory handle”
Indicates an internal parser problem.
211
“Unit not matched”
Indicates the parameter unit does not match any defined unit for this parameter.
212
“Unit not required”
Indicates no unit is required with this parameter.