Rubidiumâ„¢ MG362x1A SeriesLow-Noise RF/Microwave SignalGenerators : Instrument Overview : Installation and Power Requirements
 
Installation and Power Requirements
The signal generator accepts 85 to 264 Vac, 48 to 440 Hz, single-phase power. Power consumption is 250 VA maximum. The signal generator is intended for Installation Category (Over Voltage Category) II. To connect the MG362x1A to the power source, plug the female end of the power cable into the input line voltage receptacle on the rear panel as shown in Figure: Rear Panel Power Connection. Then plug the male end of the power cord into a three-wire power line outlet. Turn on the rear panel power button. This automatically places the signal generator in operation (front panel Operate LED on).
Warning
When supplying power to this equipment, always use a three-wire power cable connected to a three-wire power line outlet. If power is supplied without grounding the equipment in this manner, there is a risk of receiving a severe or fatal electric shock.