This section provides an alphabetical listing of the SCPI and Native commands used to program the series MG362x1A Synthesized Signal Generator.
*IDN?
Description:
Requests device identification.
Returns an identification string in IEEE 488.2 specified <NR1> format (four fields separated by commas). The fields are <manufacturer>, <model>, <serial #>, <firmware version>.
Functional Category
OUTPUT
ACW
Description:
Activates the currently open frequency parameter as CW.
Activates the frequency parameter that is currently open for data entry as CW, switches to CW if not in CW mode.
Functional Category
CW FREQUENCY
Example:
F1, ACW
Related Commands:
CF1
Commands:
L9, XL9
AM0
Description:
Turns off the internal or external AM function.
Functional Category
MODULATION
AM1
Description:
Turns on the external AM function in linear mode.
Disables the internal AM function or the external AM function in Log mode, had either of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
AMS
AM2
Description:
Turns on the external AM function in log mode.
Disables the internal AM function or the external AM function in Linear mode, had either of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
ASD
AM7
Description:
Turns on the internal AM function in linear mode.
Disables the external AM function or the internal AM function in Log mode, had either of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
ADP1, AMR, AMW(x)
AM8
Description:
Turns on the internal AM function in log mode.
Disables the external AM function or the internal AM function in Linear mode, had either of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
ADP2, AMR, AMW(x)
AMI
Description:
Turns on the AM measurement mode.
In this mode, the signal generator measures the voltage of the external modulating signal and calculates the modulation depth of the RF output signal. The measurement results will be sent to the controller upon receipt of the OMM command. When the AMI command is received, measurements will continue to be taken until the mode is exited by receipt of the MOM command.
Functional Category
MEASURE FUNCTION
Output Command:
OAD1
Related Commands:
OMM, MOM
AMO
Description:
Turns off the internal AM function generator if it is on.
Deselects the step attenuator decoupling function. This returns the MG362x1A to its normal (coupled) leveling mode. In this mode, when the desired power level is set, the correct combination of level and step attenuator setting is determined by the instrument firmware.
Functional Category
POWER LEVELING
Related Commands:
AT1, ATT(xx)
AT1
Description:
Selects ALC step attenuator decoupling.
This function decouples the step attenuator (if equipped) from the ALC system. This permits separate setting the ALC level and step attenuator.
Functional Category
POWER LEVELING
Related Commands:
AT0, ATT(xx)
ATT(xx)
Description:
Sets the step attenuator value to xx (times 10 dB).
Permits setting the step attenuator value to xx (times 10 dB) in the ALC step attenuator decoupling mode.
Functional Category
POWER LEVELING
Value:
xx: unsigned integer between 00 (0 dB attenuation) and 11 (110 dB attenuation)
Example:
AT1, ATT08
Decouples the step attenuator from the ALC system and sets the step attenuator value to 80 dB.
Related Commands:
AT0, AT1
AUT
Description:
Selects automatic sweep triggering.
Selects automatic sweep triggering for frequency, power level, or list sweeps. When automatic sweep triggering is selected, the sweep continually sweeps from its start frequency, power level, or list index to its stop frequency, power level, or list index with optimal retrace time.
Functional Category
SWEEP
Example:
SSP; SF1; AUT
Sets step sweep F1-F2 with auto triggering.
Related Commands:
EXT, TEX
BPN
Description:
Selects -3.3V for retrace and bandswitch blanking outputs.
Selects a -3.3V level for the rear panel retrace and bandswitch blanking outputs. The retrace blanking output signal is available at pin six of the AUX I/O connector. The bandswitch blanking output signal is available at pin 20 of the AUX I/O connector.
Functional Category
CONFIGURATION
BPP
Description:
Selects +3.3V for retrace and bandswitch blanking outputs.
Selects a +3.3V level for the rear panel retrace and bandswitch blanking outputs. The retrace blanking output signal is available at pin six of the AUX I/O connector. The bandswitch blanking output signal is available at pin 20 of the AUX I/O connector.
Functional Category
CONFIGURATION
CF0 or CF0 <value> <unit>
Description:
Sets CW mode at F0 and opens the F0 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) F0 frequency and causes F0 to be output. Also permits setting the F0 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Example:
CF0 5.5 GH
Places the MG362x1A at the preset (or previously set) CW F0 frequency and opens the F0 parameter for data entry. Sets the F0 frequency to 5.5 GHz.
Output Command:
OF0
Related Commands:
ACW
CF1 <value> <unit>
Description:
Sets CW mode at F1 and opens the F1 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) F1 frequency and causes F1 to be output. Also permits setting the F1 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF1
Related Commands:
CF1, ACW
CF2 or CF2 <value> <unit>
Description:
Sets CW mode at F2 and opens the F2 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) F2 frequency and causes F2 to be output. Also permits setting the F2 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OAD1
Related Commands:
F2, ACW
CF3 or CF3 <value> <unit>
Description:
Sets CW mode at F3 and opens the F3 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) F3 frequency and causes F3 to be output. Also permits setting the F3 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Related Commands:
F3, ACW
Output Command:
OF3
CF4 or CF4 <value> <unit>
Description:
Sets CW mode at F4 and opens the F4 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) F4 frequency and causes F4 to be output. Also permits setting the F4 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF4
Related Commands:
F4, ACW
CF5 or CF5 <value> <unit>
Description:
Sets CW mode at F5 and opens the F5 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) F5 frequency and causes F5 to be output. Also permits setting the F5 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF5
Related Commands:
F5, ACW
CF6 or CF6 <value> <unit>
Description:
Sets CW mode at F6 and opens the F6 parameter. Places the MG362x1A in CW mode at the preset (or previously set) F6 frequency and causes F6 to be output. Also permits setting the F6 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF6
Related Commands:
F6, ACW
CF7 or CF7 <value> <unit>
Description:
Sets CW mode at F7 and opens the F7 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) F7 frequency and causes F7 to be output. Also permits setting the F7 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF7
Related Commands:
F7, ACW
CF8 or CF8 <value> <unit>
Description:
Sets CW mode at F8 and opens the F8 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) F8 frequency and causes F8 to be output. Also permits setting the F8 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF8
Related Commands:
F8, ACW
CF9 or CF9 <value> <unit>
Description:
Sets CW mode at F9 and opens the F9 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) F9 frequency and causes F9 to be output. Also permits setting the F9 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category:
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF9
Related Commands:
F9, ACW
CLO
Description:
Closes the previously opened parameter.
Functional Category
PARAMETER ENTRY
Example:
F1 2 GH; CLO
CLR
Description:
Clears the data entry.
Functional Category
DATA ENTRY
CM0 or CM0 <value> <unit>
Description:
Sets CW mode at M0 and opens the M0 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) M0 marker frequency and causes M0 to be output. Also permits setting the M0 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Example:
CM0 5.5 GH
Places the MG362x1A at the preset (or previously set) CW M0 marker frequency and opens the M0 parameter for data entry. Sets the M0 frequency to 5.5 GHz.
Output Command:
OM0
Related Commands:
M0, ACW
CM1 or CM1 <value> <unit>
Description:
Sets CW mode at M1 and opens the M1 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) M1 marker frequency and causes M1 to be output. Also permits setting the M1 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM1
Related Commands:
M1, ACW
CM2 or CM2 <value> <unit>
Description:
Sets CW mode at M2 and opens the M2 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) M2 marker frequency and causes M2 to be output. Also permits setting the M2 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM2
Related Commands:
M2, ACW
CM3 or CM3 <value> <unit>
Description:
Sets CW mode at M3 and opens the M3 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) M3 marker frequency and causes M3 to be output. Also permits setting the M3 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM3
Related Commands:
M3, ACW
CM4 or CM4 <value> <unit>
Description:
Sets CW mode at M4 and opens the M4 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) M4 marker frequency and causes M4 to be output. Also permits setting the M4 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM4
Related Commands:
M4, ACW
CM5 or CM5 <value> <unit>
Description:
Sets CW mode at M5 and opens the M5 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) M5 marker frequency and causes M5 to be output. Also permits setting the M5 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM5
Related Commands:
M5, ACW
CM6 or CM6 <value> <unit>
Description:
Sets CW mode at M6 and opens the M6 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) M6 marker frequency and causes M6 to be output. Also permits setting the M6 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM6
Related Commands:
M6, ACW
CM7 or CM7 <value> <unit>
Description:
Sets CW mode at M7 and opens the M7 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) M7 marker frequency and causes M7 to be output. Also permits setting the M7 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM7
Related Commands:
M7, ACW
CM8 or CM8 <value> <unit>
Description:
Sets CW mode at M8 and opens the M8 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) M8 marker frequency and causes M8 to be output. Also permits setting the M8 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Related Output Command:
OM8
Commands:
M8, ACW
CM9 or CM9 <value> <unit>
Description:
Sets CW mode at M9 and opens the M9 parameter.
Places the MG362x1A in CW mode at the preset (or previously set) M9 marker frequency and causes M9 to be output. Also permits setting the M9 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
CW FREQUENCY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM9
Related Commands:
M9, ACW
CSB
Description:
Clears all Native status bytes.
Functional Category
STATUS
CTL
Description:
Copy current CW frequency and power level to current list index.
The command copies the current CW frequency and power level to the current list index. This is another method of entering frequency and power level information into the current list index. The MG362x1A must be list sweep mode and the desired list selected and list index set before invoking this command. The command, LST, places the unit in list sweep mode; the command, ELN(x), is used to select which of the four lists is to be accessed; and the command, ELI(xxxx), sets the list index for the current list.
Functional Category
LIST SWEEP
Example:
LST; ELN2; ELI1023; CTL
Places the MG362x1A in list sweep mode, selects list number 2, sets the list index to 1023, and copies the current CW frequency and power level to list index 1023.
Related Commands:
LST, ELN(x), ELI(xxxx)
D1 <value> <unit>
Description:
Opens the internal pulse delay1 parameter.
Functional Category
PARAMETER ENTRY
Value:
30 ns to 167 ms at 100 MHz pulse clock rate; 100 ns to 419 ms at 40 MHz pulse clock rate; 300 ns to 1.6 s at 10 MHz pulse clock rate
Units:
SEC, MS, US, NS
Example:
PTG3; D1 100 US; IP
Turns on internal pulse trigger delayed by 100 μs.
Output Command:
OD1 (μs)
Related Commands:
IP, PMD(x), PTG(x)
D2 <value> <unit>
Description:
Opens the internal pulse delay2 parameter.
Functional Category
PARAMETER ENTRY
Value:
30 ns to 167 ms at 100 MHz pulse clock rate; 100 ns to 419 ms at 40 MHz pulse clock rate; 300 ns to 1.6 s at 10 MHz pulse clock rate
Units:
SEC, MS, US, NS
Example:
PMD2; D2 100 US; IP
Turns on internal pulse doublet with a 100 μs delay before the 2nd pulse.
Output Command:
OD2 (μs)
Related Commands:
IP, PMD(x), PTG(x)
D3 <value> <unit>
Description:
Opens the internal pulse delay3 parameter.
Functional Category
PARAMETER ENTRY
Value:
30 ns to 167 ms at 100 MHz pulse clock rate; 100 ns to 419 ms at 40 MHz pulse clock rate; 300 ns to 1.6 s at 10 MHz pulse clock rate
Units:
SEC, MS, US, NS
Example:
PMD3; D3 100 US; IP
Turns on internal pulse triplet with a 100 μs delay before the 3rd pulse.
Output Command:
OD3 (μs)
Related Commands:
IP, PMD(x), PTG(x)
D4 <value> <unit>
Description:
Opens the internal pulse delay4 parameter.
Functional Category
PARAMETER ENTRY
Value:
30 ns to 167 ms at 100 MHz pulse clock rate; 100 ns to 419 ms at 40 MHz pulse clock rate; 300 ns to 1.6 s at 10 MHz pulse clock rate
Units:
SEC, MS, US, NS
Example:
PMD4; D4 100 US; IP
Turns on internal pulse quadruplet with a 100 μs delay before the 4th pulse.
Output Command:
OD4 (μs)
Related Commands:
IP, PMD(x), PTG(x)
DF5
Description:
Selects F5-delta frequency sweep mode.
Selects a symmetrical frequency sweep around F5. The width of the sweep is determined by the Delta Frequency parameter.
Functional Category
SWEEP
Example:
SSP; DF5; AUT
Sets step sweep F5-ΔF with auto triggering.
Related Commands:
F5, DLF, DFF, DFM
DF6
Description:
Selects F6-delta frequency sweep mode.
Selects a symmetrical frequency sweep around F6. The width of the sweep is determined by the Delta Frequency parameter.
Functional Category
SWEEP
Example:
SSP; DF6; AUT
Sets step sweep F6-ΔF with auto triggering.
Related Commands:
F6, DLF, DFF, DFM
DFF <value> <unit>
Description:
Opens the delta frequency parameter.
Permits setting the Delta Frequency parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model. The commands, DFM and DLF, are the same as DFF.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Example:
F1 3 GH; DFF 100 MH; DF1; SSP; AUT
Sets auto trigger step sweep F1-ΔF 3 GHz - 50 MHz to 3 GHz + 50 MHz.
Output Command:
ODF returns the DF frequency value (in MHz) to the controller.
DFM <value> <unit>
Description:
Opens the delta frequency parameter.
Permits setting the Delta Frequency parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model. The commands, DFF and DLF, are the same as DFM.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Example:
F1 3 GH; DFM 100 MH; DF1; SSP; AUT
Sets auto trigger step sweep F1-ΔF 3GHz-50MHz to 3GHz+50MHz.
Output Command:
ODF returns the DF frequency value (in MHz) to the controller.
DLF <value> <unit>
Description:
Opens the delta frequency parameter.
Permits setting the Delta Frequency parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model. The commands, DFF and DFM, are the same as DLF.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Example:
F1 3 GH; DLF 100 MH, DF1; SSP; AUT
Sets auto trigger step sweep F1-ΔF 3 GHz - 50 MHz to 3 GHz + 50 MHz.
Output Command:
ODF returns the DF frequency value (in MHz) to the controller.
DPT
Description:
Internal triggered pulse with delay.
Sets the internal pulse mode to single, the internal pulse trigger to triggered with delay, and turns on the internal pulse modulation function. Disables the external pulse modulation function, if previously programmed.
Functional Category
MODULATION
Example:
W1 50 US; D1 200 US; DPT
(50 μs pulse after 200 μs delay from trigger falling edge)
Related Commands:
IP, PTG5, PMD1, PTF, W1, D1
DS0
Description:
Turns on the secure mode.
In the secure mode, the front panel display of all frequency, power level, and modulation parameters is disabled. The instrument continues to function normally in all other respects.
Functional Category
MISCELLANEOUS
DS1
Description:
Turns off the secure mode.
Restores the front panel display of all frequency, power level, and modulation parameters.
Functional Category
MISCELLANEOUS
EL0
Description:
Inhibits updating of the ESB2 bit 4.
Blocks updating of the Extended Status Byte two bit four (RF Unlocked). This is the default setting because it is normal for the RF to be momentarily unlocked during sweeps and sweep retrace.
Functional Category
STATUS
Related Commands:
SB0, SQ0
EL1
Description:
Enables RF Unlocked bit update. ESB2 bit 4.
Unblocks updating of the Extended Status Byte bit four (RF Unlocked).
In list sweep mode, permits setting the list index for the current list.
Functional Category
LIST SWEEP
Value:
xxxx: 4-digit integer between 0000 and 1999
Example:
LST; ELN3; ELI1263
Places the MG362x1A in list sweep mode, selects list number 3, and sets the list index to 1263.
Related Commands:
LST, ELN(x)
ELN(x)
Description:
Sets the list number to x.
In list sweep mode, permits selection of which of the four lists is to be accessed.
Functional Category
LIST SWEEP
Value:
x: 1-digit integer between 0 and 3
EP0
Description:
Selects TTL-low to turn RF on during pulse modulation.
Permits selection of a TTL-low signal to turn the RF on during pulse modulation. If pulse modulation is not installed, this command produces a syntax error.
Functional Category
MODULATION
Related Commands:
EP1, P0, XP
EP1
Description:
Selects TTL-high to turn RF on during pulse modulation.
Permits selection of a TTL-high signal to turn the RF on during pulse modulation. If pulse modulation is not installed, this command produces a syntax error.
Functional Category
MODULATION
Related Commands:
EP0, P0, XP
EXT
Description:
Selects single sweep triggering.
Selects single sweep triggering for frequency, power level, or list sweeps. When single sweep triggering is selected, a single sweep starts when a TRG or TRS command is received. The RSS command resets the sweep to its start frequency, power level, or list index whenever the command is received while a single sweep is in progress.
Functional Category
SWEEP
Example:
SSP; SF1; EXT; TRG
Turns on step sweep F1-F2, single trigger, and trigger one sweep.
Related Commands:
TRG, TRS, RSS
F0 <value> <unit>
Description:
Opens the F0 parameter.
Permits setting the F0 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF0 returns the F0 frequency value (in MHz) to the controller.
F1 <value> <unit>
Description:
Opens the F1 parameter.
Permits setting the F1 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF1 returns the F1 frequency value (in MHz) to the controller.
F2 <value> <unit>
Description:
Opens the F2 parameter.
Permits setting the F2 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF2 returns the F2 frequency value (in MHz) to the controller.
F3 <value> <unit>
Description:
Opens the F3 parameter.
Permits setting the F3 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF3 returns the F3 frequency value (in MHz) to the controller.
F4 <value> <unit>
Description:
Opens the F4 parameter.
Permits setting the F4 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF4 returns the F4 frequency value (in MHz) to the controller.
F5 <value> <unit>
Description:
Opens the F5 parameter.
Permits setting the F5 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF5 returns the F5 frequency value (in MHz) to the controller.
F6 <value> <unit>
Description:
Opens the F6 parameter.
Permits setting the F6 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF6 returns the F6 frequency value (in MHz) to the controller.
F7 <value> <unit>
Description:
Opens the F7 parameter.
Permits setting the F7 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF7 returns the F7 frequency value (in MHz) to the controller.
F8 <value> <unit>
Description:
Opens the F8 parameter.
Permits setting the F8 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF8 returns the F8 frequency value (in MHz) to the controller.
F9 <value> <unit>
Description:
Opens the F9 parameter.
Permits setting the F9 parameter. The frequency setting is determined by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OF9 returns the F9 frequency value (in MHz) to the controller.
FB0
Description:
Inhibits extended status byte 1 SRQ generation.
Inhibits an SRQ from being generated when the Extended Status Byte one bit is set. This is the default mode.
Functional Category
STATUS
Related Commands:
SQ0, FB1
FB1
Description:
Enables extended status byte 1 SRQ generation.
Enables an SRQ to be generated when Primary Status Byte bit zero (Extended Status Byte one) is set and SQ1 has been programmed. The Extended Status Byte one bit is set whenever one (or more) of the unmasked status reporting functions in Extended Status Byte one is true. SQ1 enables the SRQ generation function.
10 kHz to 20 MHz in Locked, Locked Low-Noise, and Unlocked Narrow FM; 100 kHz to 100 MHz in Unlocked Wide FM
Units:
MH, KH
Example:
FM9; FDV 500 KH
Turns on FM locked mode with a deviation of 500 kHz.
Output Command:
OFD (in Hz)
Related Commands:
FM7, FM8, FM9, FMN
FM0
Description:
Turns off the internal or external FM function.
Functional Category
MODULATION
FM1
Description:
Turns on the external FM function in unlocked narrow mode.
The signal generator output is not phase-locked. Disables the internal FM function or the external FM function in Unlocked Wide, Locked, or Locked Low-Noise mode, had any of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
FM0, FM2, FML, FMU, FMW
FM2
Description:
Turns on the external FM function in locked low-noise mode.
The signal generator output is phase-locked. Disables the internal FM function or the external FM function in Unlocked Narrow, Unlocked Wide, or Locked mode, had any of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
FM0, FM1, FML, FMU, FMW
FM7
Description:
Turns on the internal FM function in unlocked narrow mode.
The signal generator output is not phase-locked. Disables the external FM function or the internal FM function in Unlocked Wide, Locked, or Locked Low-Noise mode, had any of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
FM0, FM8, FM9, FMN
FM8
Description:
Turns on the internal FM function in unlocked wide mode.
The signal generator is not phase-locked. Disables the external FM function or the internal FM function in Unlocked Narrow, Locked, or Locked Low-Noise mode, had any of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
FM0, FM7, FM9, FMN
FM9
Description:
Turns on the internal FM function in locked mode.
The signal generator output is phase-locked. Disables the external FM function or the internal FM function in Unlocked Narrow, Unlocked Wide, or Locked Low-Noise mode, had any of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
FM0, FM7, FM8, FMN
FMD
Description:
Turns on the FM measurement mode.
Turns on the FM Measurement mode. In this mode, the signal generator measures the voltage of the external modulating signal and calculates the peak frequency deviation of the RF output signal. The measurement results will be sent to the controller upon receipt of the OMM command. When the FMD command is received, measurements will continue to be taken until the mode is exited by receipt of the MOM command.
Functional Category
MEASURE FUNCTION
Output Command:
OMM
Related Commands:
OMM, MOM
FML
Description:
Turns on the external FM function in locked mode.
The signal generator output is phase-locked. Disables the internal FM function or the external FM function in Unlocked Narrow, Unlocked Wide, or Locked Low-Noise mode, had any of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
FM0, FM1, FM2, FMU, FMW
FMN
Description:
Turns on the internal FM function in locked low-noise mode.
The signal generator is phase-locked. Disables the external FM function or the internal FM function in Unlocked Narrow, Unlocked Wide, or Locked mode, had any of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
FM0, FM7, FM8, FM9
FMR <value> <unit>
Description:
Opens the internal FM rate parameter.
Functional Category
PARAMETER ENTRY
Value:
0.1 Hz to 1 MHz for sine wave; 0.1 Hz to 100 kHz for square, triangle, and ramp waveforms
Units:
MH, KH, HZ
Example:
FM9; FMR 1 KH
Turns on FM internal locked mode with a 1 kHz rate.
Output Command:
OFR (in Hz)
Related Commands:
FM7, FM8, FM9, FMN
FMS <value> <unit>
Description:
Opens the external FM sensitivity parameter.
Functional Category
PARAMETER ENTRY
Value:
±10 kHz/V to ±20 MHz/V in Locked, Locked Low-Noise, and Unlocked Narrow FM; ±100 kHz/V to ±100 MHz/V in Unlocked Wide FM
Units:
KV, MV
Example:
FML; FMS 500 KH
Turns on FM external locked with a sensitivity of 500 kHz/Volt.
Output Command:
OFS (in Hz)
Related Commands:
FM1, FM2, FML, FMU, FMW
FMU
Description:
Turns on the external FM function in unlocked narrow mode (Same as FM1).
The signal generator output is not phase-locked. Disables the internal FM function or the external FM function in Unlocked Wide, Locked, or Locked Low-Noise mode, had any of these modes been previously programmed.
Functional Category
MODULATION
Related Commands:
FM0, FM1, FM2, FML, FMW
FMW
Description:
Turns on the external FM function in unlocked wide mode.
The signal generator output is not phase-locked. Disables the internal FM function or the external FM function in Unlocked Narrow, Locked, or Locked Low-Noise mode, had any of these modes been previously programmed.
Sets the internal pulse mode to single, the internal pulse trigger to gated, and turns on the internal pulse modulation function. Disables the external pulse modulation function, if previously programmed.
Functional Category
MODULATION
Related Commands:
P0, PTG(x), IP
HWT
Description:
Selects external sweep trigger.
Selects external sweep triggering for frequency, power level, or list sweeps. When external sweep triggering is selected , a single sweep occurs when triggered by an external TTL-compatible clock pulse to the rear panel AUX I/O connector.
Functional Category
SWEEP
Related Commands:
AUT, EXT, TEX
IM1
Description:
Turns on the intensity marker mode.
Functional Category
MARKER
Related Commands:
MK0, VM1
IP
Description:
Turns on internal pulse modulation.
Disables the external pulse modulation function, if previously programmed.
Functional Category
MODULATION
Related Commands:
P0
L0 or L0 <value> <unit>
Description:
Sets RF output power level to L0. Opens L0 parameter.
Causes the RF output power level to be set to the preset (or previously set) L0 power level. Also permits setting the L0 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
POWER LEVELING
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Example:
L0 5 DM
Places the RF output power level at the preset (or previously set) L0 power level and opens the L0 parameter for data entry. Sets the L0 power level to 5 dBm.
Output Command:
OL0 returns the L0 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
L1 or L1 <value> <unit>
Description:
Sets RF output power level to L1. Opens L1 parameter.
Causes the RF output power level to be set to the preset (or previously set) L1 power level. Also permits setting the L1 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
POWER LEVELING
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL1 returns the L1 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
L2 or L2 <value> <unit>
Description:
Sets RF output power level to L2. Opens L2 parameter.
Causes the RF output power level to be set to the preset (or previously set) L2 power level. Also permits setting the L2 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
POWER LEVELING
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL2 returns the L2 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
L3 or L3 <value> <unit>
Functional Category
POWER LEVELING
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Description:
Causes the RF output power level to be set to the preset (or previously set) L3 power level. Also permits setting the L3 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Output Command:
OL3 returns the L3 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
L4 or L4 <value> <unit>
Description:
Sets RF output power level to L4. Opens L4 parameter.
Causes the RF output power level to be set to the preset (or previously set) L4 power level. Also permits setting the L4 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
POWER LEVELING
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL4 returns the L4 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
L5 or L5 <value> <unit>
Description:
Sets RF output power level to L5. Opens L5 parameter.
Causes the RF output power level to be set to the preset (or previously set) L5 power level. Also permits setting the L5 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
POWER LEVELING
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL5 returns the L5 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
L6 or L6 <value> <unit>
Description:
Sets RF output power level to L6. Opens L6 parameter.
Causes the RF output power level to be set to the preset (or previously set) L6 power level. Also permits setting the L6 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
POWER LEVELING
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Related Commands:
LOG, LIN, PU0, PU1, PU2
Output Command:
OL6 returns the L6 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
L7 or L7 <value> <unit>
Description:
Sets RF output power level to L7. Opens L7 parameter.
Causes the RF output power level to be set to the preset (or previously set) L7 power level. Also permits setting the L7 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
POWER LEVELING
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL7 returns the L7 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
L8 or L8 <value> <unit>
Description:
Sets RF output power level to L8. Opens L8 parameter.
Causes the RF output power level to be set to the preset (or previously set) L8 power level. Also permits setting the L8 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
POWER LEVELING
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL8 returns the L8 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
L9 or L9 <value> <unit>
Description:
Sets RF output power level to L9. Opens L9 parameter.
Causes the RF output power level to be set to the preset (or previously set) L9 power level. Also permits setting the L9 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
POWER LEVELING
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL9 returns the L9 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
LDT <value> <unit>
Description:
Opens the list sweep dwell time parameter.
Permits setting the dwell-time-per-step of the list sweep.
Functional Category
LIST SWEEP
Value:
1 ms to 99 s
Units:
MS, SEC
Example:
LIB1234; LIE1237; EXT; LDT 10 MS; TRG
Implements a list sweep from current list index 1234 to index 1237 in single trigger mode with a 10 ms dwell-time-per-step, then triggers a single sweep.
Related Commands:
LIB(xxxx), LIE(xxxx)
LE0
Description:
Inhibits lock error SRQ generation.
Inhibits an SRQ from being generated when the Lock Error bit (Primary Status Byte bit three) is set. This is the default mode.
Functional Category
STATUS
LE1
Enables lock error SRQ generation.
Functional Category
STATUS
LE1
Description:
Enables lock error SRQ generation.
Enables an SRQ to be generated when Primary Status Byte bit three (Lock Error) is set and SQ1 has been programmed.
Example:
CSB; SQ1; LE1
Enable an SRQ on lock error.
LF <value> <unit>, <value> <unit>
Sets list frequencies starting at the list index.
Functional Category
LIST SWEEP
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Description:
Permits setting the list frequencies of the selected list starting at the list index specified by the command ELI(xxxx). Any number of frequencies can follow this command. This command does not change the value of the list index.
Places the instrument in list sweep mode, selects list number 1, and sets the list index to 1234. List index 1234 is set to 2 GHz at 2 dBm, list index 1235 is set
Related Commands:
LST, ELN(x), ELI(xxxx)
LGS
Description:
Selects logarithmic step sweep.
LGS, by itself, does not provide a swept-frequency output. It only determines that the swept-frequency output will be a logarithmic step sweep. However, if a frequency sweep is being output by the instrument when LGS is received, the sweep will become a logarithmic step sweep.
Functional Category
SWEEP
Related Commands:
SWP, SSP, LIS
Example:
SDT 3 MS; SNS 100 SPS; LGS; SF1; AUT
Initiates a continuous logarithmic step frequency sweep between F1 and F2. The sweep consists of 100 steps with a dwell-time-per-step of 3 ms.
Output Command:
OSD returns the dwell time of the step sweep (in ms) to the controller. OSS returns the number of steps of the step sweep to the controller. OST returns the sweep time value (in ms) to the controller.
LIB(xxxx)
Description:
Sets the list start index to xxxx.
In list sweep mode, permits setting the list start index for a list sweep of the current list.
Functional Category
LIST SWEEP
Value:
xxxx: 4-digit integer between 0000 and 1999
Example:
LST; ELN1; LIB1097; LIE1197; EXT; LDT 10 MS; TRG
Places the unit in list sweep mode and implements a list sweep from list 1 index 1097 to index 1197 in single trigger mode with a 10 ms dwell-time-per-step, then triggers a single sweep.
Related Commands:
LST, ELN(x), LIE(xxxx)
LIE(xxxx)
Description:
Sets the list stop index to xxxx.
In list sweep mode, permits setting the list stop index for a list sweep of the current list.
Functional Category
LIST SWEEP
Value:
xxxx: 4-digit integer between 0000 and 1999
Example:
LST; ELN1; LIB1097; LIE1197; EXT; LDT 10 MS; TRG
Places the unit in list sweep mode and implements a list sweep from list 1 index 1097 to index 1197 in single trigger mode with a 10 ms dwell-time-per-step, then triggers a single sweep.
Related Commands:
LST, ELN(x), LIB(xxxx)
LIN
Description:
Selects linear power level operation in mV.
In linear power level operation, power level entries and outputs are in mV and power level sweeps are linear.
Functional Category
POWER LEVELING
Related Commands:
LOG, PU0, PU1, PU2
LIS
Description:
Selects linear step sweep.
LIS, by itself, does not provide a swept-frequency output. It only determines that the swept-frequency output will be a linear step sweep. However, if a frequency sweep is being output by the instrument when LIS is received, the sweep will become a linear step sweep. This is the default mode.
Functional Category
SWEEP
Related Commands:
SWP, SSP, LGS
Example:
SDT 3 MS; SNS 100 SPS; LIS; SF1; AUT
Initiates a continuous linear step frequency sweep between F1 and F2. The sweep consists of 100 steps with a dwell-time-per-step of 3 ms.
Output Command:
OSD returns the dwell time of the step sweep (in ms) to the controller. OSS returns the number of steps of the step sweep to the controller. OST returns the sweep time value (in ms) to the controller.
LO0
Description:
Turns off the level offset function.
Functional Category
POWER LEVELING
Related Commands:
LO1, LOS
LO1
Description:
Turns on the level offset function.
The Level Offset function lets you compensate for a device on the MG362x1A’s output that alters the RF output power level at the point of interest. Using the LOS command, you can apply a constant to the level measured by the internal leveling loop that compensates for this loss or gain. The displayed power level will then reflect the actual power level at the test device.
Functional Category
POWER LEVELING
Related Commands:
LO0, LOS
Output Command:
OLO returns the Level Offset value (in dB when in Log mode; in mV when in Linear mode) to the controller.
LOG
Description:
Selects logarithmic power level operation in dBm.
In logarithmic power level operation, power level entries and outputs are in dBm and power level sweeps are logarithmic. This is the default mode.
Functional Category
POWER LEVELING
Related Commands:
LIN, PU0, PU1, PU2
LOS <value> <unit>
Description:
Opens the level offset parameter.
Permits setting the Level Offset constant to be applied to the level measured by the internal leveling loop by the LO1 command.
Functional Category
PARAMETER ENTRY
Value:
+100dB to 100dB (logarithmic); +xxx mV to xxx mV (linear)
Units:
DB (log), VT (linear)
Output Command:
OLO returns the Level Offset value (in dB when in Log mode; in mV when in Linear mode) to the controller.
Related Commands:
LO0, LO1
LP <value> <unit>, <value> <unit>
Description:
Sets list power levels starting at the list index.
Permits setting the power levels of the selected list starting at the list index specified by the command ELI(xxxx). Any number of power levels can follow this command. This command does not change the value of the list index.
Places the instrument in list sweep mode, selects list number 1, and sets the list index to 1234. List index 1234 is set to 2 GHz at 2 dBm, list index 1235 is set
Related Commands:
LST, ELN(x), ELI(xxxx)
LS0
Description:
Inhibits updating of the ESB1 bit 3.
Blocks updating of the Extended Status Byte one bit three (RF Locked). This is the default setting.
Functional Category
STATUS
LS1
Description:
Enables updating of the ESB1 bit 3.
Unblocks updating of the Extended Status Byte one bit three (RF Locked). This bit is only used with the Model 360B interface.
In Power Level Sweep mode, the power level will sweep as determined by the preset (or previously set) dwell-time and number-of-steps parameters.
Functional Category
POWER LEVELING
Related Commands:
LOG, LIN, PDT, PNS, PU0, PU1, PU2
LST
Description:
Places the MG362x1A in list sweep mode.
In list sweep mode, up to four lists of 2000 non-sequential frequency/power level sets can be stored and accessed. A list index (0 through 1999) identifies each frequency/power level set in a list. When commanded, the MG362x1A generates a phase-locked step sweep between the specified list start index and list stop index.
Permits setting the M0 marker frequency parameter. The marker frequency setting is limited by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Output Command:
OM0 returns the M0 marker frequency value (in MHz) to the controller.
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
M1 <value> <unit>
Description:
Opens the M1 parameter.
Permits setting the M1 marker frequency parameter. The marker frequency setting is limited by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Output Command:
OM1 returns the M1 marker frequency value (in MHz) to the controller.
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
M2 <value> <unit>
Description:
Opens the M2 parameter.
Permits setting the M2 marker frequency parameter. The marker frequency setting is limited by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Output Command:
OM2 returns the M2 marker frequency value (in MHz) to the controller.
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
M3 <value> <unit>
Description:
Permits setting the M3 marker frequency parameter. The marker frequency setting is limited by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM3 returns the M3 marker frequency value (in MHz) to the controller.
M4 <value> <unit>
Description:
Opens the M4 parameter.
Permits setting the M4 marker frequency parameter. The marker frequency setting is limited by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM4 returns the M4 marker frequency value (in MHz) to the controller.
M5 <value> <unit>
Description:
Opens the M5 parameter.
Permits setting the M5 marker frequency parameter. The marker frequency setting is limited by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM5 returns the M5 marker frequency value (in MHz) to the controller.
M6 <value> <unit>
Description:
Opens the M6 parameter.
Permits setting the M6 marker frequency parameter. The marker frequency setting is limited by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM6 returns the M6 marker frequency value (in MHz) to the controller.
M7 <value> <unit>
Description:
Opens the M7 parameter.
Permits setting the M7 marker frequency parameter. The marker frequency setting is limited by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM7 returns the M7 marker frequency value (in MHz) to the controller.
M8 <value> <unit>
Description:
Opens the M8 parameter.
Permits setting the M8 marker frequency parameter. The marker frequency setting is limited by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM8 returns the M8 marker frequency value (in MHz) to the controller.
M9 <value> <unit>
Description:
Opens the M9 parameter.
Permits setting the M9 marker frequency parameter. The marker frequency setting is limited by the frequency range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Frequency range of the MG362x1A model
Units:
GH, MH, KH, HZ
Output Command:
OM9 returns the M9 marker frequency value (in MHz) to the controller.
MB0
Description:
Sets the enable mask byte for the primary status byte.
Sets an eight-bit data mask that is used to enable specific bits of the primary status byte. This enables any or all of the bits (except for bit six) in the primary status byte to generate an SRQ using one eight-bit byte. This command can be equivalent to sending FB1, LE1, PE1, SE1, SB1, and UL1.
Functional Category
STATUS
MB1
Description:
Sets the enable mask byte for extended status byte 1.
Sets an eight-bit data mask that is used to enable specific bits of the extended status byte one. This enables any or all of the bits in the extended status byte one to generate an SRQ using one eight-bit byte.
Sets the enable mask byte for extended status byte 2.
Sets an eight-bit data mask that is used to enable specific bits of the extended status byte two. This enables any or all of the bits in the extended status byte two to generate an SRQ using one eight-bit byte.
Functional Category
STATUS
Example:
C code snippet to enable the RF unlocked bit: strcpy(buf, “mb2 ”); buf[3] = 16; snd(source_addr, buf);
ME0
Description:
Disables the marker at the active frequency.
Functional Category
MARKER
Example:
F1; ME0
Disables the F1 frequency marker.
Related Commands:
ME1
ME1
Description:
Enables a marker at the active frequency.
This command will enable a marker at the current frequency that is open for update (F0-F9 or M0-M9).
Functional Category
MARKER
Example:
F6; ME1
Enables the F6 frequency marker.
MKO
Description:
Turns off markers.
Enabled markers remain enabled, but are not active.
Functional Category
MARKER
MNT
Description:
Selects manual trigger.
In manual trigger mode, the list index is incremented by one each time the UP command is received or each time an external TTL trigger is received. The list index is decremented by one each time the DN command is received.
Functional Category
SWEEP
Related Commands:
DN, UP
MOM
Description:
Turns off the AM and FM measurement modes.
Functional Category
MEASURE FUNCTION
Related Commands:
OMM, AMI, FMD
MPN
Description:
Sets the rear panel marker polarity negative.
Functional Category
MARKER
MPP
Description:
Sets the rear panel marker polarity positive.
Functional Category
MARKER
OAD1
Description:
Returns the internal AM depth value (in %) to the controller.
Functional Category
OUTPUT
OAD2
Description:
Returns the internal AM depth value (in dB) to the controller.
Functional Category
OUTPUT
OAR
Description:
Returns the internal AM rate (in Hz) to the controller.
Functional Category
OUTPUT
OAS
Description:
Returns the external AM sensitivity value (in %/V) to the controller.
Functional Category
OUTPUT
OAW
Description:
Returns the name of the internal AM waveform to the controller.
Possible returned values: “SINE”, “SQUARE WAVE”, “RAMP UP”, “RAMP DOWN”, “GAUSSIAN NOISE”, “UNIFORM NOISE”, “TRIANGLE”
Functional Category
OUTPUT
OD1
Description:
Returns the internal pulse delay1 value (in μs) to the controller.
Functional Category
OUTPUT
OD2
Description:
Returns the internal pulse delay2 value (in μs) to the controller.
Functional Category
OUTPUT
OD3
Description:
Returns the internal pulse delay3 value (in μs) to the controller.
Functional Category
OUTPUT
OD4
Description:
Returns the internal pulse delay4 value (in μs) to the controller.
Functional Category
OUTPUT
ODF
Description:
Returns the DF frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
DLF, DFF, DFM
ODP
Description:
Returns the internal pulse delay1 value (in ms) to the controller (Same as OD1).
Functional Category
OUTPUT
OEM
Description:
Returns the Extended SRQ Mask bytes (3 binary bytes) to the controller
Functional Category
OUTPUT
Related Commands
MB0, MB1, MB2
OES
Description:
Returns the GPIB Status bytes (3 binary bytes) to the controller. Returns the contents of the primary status byte and the two extended status bytes to the controller.
Functional Category
OUTPUT
OF0
Description:
Returns the F0 frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
F0
OF1
Description:
Returns the F1 frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
F1
OF2
Description:
Returns the F2 frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
F2
OF3
Description:
Returns the F3 frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
F3
OF4
Description:
Returns the F4 frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
F4
OF5
Description:
Returns the F5 frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
F5
OF6
Description:
Returns the F6 frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
F6
OF7
Description:
Returns the F7 frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
F7
OF8
Description:
Returns the F8 frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
F8
OF9
Description:
Returns the F9 frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
F9
OFD
Description:
Returns the internal FM deviation value (in Hz) to the controller.
Functional Category
OUTPUT
OFH
Description:
Returns the high-end frequency value (in MHz) to the controller.
Functional Category
OUTPUT
OFK
Description:
Returns the internal or external FM lock status to the controller.
Possible returned values: “UNLOCKED NARROW”, “UNLOCKED WIDE”, “LOCKED”
Functional Category
OUTPUT
OFL
Description:
Returns the low-end frequency value (in MHz) to the controller.
Functional Category
OUTPUT
OFM
Description:
Returns the frequency multiplier parameter.
Functional Category
OUTPUT
OFP
Description:
Outputs the current frequency preset.
Functional Category
OUTPUT
OFR
Description:
Returns the internal FM rate (in Hz) to the controller.
Functional Category
OUTPUT
OFS
Description:
Returns the external FM sensitivity value (in MHz/V) to the controller.
Functional Category
OUTPUT
OFW
Description:
Returns the name of the internal FM waveform to the controller.
Possible returned values: “SINE”, “SQUARE WAVE”, “RAMP UP”, “RAMP DOWN”, “GAUSSIAN NOISE”, “UNIFORM NOISE”, “TRIANGLE”
Functional Category
OUTPUT
OI
Description:
Returns the instrument identification string to the controller.
The instrument identification string provides the following information: model number, low-end frequency, high-end frequency, minimum output power level, maximum output power level, software revision number, serial number, model prefix, and series. The string is 36 characters long.
Functional Category
OUTPUT
Example:
The return string “3692 0.0020.00-110.014.02.07000001C” indicates:
Model Series 36, Model Number 92, Minimum Frequency 0.00 GHz (0.01 Hz), Maximum Frequency 20 GHz, Min Power -110 dBm, Max Power 14 dBm, Firmware Version 2.07, Serial Number 000001, and instrument model version letter, in this case a “C” indicating the instrument is an MG362x1A.
OL0
Description:
Returns the L0 power level value to the controller.
Returned power level value is in dBm when in log mode and in mV when in linear mode.
Functional Category
OUTPUT
Related Commands:
L0, XL0
OL1
Description:
Returns the L1 power level value to the controller.
Returned power level value is in dBm when in log mode and in mV when in linear mode.
Functional Category
OUTPUT
Related Commands:
L1, XL1
OL2
Description:
Returns the L2 power level value to the controller.
Returned power level value is in dBm when in log mode and in mV when in linear mode.
Functional Category
OUTPUT
Related Commands:
L2, XL2
OL3
Description:
Returns the L3 power level value to the controller.
Returned power level value is in dBm when in log mode and in mV when in linear mode.
Functional Category
OUTPUT
Related Commands:
L3, XL3
OL4
Description:
Returns the L4 power level value to the controller.
Returned power level value is in dBm when in log mode and in mV when in linear mode.
Functional Category
OUTPUT
Related Commands:
L4, XL4
OL5
Description:
Returns the L5 power level value to the controller.
Returned power level value is in dBm when in log mode and in mV when in linear mode.
Functional Category
OUTPUT
Related Commands:
L5, XL5
OL6
Description:
Returns the L6 power level value to the controller.
Returned power level value is in dBm when in log mode and in mV when in linear mode.
Functional Category
OUTPUT
Related Commands:
L6, XL6
OL7
Description:
Returns the L7 power level value to the controller.
Returned power level value is in dBm when in log mode and in mV when in linear mode.
Functional Category
OUTPUT
Related Commands:
L7, XL7
OL8
Description:
Returns the L8 power level value to the controller.
Returned power level value is in dBm when in log mode and in mV when in linear mode.
Functional Category
OUTPUT
Related Commands:
L8, XL8
OL9
Description:
Returns the L9 power level value to the controller.
Returned power level value is in dBm when in log mode and in mV when in linear mode.
Functional Category
OUTPUT
Related Commands:
L9, XL9
OLO
Description:
Returns the level offset power value to the controller.
Returned level offset power value is in dBm when in log mode and in mV when in linear mode.
Functional Category
OUTPUT
Related Commands:
LOS
OLP
Description:
Output current level preset.
Functional Category
OUTPUT
OM0
Description:
Returns the M0 marker frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
M0
OM1
Description:
Returns the M1 marker frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
M1
OM2
Description:
Returns the M2 marker frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
M2
OM3
Description:
Returns the M3 marker frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
M3
OM4
Description:
Returns the M4 marker frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
M4
OM5
Description:
Returns the M5 marker frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
M5
OM6
Description:
Returns the M6 marker frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
M6
OM7
Description:
Returns the M7 marker frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
M7
OM8
Description:
Returns the M8 marker frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
M8
OM9
Description:
Returns the M9 marker frequency value (in MHz) to the controller.
Functional Category
OUTPUT
Related Commands:
M9
OMD
Description:
Returns the name of the internal pulse mode to the controller.
Possible returned values: “SINGLE”, “DOUBLET”, “TRIPLET”, “QUADRUPLET”
Functional Category
OUTPUT
OMM
Description:
1. Returns the AM depth measurement value (in %) to the controller.
2. Returns the FM deviation measurement value (in MHz) to the controller.
The AM depth measurement is only returned when the AMI command either has been or is also programmed. The FM deviation measurement is only returned when the FMD command either has been or is also programmed.
Functional Category
OUTPUT
OMPS
Description:
Output current pulse modulation state on or off
Functional Category
OUTPUT
OP5
Description:
Returns the internal or external pulse polarity to the controller.
Functional Category
OUTPUT
OPA
Description:
Sends attenuation value in dB.
Functional Category
POWER LEVELING
OPD
Description:
Returns the power sweep dwell time (in ms) to the controller.
Functional Category
OUTPUT
Related Commands:
PDT
OPHD
Description:
Returns the internal ΦM deviation value (in radians) to the controller.
Functional Category
OUTPUT
OPHHS
Description:
Set FM on or off.
Functional Category
MODULATION
OPHM
Description:
Returns the name of the ΦM mode to the controller.
Functional Category
OUTPUT
OPHR
Description:
Returns the internal ΦM rate (in Hz) to the controller.
Functional Category
OUTPUT
OPHS
Description:
Returns the external ΦM sensitivity value (in radians/V) to the controller.
Functional Category
OUTPUT
OPHW
Description:
Returns the name of the internal ΦM waveform to the controller.
Possible returned values: “SINE”, “SQUARE WAVE”, “RAMP UP”, “RAMP DOWN”, “GAUSSIAN NOISE”, “UNIFORM NOISE”, “TRIANGLE”
Functional Category
OUTPUT
OPMS
Description:
Output PM Modulation Source
Functional Category
OUTPUT
OPO
Description:
Returns the phase offset value (in degrees) to the controller.
Functional Category
OUTPUT
Related Commands:
DG, PS0, PS1, PSO, PSZ
OPP
Description:
Returns the internal pulse period value (in ms) to the controller.
Functional Category
OUTPUT
OPR
Description:
Returns the internal pulse frequency (in MHz) to the controller.
Functional Category
OUTPUT
OPS
Description:
Returns the power sweep number-of-steps to the controller.
Functional Category
OUTPUT
Related Commands:
PNS
OPT
Description:
Returns the name of the internal pulse trigger to the controller.
Possible returned values: “FREE RUN”, “GATED”, “DELAYED”, “TRIGGERED”, “TRIG WITH DELAY”, “COMPOSITE”
Functional Category
OUTPUT
OPW
Description:
Returns the internal pulse width1 value (in μs) to the controller (Same as OW1).
Functional Category
OUTPUT
OSB
Description:
Returns the primary GPIB status byte to the controller.
Functional Category
OUTPUT
OSD
Description:
Returns the step sweep dwell time (in ms) to the controller.
Functional Category
OUTPUT
Related Commands:
SDT
OSFM
Description:
Outputs the frequency sweep mode. 1 for SWE1, 0 for CW.
Functional Category
OUTPUT
OSLM
Description:
Outputs the level sweep mode. 1 for SWE2, 0 for FIX.
Functional Category
OUTPUT
OSR
Description:
Returns the self-test results (6 binary bytes) to the controller. Can apply a constant offset to the displayed level value.
Functional Category
OUTPUT
Related Commands:
TST
OSS
Description:
Returns the step sweep number-of-steps to the controller.
Functional Category
OUTPUT
Related Commands:
SNS
OST
Description:
Returns the sweep time value (in ms) to the controller.
Functional Category
OUTPUT
Related Commands:
SWTT
OW1
Description:
Returns the internal pulse width1 value (in μs) to the controller.
Functional Category
OUTPUT
OW2
Description:
Returns the internal pulse width2 value (in μs) to the controller.
Functional Category
OUTPUT
OW3
Description:
Returns the internal pulse width3 value (in μs) to the controller.
Functional Category
OUTPUT
OW4
Description:
Returns the internal pulse width4 value (in μs) to the controller.
Functional Category
OUTPUT
P0
Description:
Turns off the internal or external pulse modulation function (This is the default mode).
Functional Category
MODULATION
Related Commands:
XP
P3
Description:
Valid for the MG362x1A product series and made available to avoid a syntax error when programming the MG362x1A product.
P4
Description:
Valid for the MG362x1A product series and made available to avoid a syntax error when programming the MG362x1A product.
PC1
Description:
Not valid for the MG362x1A product series and made available to avoid a syntax error when programming the MG362x1A product.
PC4
Description:
Not valid for the MG362x1A product series and made available to avoid a syntax error when programming the MG362x1A product.
PDT <value> <unit>
Description:
Opens the power sweep dwell time parameter.
Permits setting the dwell-time-per-step of the power level sweep.
Functional Category
PARAMETER ENTRY
Value:
1 ms to 99 s
Units:
MS, SEC
Output Command:
OPD returns the power sweep dwell time (in ms) to the controller.
PDY <value> <unit>
Description:
Opens the internal pulse delay1 parameter (Same as D1).
Functional Category
PARAMETER ENTRY
Value:
0 ms to 167 ms at 100 MHz pulse clock rate; 0 ms to 419 ms at 40 MHz pulse clock rate; 0 s to 1.6 s at 10 MHz pulse clock rate
Units:
SEC, MS, US, NS
Output Command:
OD1
PE0
Description:
Inhibits parameter range error SRQ generation.
Inhibits an SRQ from being generated when the primary status byte bit four (parameter range error) is set. This is the default mode.
Functional Category
STATUS
PE1
Description:
Enables parameter range error SRQ generation.
Enables an SRQ to be generated when primary status byte bit four (parameter range error) is set and SQ1 has been programmed. SQ1 enables the SRQ generation function.
Related Commands:
SQ1
Example:
CSB; SQ1; PE1
Enables an SRQ on a parameter range error.
PER <value> <unit>
Description:
Opens the internal pulse period parameter.
Functional Category
PARAMETER ENTRY
Value:
See MG362x1A TDS (11410-00928) for clock rates.
Units:
SEC, MS, US, NS
Example:
PER 500 US; W1 50 US; IP
Output Command:
OPP (in μs)
Turns internal pulse on with a period of 500 μs and pulse width of 50 μs.
Related Commands:
PR
PH0
Description:
Turns off the internal or external ΦM function.
If the ΦM function is not installed, this command produces a syntax error.
Functional Category
MODULATION
Related Commands:
PH1, PH2, PH7, PH8
PH1
Description:
Turns on the external ΦM function in narrow mode.
Disables the internal ΦM function or the external ΦM function in Wide mode, had either of these modes been previously programmed. If the ΦM function is not installed, this command produces a syntax error.
Functional Category
MODULATION
Related Commands:
PH0, PH2
PH2
Description:
Turns on the external ΦM function in wide mode.
Disables the internal ΦM function or the external ΦM function in Narrow mode, had either of these modes been previously programmed. If the ΦM function is not installed, this command produces a syntax error.
Functional Category
MODULATION
Related Commands:
PH0, PH1
PH7
Description:
Turns on the internal ΦM function in narrow mode.
Disables the external ΦM function or the internal ΦM function in Wide mode, had either of these modes been previously programmed. If the ΦM function is not installed, this command produces a syntax error.
Functional Category
MODULATION
Related Commands:
PH0, PH8
PH8
Description:
Turns on the internal ΦM function in wide mode.
Disables the external ΦM function or the internal ΦM function in Narrow mode, had either of these modes been previously programmed. If the ΦM function is not installed, this command produces a syntax error.
Functional Category
MODULATION
Related Commands:
PH0, PH7
PHD <value> <unit>
Description:
Opens the internal ΦM deviation parameter.
Functional Category
PARAMETER ENTRY
Value:
0.0025 to 5 radians in Phase Modulation Narrow mode; 0.25 to 500 radians in Phase Modulation Wide mode
Units:
RD
Example:
PHD 0.5 RD; PH7
Turns internal phase modulation narrow on with a phase deviation of 500 mrad.
Output Command:
OPHD (in rads)
PHN
Description:
Turns on the external ΦM function in narrow mode (Same as PH1).
Disables the internal ΦM function or the external ΦM function in Wide mode, had either of these modes been previously programmed. If the ΦM function is not installed, this command produces a syntax error.
Functional Category
MODULATION
Related Commands:
PH0, PH2
PHR <value> <unit>
Description:
Opens the internal ΦM rate parameter.
Functional Category
PARAMETER ENTRY
Value:
0.1 Hz to 1 MHz for sine wave; 0.1 Hz to 100 kHz for square, triangle, and ramp waveforms
Units:
MH, KH, HZ
Example:
PHR 10 KH; PH7
Turns internal phase modulation narrow on with a phase rate 10 kHz.
Output Command:
OPHR (in Hz)
PHS <value> <unit>
Description:
Opens the external ΦM sensitivity parameter.
Functional Category
PARAMETER ENTRY
Value:
±0.0025 radians/V to ±5 radians/V in ΦM Narrow mode; ±0.25 radians/V to ±500 radians/V in ΦM Wide mode
Units:
RV
Example:
PHS 0.5 RV; PH1
Turns external phase modulation narrow on with a phase sensitivity of 500 mrad/V.
Output Command:
OPHS (in rads/V)
PHV(x)
Description:
Selects the internal ΦM waveform.
If the ΦM function is not installed, this command produces a syntax error.
Turns square wave, internal phase modulation narrow on.
Output Command:
OPHW (text)
PHW
Description:
Turns on the external ΦM function in wide mode (Same as PH2).
If the ΦM function is not installed, this command produces a syntax error.
Functional Category
MODULATION
Related Commands:
PH0, PH1
PMD(x)
Description:
Selects the internal pulse mode.
Disables the internal ΦM function or the external ΦM function in Narrow mode, had either of these modes been previously programmed. If the ΦM function is not installed, this command produces a syntax error.
Selects internal pulse triggering on the rising edge of an external input.
The PTR command is active only when pulse trigger is gated, triggered, or triggered with delay.
Functional Category
MODULATION
Example:
PTG4; PTR; IP
Turns internal pulse on and triggers pulse on the rising edge.
Related Commands:
PTF
PU0
Description:
Selects logarithmic power level operation in dBm.
Functional Category
POWER LEVELING
Related Commands:
LIN, LOG, PU1, PU2
PU1
Description:
Selects linear power level operation in mV.
Functional Category
POWER LEVELING
Related Commands:
LIN, LOG, PU0, PU2
PU2
Description:
Selects logarithmic power level operation in dBμV.
Functional Category
POWER LEVELING
Related Commands:
LIN, LOG, PU0, PU2
PW <value> <unit>
Description:
Opens the internal pulse width1 parameter (Same as W1).
Functional Category
PARAMETER ENTRY
Value:
See MG362x1A TDS (11410-00928) for clock rates.
Units:
SEC, MS, US, NS
Related Commands:
W1
Example:
PMD1; PTG1; PW 50 US; IP
Turns internal pulse on with a single pulse, free run of width 50 μs.
Output Command:
OPW (in μs)
RC0
Description:
Selects RF to be off during frequency switching in CW, step sweep, and list sweep modes.
Functional Category
CONFIGURATION
RC1
Description:
Selects RF to be on during frequency switching in CW, step sweep, and list sweep modes.
Functional Category
CONFIGURATION
RF0
Description:
Turns off the RF output.
Functional Category
POWER LEVELING
RF1
Description:
Turns on the RF output. This is the default mode.
Functional Category
POWER LEVELING
RL
Description:
Returns the instrument to local (front panel) control.
Functional Category
MISCELLANEOUS
RO0
Description:
Selects RF to be on at reset.
This is the default mode.
Functional Category
CONFIGURATION
RO1
Description:
Selects RF to be off at reset.
Functional Category
CONFIGURATION
RSN(m)
Description:
Recalls the instrument setup stored in internal setup memory location M.
Recalls a stored instrument setup from the internal setup memory location specified (1 to 9).
Functional Category
STORED SETUP
Value:
M: memory locations 1 to 9
Related Commands:
SSN(m)
RSS
Description:
Resets a sweep if in progress.
Resets the sweep to its start frequency (or power level), whenever the command is received while a single sweep is in progress.
Functional Category
SWEEP
Related Commands:
TRG, TRS
RST
Description:
Resets the MG362x1A to its default settings.
Sending this command clears the current instrument setup. If this setup is needed for future testing, save it as a stored setup using the command SSN(M1-9) or SAF before sending RST.
Remote settings are not reset, e.g. Remote language, Ethernet settings, GPIB settings are not reset.
Functional Category
MISCELLANEOUS
RT0
Description:
Selects RF to be off during retrace.
Functional Category
CONFIGURATION
RT1
Description:
Selects RF to be on during retrace.
Functional Category
CONFIGURATION
SB0
Description:
Inhibits extended status byte 2 SRQ generation.
Inhibits an SRQ from being generated when the Extended Status Byte two bit is set. This is the default mode.
Functional Category
STATUS
SB1
Description:
Enables extended status byte 2 SRQ generation.
Enables an SRQ to be generated when primary status byte bit seven (extended status byte two) is set and SQ1 has been programmed. The extended status byte two bit is set whenever one (or more) of the unmasked status reporting functions in extended status byte two is true.
Permits setting the dwell-time-per-step value for the step sweep.
Related Commands:
LIS, LGS, SSP, SNS, SWT
Output Command:
OSD returns the dwell time of the step sweep (in ms) to the controller.
SF1
Description:
Selects the F1-F2 sweep mode.
Selects a frequency sweep between F1 and F2.
Functional Category
SWEEP
Related Commands:
F1, F2
SF3
Description:
Selects the F3-F4 sweep mode.
Selects a frequency sweep between F3 and F4.
Functional Category
SWEEP
Related Commands:
F3, F4
SNS <value> <unit>
Description:
Opens the step sweep number of steps parameter.
Permits setting the number of steps for the step sweep.
Functional Category
PARAMETER ENTRY
Value:
1 to 10,000
Units:
SPS
Related Commands:
LIS, LGS, SDT, SSP, SWT
Output Command:
OSS returns the number of steps of the step sweep to the controller.
SQ0
Description:
Disables the SRQ generation function. This is the default mode.
Functional Category
STATUS
SQ1
Description:
Enables the SRQ generation function. Enables SRQ generation when a Primary byte bit is set.
This command allow a status reporting function, that is true and enabled, to pull the SRQ line LOW (true) and request service from the controller.
Functional Category
STATUS
SQP
Description:
Turns on internal 1 kHz square wave pulse modulation.
Sets the internal pulse width (W1) to 500 ms, PRF to 1 kHz, pulse mode to single, pulse trigger to free run, and turns on the internal pulse modulation function. Disables the external pulse modulation function, if previously programmed.
Functional Category
MODULATION
Units:
MH
Related Commands:
PMD(x), PTG(x), IP, W1, PER
SSN(m)
Description:
Saves the current instrument setup to internal setup memory location M specified as (1 to 9).
Functional Category
STORED SETUP
Value:
m: memory locations 1 to 9
Related Commands:
RSN(m)
SSP
Description:
Selects linear step sweep.
Same as LIS. SSP, by itself, does not provide a swept-frequency output. It only determines that the swept-frequency output will be a linear step sweep. However, if a frequency sweep is being output by the instrument when SSP is received, the sweep will become a linear step sweep.
Functional Category
SWEEP
Example:
SDT 3 MS; SNS 100 SPS; SSP; SF1; AUT
Initiates a continuous linear step frequency sweep between F1 and F2. The sweep consists of 100 steps with a dwell-time-per-step of 3 ms.
Output Command:
OSD returns the dwell time of the step sweep (in ms) to the controller. OSS returns the number of steps of the step sweep to the controller. OST returns the sweep time value (in ms) to the controller.
Related Commands:
LGS, LIS, SWP
SW0
Description:
Turns off the internal or external pulse modulation function (Same as P0).
Functional Category
MODULATION
SW1
Description:
Turns on internal 400 Hz square wave pulse modulation.
Sets the internal pulse width (W1) to 1.25 ms, PRF to 400 Hz, pulse mode to single, trigger to free run, pulse clock rate to 100 MHz (or 40 MHz with Option 24), and turns on the internal pulse modulation function. Disables the external pulse modulation function, if previously programmed.
Functional Category
MODULATION
SW2
Description:
Turns on internal 1 kHz square wave pulse modulation (Same as SQP).
Sets the internal pulse width (W1) to 500 ms, PRF to 1 kHz, pulse mode to single, pulse trigger to free run, and turns on the internal pulse modulation function. Disables the external pulse modulation function, if previously programmed.
Functional Category
MODULATION
SW3
Description:
Turns on internal 7.8125 kHz square wave pulse modulation.
Sets the internal pulse width (W1) to 64 ms, PRF to 7.8125 kHz, pulse mode to single, pulse trigger to free run, and turns on the internal pulse modulation function. Disables the external pulse modulation function, if previously programmed.
Functional Category
MODULATION
SW4
Description:
Turns on internal 27.8 kHz square wave pulse modulation.
Sets the internal pulse width (W1) to 18 ms, PRF to 27.8 kHz, pulse mode to single, pulse trigger to free run, pulse clock rate to 100 MHz (or 40 MHz with Option 24), and turns on the internal pulse modulation function. Disables the external pulse modulation function, if previously programmed.
Functional Category
MODULATION
SWP
Description:
Selects analog sweep.
Functional Category
SWEEP
SWT <value> <unit>
Description:
Opens the sweep time parameter.
Permits setting the time for the CW ramp, analog sweep, or step sweep.
Functional Category
PARAMETER ENTRY
Value:
30ms to 99s
Units:
MS, SEC
Output Command:
OST returns the sweep time value (in ms) to the controller.
Related Commands:
LIS, LGS, SSP
SYST:LANG "SCPI”
Description:
Selects the instrument's external interface language. The double quotes are required and will be returned with the query reply.
"NATIVE" changes the instrument's external interface language to the Anritsu Native language mode. Any commands issued within 1 second of the change may be garbled or lost.
Note: When changing from NATIVE to SCPI interface language, use the command SYST:LANG "SCPI". Do not use the long form of the command, do not use a leading colon (:), and use exactly one space ( ) between the 'G' and the opening double quote ("). Deviations from this exact command may result in a syntax error.
Related Commands:
SYST:LANG
Default Value:
SCPI
TEX
Description:
Sets sweep trigger to external.
Functional Category
SWEEP
Related Commands:
EXT, AUT
TRG
Description:
Triggers a single sweep.
When single sweep triggering is selected for frequency, power level, or list sweeps, a single sweep starts when the TRG command is received.
Functional Category
SWEEP
Related Commands:
EXT, TRS, RSS
TRS
Description:
Triggers a single sweep.
Same as TRG. When single sweep triggering is selected for frequency, power level, or list sweeps, a single sweep starts when the TRS command is received.
Functional Category
SWEEP
Related Commands:
EXT, TRG, RSS
TST
Description:
Starts an instrument self-test.
When a TST command is received, the signal generator performs a self test, then returns a single byte, unterminated. A "P" (for pass) or an "F" (for fail). Extended Status Byte one bit zero is set if self test fails; bit two is set when self test is complete.
Functional Category
SELF TEST
Output Command:
OSR returns the six self test results bytes to the controller.
UL0
Description:
Inhibits RF unleveled SRQ generation.
Inhibits an SRQ from being generated when the RF Unleveled bit is set. This is the default mode.
Functional Category
STATUS
UL1
Description:
Enables RF unleveled SRQ generation. Enables RF Unleveled bit update. Primary byte bit 2.
Enables an SRQ to be generated when Primary Status Byte bit two (RF Unleveled) is set and SQ1 has been programmed.
Functional Category
STATUS
Example:
SQ1; UL1
Enables an unleveled SRQ generation.
VM1
Description:
Turns on the video marker mode.
When the video marker mode is turned on, the MG362x1A generates a pulse output at each marker frequency that can be seen on a CRT display.
Functional Category
MARKER
Related Commands:
MK0, ME1, ME0
W1 <value> <unit>
Description:
Opens the internal pulse width 1 parameter.
Functional Category
PARAMETER ENTRY
Value:
30 ns to 167 ms with 100 MHz pulse clock rate; 25 ns to 419 ms at 40 MHz pulse clock rate; 100 ns to 1.6 s at 10 MHz pulse clock rate
Units:
SEC, MS, US, NS
Example:
W1 50 US; PER 100 US; PMD1; PTG1; IP
Turns internal pulse on with a 50 μs pulse every 100 μs.
Output Command:
OW1 (in μs)
W2 <value> <unit>
Description:
Opens the internal pulse width 2 parameter.
Functional Category
PARAMETER ENTRY
Value:
30 ns to 167 ms with 100 MHz pulse clock rate; 25 ns to 419 ms at 40 MHz pulse clock rate; 100 ns to 1.6 s at 10 MHz pulse clock rate
Units:
SEC, MS, US, NS
Output Command:
OW2 (in μs)
W3 <value> <unit>
Description:
Opens the internal pulse width 3 parameter.
Functional Category
PARAMETER ENTRY
Value:
30 ns to 167 ms with 100 MHz pulse clock rate; 25 ns to 419 ms at 40 MHz pulse clock rate; 100 ns to 1.6 s at 10 MHz pulse clock rate
Units:
SEC, MS, US, NS
Output Command:
OW3 (in μs)
W4 <value> <unit>
Description:
Opens the internal pulse width 4 parameter.
Functional Category
PARAMETER ENTRY
Value:
30 ns to 167 ms with 100 MHz pulse clock rate; 25 ns to 419 ms at 40 MHz pulse clock rate; 100 ns to 1.6 s at 10 MHz pulse clock rate
Units:
SEC, MS, US, NS
Output Command:
OW4 (μs)
XL0 <value> <unit>
Description:
Opens the L0 parameter.
Permits setting the L0 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL0 returns the L0 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
XL1 <value> <unit>
Description:
Opens the L1 parameter.
Permits setting the L1 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL1 returns the L1 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
XL2 <value> <unit>
Description:
Opens the L2 parameter.
Permits setting the L2 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL2 returns the L2 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
XL3 <value> <unit>
Description:
Opens the L3 parameter.
Permits setting the L3 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL3 returns the L3 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
XL4 <value> <unit>
Description:
Opens the L4 parameter.
Permits setting the L4 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL4 returns the L4 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
XL5 <value> <unit>
Description:
Opens the L5 parameter.
Permits setting the L5 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL5 returns the L5 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
XL6 <value> <unit>
Description:
Opens the L6 parameter.
Permits setting the L6 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL6 returns the L6 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
XL7 <value> <unit>
Description:
Opens the L7 parameter.
Permits setting the L7 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL7 returns the L7 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
XL8 <value> <unit>
Description:
Opens the L8 parameter.
Permits setting the L8 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL8 returns the L8 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
XL9 <value> <unit>
Description:
Opens the L9 parameter.
Permits setting the L9 parameter. The power level setting is determined by the power level range of the particular MG362x1A model.
Functional Category
PARAMETER ENTRY
Value:
Power level range of the MG362x1A model
Units:
DM (log), VT (linear)
Output Command:
OL9 returns the L9 power level (in dBm when in log mode; in mV when in linear mode) to the controller.
Related Commands:
LOG, LIN, PU0, PU1, PU2
XP
Description:
Turns on the external pulse modulation function. Disables the internal pulse modulation function, if previously programmed.
Turns on the pulse modulation function. If pulse modulation is not installed, this command produces a syntax error.